BSM- 4101A BSM- 4101J BSM- 4101K BSM- 4102A BSM- 4103A BSM- 4103J BSM- 4103K BSM- 4104A BSM- 4111A BSM- 4111J BSM- 4111K BSM- 4112A BSM- 4113A BSM- 4113J BSM- 4113K BSM- 4114A SERVICE MANUAL Life Scope P BEDSIDE MONITOR BSM-4100 0634-001539E
CONTENTS Contents GENERAL HANDLING PRECAUTIONS ......................................................................... i WARRANTY POLICY .................................................................................................... ii EMC RELATED CAUTION ............................................................................................ iii Conventions Used in this Manual and Instrument ......................................................... v Warnings, Cautions and Notes ...............
CONTENTS Names and Function of Parts ............................................................................................. 1.18 Front Panel ............................................................................................................... 1.18 Socket Panel ............................................................................................................ 1.20 On the BSM-4101/4103/4111/4113 Monitors .................................................. 1.
CONTENTS Hard Key Check ............................................................................................. 3.12 Touch Key Check ........................................................................................... 3.13 Sound Check ................................................................................................. 3.14 COM RAM Check .......................................................................................... 3.15 ALARM INDICATOR Check .............................
CONTENTS Section 5 Disassembly and Assembly ............................................... 5C.1 Before You Begin .................................................................................................................. 5.1 Warnings and Caution ................................................................................................. 5.1 Required Tools ............................................................................................................ 5.1 Opening the Instrument ...
CONTENTS Section 8 Replaceable Parts List......................................................... 8C.1 Bedside Monitor BSM-4101/4103/4111/4113 ........................................................................ 8.2 Bedside Monitor BSM-4102/4104/4112/4114 (MP304* included) ........................................ 8.10 Bedside Monitor BSM-4102/4104/4112/4114 (MP506* included) ........................................ 8.18 RG-921P Recorder Unit ...........................................................
CONTENTS CN101 (for LCD unit) ...................................................................................... 9.21 CN102 (for touch screen) ............................................................................... 9.22 CN103 (for MAIN board) ................................................................................. 9.22 Inverter Board .................................................................................................................... 9.23 CN1 (for MAIN board) ............
GENERAL HANDLING PRECAUTIONS This device is intended for use only by qualified medical personnel. Use only Nihon Kohden approved products with this device. Use of non-approved products or in a non-approved manner may affect the performance specifications of the device. This includes, but is not limited to, batteries, recording paper, pens, extension cables, electrode leads, input boxes and AC power. Please read these precautions thoroughly before attempting to operate the instrument. 1.
(3) Technical information such as parts list, descriptions, calibration instructions or other information is available for qualified user technical personnel upon request from your Nihon Kohden distributor. 9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or location of electrodes and/or transducers to avoid possible burn to the patient. 10.
EMC RELATED CAUTION This equipment and/or system complies with the International Standard IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or system. However, an electromagnetic environment that exceeds the limits or levels stipulated in the IEC 60601-1-2, can cause harmful interference to the equipment and/or system or cause the equipment and/or system to fail to perform its intended function or degrade its intended performance.
In IEC 60601-1-2 Medical Electronic Equipment, Part 1: General Requirements for Safety, 2. Collateral Standard: Electromagnetic compatibility-Requirements and test. Section 36. 202. 2 Radiated radio-frequency electromagnetic fields, PATIENT COUPLED EQUIPMENT and/or SYSTEMS applicable IMMUNITY test methods are under consideration at SC62A/WG13.
Conventions Used in this Manual and Instrument Warnings, Cautions and Notes Warnings, cautions and notes are used in this manual to alert or signal the reader to specific information. WARNING A warning alerts the user to possible injury or death associated with the use or misuse of the instrument.
Section 1 General Introduction ......................................................................................................................... 1.1 General Information on Servicing ........................................................................................ 1.2 Service Policy, Service Parts and Patient Safety Checks ................................................... 1.4 Service Policy .....................................................................................................
Environment .................................................................................................. 1.17 Dimensions and Weight (approximate) .......................................................... 1.17 Electromagnetic Compatibility ....................................................................... 1.17 Safety Standard ............................................................................................ 1.17 Names and Function of Parts ...............................................
1. GENERAL Introduction This service manual provides useful information to qualified personnel to understand, troubleshoot, service, maintain and repair the BSM-4100A/J/K Bedside Monitor (referred to as “the instrument” in this service manual). All replaceable parts or units of this instrument are clearly listed with exploded illustrations to help you locate the parts quickly. The “Maintenance” section in this service manual describes the maintenance that should be performed by qualified service personnel.
1. GENERAL General Information on Servicing Note the following information when servicing the instrument. CAUTION Safety • There is the possibility that the outside surface of the instrument, such as the operation keys, could be contaminated by contagious germs, so disinfect and clean the instrument before servicing it. When servicing the instrument, wear rubber gloves to protect yourself from infection.
1. GENERAL Transport • Use the specified shipment container and packing material to transport the instrument. If necessary, double pack the instrument. Also, put the instrument into the shipment container after packing so that the buffer material does not get inside the instrument. • When transporting a board or unit of the instrument, be sure to use a conductive bag. Never use an aluminum bag when transporting the power board, power unit or board on which a lithium battery is mounted.
1. GENERAL Service Policy, Service Parts and Patient Safety Checks Service Policy Our technical service policy for this instrument is to replace the faulty unit, board or part or damaged mechanical part with a new one. Do not perform electrical device or component level repair of the multilayer board or unit.
