Service manual
3. DIAGNOSTIC CHECK
Service Manual BSM-4100 3.51
SELF-TEST-PRO No.2
This check program checks the paper speed, slip and snaking. A 0.5 Hz rectangular
waveform and bar graph are recorded on the paper at 25 mm/s paper speed by
pressing the NIBP INTERVAL key.
To stop the recording, press the NIBP INTERVAL key again.
25 mm/sec
0.5 Hz rectangular waveform
10 mm
25 mm/s +5%
−
(47.5 to 52.5 mm)