1. GENERAL Patient Safety Checks Periodic maintenance procedures and diagnostic check procedures are provided in this manual to ensure that the instrument is operating in accordance with its design and production specifications. To verify that the instrument is working in a safe manner with regard to patient safety, patient safety checks should be performed on the instrument before it is first installed, periodically after installation, and after any repair is made on the instrument.
1. GENERAL Composition BSM-4101/4103/4111/4113 BSM-4101A BSM-4101J BSM-4101K BSM-4111A BSM-4111J BSM-4111K BSM-4103A BSM-4103J BSM-4103K BSM-4113A BSM-4113J BSM-4113K 1.6 CD-225P Chassis RG-921P Recorder Unit (BSM-4101/4103) UR-3586 MAIN Board UR-3563 DPU Board UR-3566 Motherboard UR-3541/ UR-3939* Connection Board * Hardware revision no.
1. GENERAL Service Manual BSM-4100 YS-024R0 Operation Panel YS-024R1 Input Panel (BSM-4101/4111) YS-024R3 Input Panel (BSM-4103/4113) QM-421P Flash Disk Card (Option) ZB-800PA Transmitter (Option) ZB-800PG Transmitter (Option) ZB-800PK Transmitter (Option) QI-101P Network Card (Option) QI-111P Network Printer Card (Option) 1.
1. GENERAL BSM-4102/4104/4112/4114 BSM-4102A BSM-4112A BSM-4104A BSM-4114A 1.8 CD-226P Chassis RG-921P Recorder Unit (BSM-4102/4104) UR-3586 MAIN Board UR-35631 DPU Board UR-35661 Motherboard UR-3541/ UR-3939* Connection Board * Hardware revision no.
1. GENERAL Service Manual BSM-4100 YS-024R0 Operation Panel YS-024R2 Input Panel (BSM-4102/4112) YZ-024R4 Input Panel (BSM-4104/4114) QM-421P Flash Disk Card (Option) ZB-800PA Transmitter (Option) ZB-800PG Transmitter (Option) ZB-800PK Transmitter (Option) QI-101P Network Card (Option) QI-111P Network Printer Card (Option) 1.
1. GENERAL Specifications Display Display size: Waveform display mode: Viewing area: Resolution: Maximum number of waveform trace: Sweep speed: Sweep time: Waveform display color: Numeric display color: Waveform freeze: Display waveforms: Numerical data display: Synchronization mark: 12.1 inch, TFT type LCD Non-fade moving or non-fade fixed 246 mm × 184.5 mm 800 × 600 dots 13 traces (12 lead display) 25 mm/s, 50 mm/s (At compressed: 6.25 mm/s, Respiration low speed: 1.56 mm/s) 7.
1. GENERAL Arrhythmia message: Arrhythmia recall: Number of recall files: ASYSTOLE, VT, VF, VPC RUN, COUPLET, EARLY VPC, BIGEMINY, FREQ VPC, TACHYCARDIA, BRADYCARDIA, NOISE, Electrode OFF, LEARNING 16 With QM-421P flash disk card, 100 8s Storage time per file: ST level measurement: Number of measurement channels:3-electrodes: 1 ch 6-electrodes: 8 ch 10-electrodes: 12 ch Measuring range: ±2.
1.
1. GENERAL Non Invasive Blood pressure, NIBP Measuring method: Oscillometric Cuff pressure display range: 0 to 300 mmHg Accuracy: ±3 mmHg (0 mmHg ≤ NIBP ≤ 200 mmHg) ±4 mmHg (200 mmHg ≤ NIBP ≤ 300 mmHg) Safety: Maximum pressurization value cuff inflation limiter: Adult 300 to 330 mmHg Neonates 150 to 165 mmHg Cull inflation time limiter: Adult 161 to 165 s Neonates 81 to 84 s Measurement mode: Manual Continuous Periodic: 2, 2.
1.
1.
1. GENERAL ECG/BP Output Outputs 100 mmHg/V IBP waveform of the IBP connection cord connected to the multi-parameter 1 socket, 1 mV/V ECG waveform and heart rate trigger. Output impedance ECG: ≤100 Ω BP: ≤100 Ω Output-waveform ECG: ±5.0 V (at 1 mV/V sensitivity) BP: −0.5 to +3.0 V(at 100 mmHg/V sensitivity) Heart rate Trigger: Open collector output Frequency response ECG: 0.
1.
1. GENERAL Names and Function of Parts Front Panel 2 1 3 13 SILENCE ALARMS 4 5 NIBP INTERVAL START/STOP 6 MENU 7 HOME 8 12 9 11 10 20 14 1.
1. GENERAL No. Name Description Red or orange lamp blinks according to the alarm settings. Green lamp blinks in synchronization with the patient’s QRS. 1 Alarm indicator 2 Handle For carrying the instrument. 3 SILENCE ALARMS key Silences the alarm sound. 4 Function dial 1 and 2 Used for setting alarm limits and scrolling list or trendgraphs. When the function dial lamp lights, the function dial is available. 5 NIBP INTERVAL key Selects NIBP measurement mode.
1. GENERAL Socket Panel On the BSM-4101/4103/4111/4113 Monitors Refer to Warning and Caution in the "General Safety Information" in the operator's manual. SpO2 socket Connects to the SpO2 connection cord. ECG/RESP socket Connects to the ECG connection cord. Multi-parameter sockets Connect to the connection cord of the parameter to be monitored (IBP, temperature, CO, CO2, FiO2 or respiration by thermistor method). The type of parameter is automatically recognized.
1. GENERAL Using the Output Signal from the ECG/BP OUT Socket WARNING When using the output signal from the monitor as the synchronization signal for other equipment such as an IABP (intra-aortic balloon pump) or defibrillator: • Set the timing of the IABP by checking the waveform on the IABP screen. • Check the condition of the bedside monitor at all times. The output signal may become unstable.
1. GENERAL Left Side Panel CAUTION • Use only the Nihon Kohden card. • Do not press the PC card eject button while the PC card lamp is lit. This may damage the PC card data and interrupt network communication. PC card eject button PC card lamp Refer to "Connecting the Power Cord and Grounding the Monitor" in Section 2. Fuse holder AC SOURCE power cord socket For the AC power cord. Equipotential grounding terminal For an equipotential grounding lead. PC card slot For an optional QI-101P network card.
1. GENERAL Right Side Panel CAUTION Connect only the ZB-800P transmitter to the ZB socket by following the specified procedure. Otherwise, electrical leakage current may harm the patient and operator. ZB socket For the ZB-800P transmitter. Battery pack holder Service Manual BSM-4100 1.
1. GENERAL Storage and Transport Follow these procedures when storing or transporting the instrument. Storage Before storing the instrument for a long time, perform the following steps: 1. Disconnect the power cord from the instrument. 2. Cover the instrument with a dust cover. 3. If possible, store the instrument in its original shipping container. 4. Make sure the storage place meets the following storage conditions for the duration of the storage.
1. GENERAL Hard Keys and Soft Keys Hard Keys The instrument has six hard keys: SILENCE ALARMS, NIBP INTERVAL, NIBP START/STOP, MENU, HOME and RECORD. These keys always have the same functions, regardless of the screen display. In this manual, hard keys are indicated by CAPITAL LETTERS. Soft Keys When the MENU key is pressed or the screen is touched, the screen displays several keys which have different functions depending on the screen display.
1. GENERAL Upgrading the System Software and Changing Language on the Screen CAUTION Upgrading the system software and changing the language on screen erases all system and monitoring settings. Write down these settings so they can be re-entered after the software upgrade. The instrument uses a program card for upgrading its system software and changing the screen language.
Service Manual BSM-4100 MP3 MP2 *2BSM-4103/4104/4113/4114 only MP4 Board MP3 Board*2 UR-3559/35592 MP2 Board UR-3558 MP1 Board UR-3557 SpO2 MP1 SpO2 Board*1 UR-3561/35611 NIBP ECG RESP2 Board UR-3547 ECG RESP3 Board UR-3548 ECG/BP Out ECG/RESP ECG RESP1 Board UR-3546 Motherboard UR-3566/ 35661 NIBP Safety Board UR-3567 Transmitter *1BSM-4101/4103/4111/4113 (NK SpO2 board) BSM-4102/4104/4112/4114 (Nellcor SpO2 dummy board) Function Dial Board UR-3544 NIBP Measure Board UR-3564 DPU Board
1.
Section 2 Troubleshooting General ................................................................................................................................ 2.1 Power-Related Problems ..................................................................................................... 2.2 Display Problems ................................................................................................................ 2.3 Sound Problems ..................................................................
2. TROUBLESHOOTING General Use the troubleshooting tables to locate, identify, and solve a problem in the instrument. The problems are divided into general problem areas. Each category has its own troubleshooting table for fast and easy troubleshooting.
2. TROUBLESHOOTING Power-Related Problems Problem Possible Cause/Criteria No beep sound and No AC power input. AC power lamp does One or both AC inlet fuses is blown. not light in AC power operation. Faulty connection between CN105 on the MAIN board and power unit. Faulty connection between CN103 on the MAIN board and power SW board. Faulty power supply unit. Faulty power SW board. No beep sound and The battery is discharged. the battery power The battery is damaged.
2. TROUBLESHOOTING Display Problems Problem No display and the backlight does not light. No display, but the backlight lights. Fine vertical or horizontal lines on the LCD. Thick vertical or horizontal stripes on the screen. The characters or waveforms are distorted. The display is dim and the brightness control does not make it brighter. The Diagnostic Check screen appears in the monitor mode and “OK” is displayed at the POWER ON CHECK RESULT.
2. TROUBLESHOOTING Sound Problems Problem No sound. No sound, except for a hissing sound. The sound is muffled or distorted. Possible Cause/Criteria Faulty connection between CN006 on the MAIN board and the speaker. Faulty speaker. Action Check the connection. Replace the speaker. Faulty MAIN board. Replace the MAIN board. Faulty MAIN board. Replace the MAIN board. Faulty speaker. Replace the speaker. The speaker is blocked. Remove the object blocking the speaker.
2. TROUBLESHOOTING Recording Problems Problem Recording paper does not feed properly. Possible Cause/Criteria Faulty connection between CN103 on the Recording mark does MAIN board and Power SW board. not appear on the Faulty Power SW board. screen when the record key is pressed. Faulty MAIN board. Paper magazine open mark appears on the screen. Out of paper mark appears on the screen.
2. TROUBLESHOOTING ECG and Respiration by Impedance Method Problems Problem ATTACH ELECTRODES message is displayed even after each electrode contact part of the electrode lead is short-circuited. The time constant setting cannot be changed. Pacing spike is not detected. The ECG waveform is displayed with a different ECG sensitivity from the sensitivity setting. Monitoring lead setting cannot be changed but the time constant setting can be changed. Display of 12 lead ECG has something wrong.
2. TROUBLESHOOTING SpO2 Problems Problem The instrument does not recognize the NK SpO2 connection cord connected to the SpO2 socket on the instrument. PULSE SEARCH message is displayed even after the probe is replaced and attachment site is changed. WEAK PULSE message is displayed even after the probe is replaced and attachment site is changed. SpO2 MODULE ERROR message is displayed.
2. TROUBLESHOOTING Non-invasive Blood Pressure Problems Problem NIBP MODULE FAILURE message is displayed. NIBP SAFETY CIRCUIT RUNNING message is displayed. AIR LEAK message is displayed. The instrument does not recognize the air hose connected to the NIBP socket on the instrument. Possible Cause/Criteria DPU board or NIBP Safety board is faulty. Action Replace the faulty board with a new one. Air hose is blocked.
2. TROUBLESHOOTING Other Vital Sign Input Problems Problem The instrument does not recognize a connection cord at a multi-parameter socket. Possible Cause/Criteria The connection cord is faulty. MPU MODULE ERROR message is displayed. The connection cable between the motherboard and DPU board is disconnected or the MP2 board, motherboard or DPU board is faulty if this message always appears even after the instrument is turned off and on again. MPU FAILURE message is displayed.
2. TROUBLESHOOTING Problem IBP measurement data on the screen is wrong. Temperature cannot be measured. Possible Cause/Criteria The blood pressure transducer is faulty. The MP2 board or MP1 board is faulty if the blood pressure data is wrong at all the multi-parameter sockets and temperature, FiO2 and respiration rate (thermistor method) are not wrong at any multi-parameter sockets.
2. TROUBLESHOOTING Problem FiO2 cannot be measured. Possible Cause/Criteria The O2 sensor or connection cord is faulty. The MP2 board is faulty if the FiO2 cannot be measured but the T1 temperature can be measured at all the multi-parameter sockets or the FiO2 can be measured at one of the multi-parameter sockets. FiO2 data on the screen is wrong. CO2 cannot be measured. The MP3 board, MP2 board or motherboard is faulty if the FiO2 cannot be measured at the No. 3 multiparameter socket only.
2. TROUBLESHOOTING ECG/BP Output Problems Problem The instrument does not output the ECG, HT (heart rate) or BP signal from the ECG/BP OUT socket. The ECG or BP waveform from the ECG/BP OUT socket is wrong. Possible Cause/Criteria The ECG/BP output cable is faulty. The ECG and HT signals do not output without ECG measurement. The BP signal is not output if the blood pressure is measured at the No. 2 or No. 3 multi-parameter sockets.
2. TROUBLESHOOTING Option Problems Problem The instrument does not display the full disclosure window. Possible Cause/Criteria The optional QM-421P flash disk card is not installed. The QM-421P installation, QM-421P card or MAIN board is faulty if “QM-421P” is not displayed at the “OPTION” on the DIAGNOSTIC CHECK screen. “SYSTEM INITIALIZE” is not done or the QM-421P card or MAIN board is faulty if “QM-421P” is displayed at the “OPTION” on the DIAGNOSTIC CHECK screen.
Section 3 Diagnostic Check Introduction ......................................................................................................................... 3.1 Power On Self Check .......................................................................................................... 3.2 Calling Up the DIAGNOSTIC CHECK Screen ...................................................................... 3.4 Calling Up the Error History .......................................................................
COM Check Menu Items ......................................................................................... 3.43 WS RECORDER Check ................................................................................ 3.43 SERIAL I/F Check ........................................................................................ 3.44 NETWORK CARD Check .............................................................................. 3.45 MULTI PARAMETER UNIT Check ...................................................
3. DIAGNOSTIC CHECK Introduction The instrument has two check programs which automatically or manually diagnose the instrument. The automatic check program (power on self check) is executed whenever the instrument is turned on. The manual check program can be executed at any time. All errors detected during the power on self check, manual check, and any time in monitor mode are stored in an error history table that you can view.
3. DIAGNOSTIC CHECK Power On Self Check The following icon appears during the power on self check. If no error is detected, the monitor mode begins and the patient monitoring display appears. If a serious error which does not allow the host CPU to execute the system program any more is detected, the screen changes to the following DIAGNOSTIC CHECK screen and the error mesage is displayed.
3. DIAGNOSTIC CHECK The MAIN board and DPU board are checked as follows: • CPU ROM CHECK: Checks the system ROM for the host CPU on the MAIN board. The MAIN board will be faulty if an error which causes a partial screen problem is found. • CPU RAM CHECK: Checks the system RAM for the host CPU and patient data storage. The MAIN board will be faulty if an error which causes a screen problem is found. • DPU ROM CHECK: Checks the system ROM for the CPU on the DPU board.
3. DIAGNOSTIC CHECK Calling Up the DIAGNOSTIC CHECK Screen Use this screen to view the error history or perform manual check or system setup and initialization. 1. With the power off, press the Power switch while pressing the SILENCE ALARMS key. Continue pressing the SILENCE ALARMS key until the DIAGNOSTIC CHECK screen appears. 2. To exit the DIAGNOSTIC CHECK screen and return to the patient monitoring mode, touch the “MONITOR MODE” key. 3.
3. DIAGNOSTIC CHECK Calling Up the Error History The ERROR HISTORY screen shows the error history of the instrument. The error code, board name, time, and date the error occurred are listed on this screen. For an explanation of error codes, see the “Error Codes” section. The error history data is deleted whenever system initialization is performed. 1. From the DIAGNOSTIC CHECK screen, touch the “HISTORY” key. The HISTORY screen appears. Then, touch the “ERROR HISTORY” key to display the error history. 2.
3.
3. DIAGNOSTIC CHECK Initializing the System You can delete all stored waveforms and data and the error history and return all settings to the factory defaults. CAUTION Initializing the system deletes all the patient data and error history, and returns the setting conditions to their default settings. 1. From the DIAGNOSTIC CHECK screen, touch the “SYSTEM INITIALIZE” key until a SYSTEM INITIALIZE confirmation message appears. (Approximately 3 seconds) All data and user settings are deleted. 2.
3. DIAGNOSTIC CHECK Performing Manual Check and Other Checks The following checks are available.
3. DIAGNOSTIC CHECK Calling Up the MANUAL CHECK MENU Screen 1. From the DIAGNOSTIC CHECK screen, touch the “MANUAL CHECK” key. The MANUAL CHECK MENU screen appears. 2. To test each item from the MANUAL CHECK MENU screen, touch a menu item key. 3. To return to the DIAGNOSTIC CHECK screen, touch the “RETURN” key. 4. To exit the DIAGNOSTIC CHECK screen and return to the patient monitoring mode, touch the “MONITOR MODE” key. Service Manual BSM-4100 3.
3. DIAGNOSTIC CHECK CPU Check Menu Items The CPU check menu items check the function of the CPU peripherals such as ROMs, RAMs, Switches, Touch Key, Sound generators, Alarm indicators and external memory modules. ROM Check This item checks the ROMs on the MAIN board by the CRC (cyclic redundancy check) technique. This check assigns a count number when one set of checks is successfully completed. A set of checks takes about 1 minute and 30 seconds.
3. DIAGNOSTIC CHECK RAM Check This item checks the RAM on the MAIN board by comparing the test patterns it first wrote to the RAM with the test patterns it later read from the RAM. This test uses 15 test patterns and assigns a count number when a comparison of the 15 test patterns is successfully completed. The comparison of all the test patterns in this check takes about 4 minutes. This check deletes the patient data. If an error message appears, the MAIN board will be faulty.
3. DIAGNOSTIC CHECK Hard Key Check This item checks the operation of the hard keys on the front panel and the function dials. The hard key and dial layout appear. Pressing the hard key or rotating the function dial usually highlights the corresponding hard key or dial on the screen. If the pressed hard key is not highlighted, the hard key may be faulty. If more than one hard key has this symptom, the operation board, function dial board or MAIN board will be faulty.
3. DIAGNOSTIC CHECK Touch Key Check This item checks the touch key operation and positioning. When you start the check, the screen changes to the touch key check screen, and indicates the position where you touched. When you touch the screen and move your finger or touch pen, the target mark traces your position. If the position of the target mark and the touched position do not match, calibrate the touch screen.
3. DIAGNOSTIC CHECK Sound Check This item checks the 7 types of sounds generated by the instrument. This check can also be used to check the alarm circuit and the QRS beep circuit. Check the quality of sound, loudness, noise, incorrect sound, etc. If there is no sound or distorted sound, the speaker or MAIN board will be faulty. Procedure to Check the Sound 1. To check the sound, touch the “SOUND CHECK” key. 2.
3. DIAGNOSTIC CHECK COM RAM Check This item checks the dual port RAM for communication between the host CPU on the MAIN board and CPU on the DPU board. The check method is the same as the system RAM check. The check program indicates the address, write data and read data if an error is detected. This RAM is located on the DPU board. Therefore if an error is detected, the DPU board or connector which connects the MAIN and DPU boards will be faulty.
3. DIAGNOSTIC CHECK ALARM INDICATOR Check This item checks the function of the alarm indicator located on the top of the instrument. Check the light of indicator lamp according to the screen indication. If the color does not light, the connection cable is loose or disconnected or the MAIN board or indicator board will be faulty. If the indicator lights partially, the indicator board is faulty. Procedure to Start the ALARM INDICATOR Check Program 1.
3. DIAGNOSTIC CHECK OTHER Check This item checks the clock IC and backup NVRAM (EEPROM). The NVRAM contains the system setup data. The check program checks correspondence between write and read data with the stored data retention. Therefore, this check does not damage the system setup data. If the clock IC and NVRAM have no error, the “OK” message appears on the screen and current date and time appear at the top of the screen. If an error message appears, the MAIN board will be faulty.
3. DIAGNOSTIC CHECK MEMORY CARD Check This item checks the inserted PCMCIA card and its interface. There are two checks. The ATTRIBUTE CHECK displays the dump list of attribute area which is in the PCMCIA card. The COMMON CHECK is the write/read check for common area. This check continues until you touch the “STOP” key or “RETURN” key. The check method is the same as the system RAM check. If the PCMCIA card is not inserted, the program displays the CARD NOT INSERTED message on the screen.
3. DIAGNOSTIC CHECK ATA CARD Check This item checks the ATA card (QM-421P) and its interface. This check is almost the same as the MEMORY CARD check, except for the COMMON CHECK. The COMMON CHECK is the write/read check for common area by each sector. This check continues until you touch the “STOP” key or “RETURN” key. The check method is the same as the system RAM check. If an error is detected, the QM-421P or MAIN board will be faulty. Procedure to Start and Exit the ATA CARD Check Program 1.
3. DIAGNOSTIC CHECK Backlight Check This item checks the brightness control of the LCD backlight. While the brightness is manually changed, check that it changes smoothly with no twinkling. Check that the backlight goes off and has no twinkling when the brightness is set to minimum. If the brightness does not change smoothly or the light goes off in the middle of the brightness bar on the screen, it indicates the end of life of the backlight or faulty inverter board.
3. DIAGNOSTIC CHECK CRTC Check Menu Items The CRTC check menu items check the function of the display control block, including the LCD unit. These check results cannot be recognized by the host CPU, so you must judge the result. If an error is found, the MAIN board will be faulty. GRAPHIC MEMORY Check and CHARACTER MEMORY Check These items check the Graphic RAM and Character RAM on the MAIN board for the waveform display or character display by the write/read check technique.
3. DIAGNOSTIC CHECK GRAPHIC DRAW Check This item checks the function of the graphic draw plane. Check the figure of display and colors. At the same time, it checks for defects of LCD pixels all over the screen. The plane for wave/graphic screen is in 20 screens and the check program displays 4 planes at a time. If the screen has a wrong display, the MAIN board or LCD unit will be faulty. Procedure to Execute the GRAPHIC DRAW Check Program 1. To select the Check, touch the “GRAPHIC DRAW CHECK” key.
3. DIAGNOSTIC CHECK Screen Flow of GRAPHIC DRAW Check Program Touch Key Record Touch Key MENU For the TFT LCD screen, it is considered normal if some pixels have randomly abnormal color or do not light. (Up to 25 wrong pixels are considered normal at the maximum brightness.) Service Manual BSM-4100 3.
3. DIAGNOSTIC CHECK CHARACTER DRAW Check This item checks the function of the character draw plane. Check the character display and colors. The plane for character screen is prepared in 4 screens and the check program displays one plane at a time. Procedure to Execute the CHARACTER DRAW Check Program 1. To select the CHARACTER DRAW Check, touch the “CHARACTER DRAW CHECK” key and select a plane. 2. When the program starts, it draws the green characters on the black plane.
3. DIAGNOSTIC CHECK Screen Flow of CHARACTER DRAW Check Program MENU Function Dial Function Dial Function Dial Function Dial Service Manual BSM-4100 3.
3. DIAGNOSTIC CHECK IMAGE DRAW Check This item checks the function of the image draw. All the icons and messages, bitmap images, are displayed one by one. Check the image of the display and background colors. This function helps multi-language capability. Procedure to Start the IMAGE DRAW Check Program 1. To select the IMAGE DRAW Check, touch the “IMAGE DRAW CHECK” key. Touch the “CHANGE BACK COLOR” key to reverse the plane color. 2. When the program starts, it draws the white image on the black plane. 3.
3. DIAGNOSTIC CHECK DPU Check Menu Items The DPU board is a separate subsystem from the MAIN board. The DPU board consists of CPU, ROMs, RAMs, A to D converter and DMA. D/A ADJUST This item is used to check the output voltage level of the D/A converter. Connect a digital multimeter or oscilloscope to the ZB-800P socket (Pin 4: Ground, Pin 5: Analog output) and check the following output signals. D to A output DC 0 V 0 V ±10 mV DC −V CHECK −5.120 V ±1% DC +V CHECK +5.
3. DIAGNOSTIC CHECK AD TEST This is not a check. The AD Test table shows the A to D converted values of all the analog parameters such as ECG, SpO2 and NIBP. The other parameters are shown on the Multi Parameter Unit Check. Each vital sign data are read from the A to D converter every 2 ms. The maximum and minimum data of each vital sign data are displayed every 512 ms.
3. DIAGNOSTIC CHECK AUX I/F Check This check is used to check the waveform signal output when the ZB-800PG/K transmitter is connected to the instrument. The check program outputs a 15 Hz sawtooth waveform to the transmitter. The radio telemetry receiver, such as a cardiac telemetry system, receives the transmitted sawtooth waveform signal and displays it on the screen. Procedure to start the AUX I/F Check 1. To select the AUX I/F Check, touch the “AUX I/F CHECK” key at the DPU CHECK column.
3. DIAGNOSTIC CHECK NIBP Check This item checks the function and safety of NIBP circuit and pump. The NIBP check consist of the following check items: 1) Calibration (1) 2) Calibration (2) 3) Safety Check 4) Inflation Speed 5) Deflation Speed 6) Step Deflation 7) Air leak To perform the test, prepare the following tools: Hand bulb pump Manometer Y-shape hose connector (The Y-shape hose connector connects one tube of the dummy cuff to the handbulb pump and manometer.
3. DIAGNOSTIC CHECK 1. Calibration (1) The Calibration (1) Check program is not used for calibration but is used to display the accuracy of the Pressure Sensor 1. The accuracy of the sensor changes with the measuring pressure. The following table shows the acceptable accuracy range of the Pressure Sensor 1 at different pressure ranges.
3. DIAGNOSTIC CHECK 2. Calibration (2) The Calibration (2) Check program is not used for calibration but is used to display the accuracy of the Pressure Sensor 2. The accuracy of the sensor changes with the measuring pressure. The following table shows the acceptable accuracy range of the Pressure Sensor 2 at different pressure ranges.
3. DIAGNOSTIC CHECK 3. Safety Check There are the following 6 check items monitored by the NIBP safety circuit. Check item Description PRESS 1 Pressure limiter in the adult mode PRESS 2 Pressure limiter in the neonatal mode TIME 1 Time limiter in the adult mode TIME 2 Time limiter in the neonatal mode TIME 3 Interval error in the adult mode TIME 4 Interval error in the neonatal mode Range 315 +/− 15 mmHg 157.5 +/− 7.
3. DIAGNOSTIC CHECK Checking the PRESS 1 Safety Check Item 1. Touch the “SELECT” key to select the item number 3. 2. Touch the “AUTO” key until the key name changes to “PRESS 1” so that the PRESS 1 Safety Check item is selected. 3. Touch the “START” key to start the check. 4. Start pumping the hand bulb pump as soon as the “CHECK” message appears. 5. Quickly increase the pressure until the manometer reads 300 mmHg. From the 300 mmHg pressure point, slowly increase the pressure further.
3. DIAGNOSTIC CHECK Checking the PRESS 2 Safety Check Item 1. Touch the “SELECT” key to select item number 3. 2. Touch the “AUTO” key until the key name changes to “PRESS 2” so that the PRESS 2 Safety Check item is selected. 3. Touch the “START” key to start the check. 4. Start pumping the hand bulb pump as soon as the “CHECK” message appears. 5. Quickly increase the pressure until the manometer reads 150 mmHg. From the 150 mmHg pressure point, slowly increase the pressure further.
3. DIAGNOSTIC CHECK 4. Inflation Speed This is used to check the inflation speed of the pressure pump. The specifications state that the cuff pressure must be able to reach 250 mmHg in less than 7 seconds. Checking the Inflation Speed 1. Connect the 700 ml dummy cuff to the instrument with the 3.5 m air hose as shown below. 2. Touch the “SELECT” key to select item number 4. 3. Touch the “START” key to start the check. If the inflation speed is less than 7 seconds, an “OK” message appears.
3. DIAGNOSTIC CHECK 5. Deflation Speed NOTE Skip this deflation speed check because it is covered in the “6. Step Deflation” check. This is used to check the deflation speed through the two solenoid valves (V1 and V2) at different stages of pressure drop. An “OK” message is displayed if it takes 1.4 to 3.0 seconds for V1 and 2.5 to 4.3 seconds for V2. If not, an “ERROR” message is displayed. Checking the Deflation Speed 1. Connect the 700 ml dummy cuff to the instrument with the 3.
3. DIAGNOSTIC CHECK 6. Step Deflation This checks the step deflation of the cuff at given pressure. HIGH pressure: Valve 1 opens for 40 ms and Valve 2 opens for 70 ms at 250 mmHg. The “OK” message appears if the pressure drop is 5 to 20 mmHg. LOW pressure: Valve 1 opens for 60 ms and Valve 2 opens for 80 ms at 20 mmHg. The “OK” message appears if the pressure drop is 1 to 10 mmHg. Checking the Step Deflation 1. Connect the 250 ml dummy cuff to the instrument with the 3.5 m air hose as shown below. 2.
3. DIAGNOSTIC CHECK 7. Air Leak This checks air leakage in the pneumatic circuit of NIBP module. In this check, the air pressure of the dummy cuff goes up to 300 mmHg. It then compares the pressure readings of the dummy cuff taken at 30 seconds and 90 seconds after the pressure of the dummy cuff had reached 300 mmHg. If the pressure readings differ by 5 mmHg or less, an “OK” message appears. If not, an “ERROR” message appears.
3. DIAGNOSTIC CHECK DPU ROM Check This item checks the ROMs on the DPU board by the CRC (cyclic redundancy check) technique. This check assigns a count number when one set of checks is successfully completed. If an error message appears, the DPU board will be faulty. Procedure to Start and Exit the DPU ROM Check Program 1. To select the DPU ROM Check, touch the “ROM CHECK” key at the DPU CHECK column. 2. The instrument repeats the check and increments the pass count until you touch the “RETURN” key. 3.
3. DIAGNOSTIC CHECK DPU RAM Check This item checks the RAM on the DPU board by comparing the test patterns (first written to the RAM) with the test patterns (later read from the RAM). If an error message appears, the DPU board will be faulty. Procedure to Start, Stop, and Exit the DPU RAM Check Program 1. To select the DPU RAM Check, touch the “RAM CHECK” key at the DPU CHECK column. The check program automatically starts. The “START” key is highlighted to indicate the process of checking. 2.
3. DIAGNOSTIC CHECK DPU COM RAM Check This item checks the dual port RAM for communication between the host CPU on the MAIN board and CPU on the DPU board. The check method is the same as the system RAM check. The check program indicates the address, write data and read data if an error is detected. This RAM is located on the DPU board. Therefore if the error is detected, the DPU board will be faulty. Procedure to Start, Stop, and Exit the DPU COM RAM Check Program 1.
3. DIAGNOSTIC CHECK COM Check Menu Items The COM check menu items check the communication with peripheral units such as recorder and external equipment. WS RECORDER Check This item checks the dual port RAM in the ASIC (RACOON) and the other recorder control block with the communication between the host CPU and local CPU on the MAIN board. The built-in recorder is one of the peripheral units for the host CPU. Physically, both the recorder control block and host CPU are on the MAIN board.
3. DIAGNOSTIC CHECK SERIAL I/F Check This item checks the communication interface for external equipment such as a personal computer. To perform this check without a personal computer, you must make the following short circuits at the AUX socket. Pins 2 and 27 Pins 3 and 28 Pins 4 and 29 Procedure to Start the SERIAL I/F Check Program To select the SERIAL I/F Check, touch the “SERIAL I/F CHECK” key at the COM CHECK column. The check program automatically starts.
3. DIAGNOSTIC CHECK NETWORK CARD Check This item checks the function of the QI-101P network card and its interface. To perform this check, the optional QI-101P network card is required. Procedure to Start the NETWORK CARD Check Program 1. To select the NETWORK CARD Check, touch the “NETWORK CARD CHECK” key at the COM CHECK column. The check program recognizes the type of card and displays the attribute data which is memorized in the card.
3. DIAGNOSTIC CHECK • CHECK: Displays the check result, i.e. “OK” or “ERROR”. When the CALCULATED SUM data is same as the SUM data, “OK” appears. • SUM: The sum data of the program data stored in the ROM • CALCULATED SUM: The sum data calculated from the program data in the ROM 3. Touch the “SYSTEM RAM CHECK” key on the screen.
3. DIAGNOSTIC CHECK 5. Touch the “COMMON RAM CHECK BY HOST” key on the screen. The host CPU on the MAIN board checks the common memory of the network card and displays the check result on the screen. If an error is found, the host CPU displays the address at the error occurrence, written data and read data on the following screen. 6. To return to the MANUAL CHECK MENU screen, touch the “RETURN” key. Service Manual BSM-4100 3.
3. DIAGNOSTIC CHECK MULTI PARAMETER UNIT Check This item checks the communication with the DPU board, function and accuracy of the multi-parameter inputs. The CPU on the MP2 board sends the check result and multi-parameter socket status data to the DPU board. Procedure to Start the MULTI PARAMETER UNIT Check Program To select the MULTI PARAMETER UNIT Check, touch the “MULTI PARAMETER UNIT CHECK” key at the COM CHECK column.
3. DIAGNOSTIC CHECK ECG Check This check is for factory use. NOTE These ECG monitoring settings do not affect the monitor mode. Service Manual BSM-4100 3.
3. DIAGNOSTIC CHECK Recorder and Control Block Checks This check program checks the thermal array recorder and recorder control block on the MAIN board without communication with the host CPU. Therefore, this check is independent from the manual check menu. This check allows you to inspect the quality of print, paper speed and feed, etc. Procedure to start the Recorder and Control Block Check Program 1.
3. DIAGNOSTIC CHECK SELF-TEST-PRO No.2 This check program checks the paper speed, slip and snaking. A 0.5 Hz rectangular waveform and bar graph are recorded on the paper at 25 mm/s paper speed by pressing the NIBP INTERVAL key. To stop the recording, press the NIBP INTERVAL key again. 25 mm/sec 0.5 Hz rectangular waveform 10 mm 25 mm/s +5% − (47.5 to 52.5 mm) Service Manual BSM-4100 3.
3. DIAGNOSTIC CHECK SELF-TEST-PRO No.3 This check program checks the recording quality. There are the following three recording patterns for scale, character and thermal array head checks. The recording pattern changes automatically after the NIBP INTERVAL key is pressed for the recording start. The recording starts soon after the black square on the paper is detected. To stop the recording, press the NIBP INTERVAL key again.
3. DIAGNOSTIC CHECK Calibration of Touch Screen When you replace the touch screen or MAIN board with a new one, the touch screen must be calibrated. Or, depending on the setting position, the effective point may slightly shift due to viewing angle. In that case, re-calibrate after the instrument is set up. Procedure to Start the Calibration of TOUCH PANEL 1. To start the calibration of the TOUCH PANEL, call up the DIAGNOSTIC CHECK screen. 2.
Section 4 Board/Unit Description Block Diagram ..................................................................................................................... 4.1 MAIN Board ........................................................................................................................ 4.2 DPU Board .......................................................................................................................... 4.6 Motherboard .........................................................
Service Manual BSM-4100 Flash ROM 512 k x 16 MC68SEC000 20 MHz Local CPU Remote Buzzer Driver Battery Interface +5 V to LED Driver XTAL 2 MHz System RAM 512 k x 16 SRAM Backup RAM 32 k x 8 EEPROM Touch Screen TFT LCD 800 x 600 Inverter VBUS Touch screen Controller +5 V Function Dial Board Operation Board C RAM 512 k x 8 SRAM CG RAM-2 2 M x 32 SDRAM CG RAM-1 2 M x 32 SDRAM 1700 V AC max Connection Board LVDS Interface RG B Chip select lines Main ASIC ACORN (CPU Peripheral) Interru
4. BOARD/UNIT DESCRIPTION MAIN Board The MAIN board consists of system control block, display control block, recorder control block and power control block. The system control block has the host CPU which gets the operation key status, displays the waveforms and numeric data and controls the recorder unit indirectly.
4. BOARD/UNIT DESCRIPTION The large capacity capacitor (Super capacitor) allows the 1 MB system RAM (SRAM) to back up the trendgraph, vital sign list, arrhythmia recall, monitoring conditions for each parameter and alarm settings for 30 minutes or more after the power is turned off. When the instrument is turned on after more than 30 minutes power off is past, the host CPU regards the backup data as damaged and initializes them.
4. BOARD/UNIT DESCRIPTION b) Display control block The display control block consists of the dedicated graphic ASIC IBIS for entire display control, two 8 MB DRAMs for display of waveform and graphic data, 512 KB high-speed SRAM for display of alphanumeric data and display interfaces for the LCD and an external RGB monitor. When the IBIS receives a display control command from the host CPU, the IBIS accesses these memories to get the display data and arranges the display data for RGB outputs.
4. BOARD/UNIT DESCRIPTION System Control Block Recorder Control Block Local CPU Host CPU ROM RAM I/O RACOON Data flow in normal operation Program data flow in upgrade d) Power control block The power control block consists of the power supply microchip controller, DC/DC converters, buzzer and its driver and battery interface. +15 V from the power supply unit or +12 V from the rechargeable battery is converted to +3.
4. BOARD/UNIT DESCRIPTION DPU Board CPU on the DPU board System ROM System RAM SYLPH I/O ZB-800 I/F Analog data A to D DMA Digital data Common memory (Dual port RAM) Host CPU on the MAIN board The DPU board receives the vital sign information (analog signal or digital data) from the analog boards such as ECG RESP1 board and MP1 board. The CPU on the DPU board executes the primary processing of the vital sign information.
4. BOARD/UNIT DESCRIPTION ZB-800P interface and ECG/BP output circuits are provided on the board. These circuits send out the analog signals with almost no delay. Especially, the ECG/BP output circuit acquires ECG and BP analog signals from the ECG/RESP and multiparameter 1 sockets on the front panel. The ECG analog signal is directly sent to the ECG/BP OUT socket without multiplexer.
4. BOARD/UNIT DESCRIPTION Motherboard The motherboard isolates the analog boards such as ECG RESP1 to 3 boards, SpO2 board and MP1 to 3 boards from the ground. The motherboard supplies the isolated power to each analog board. The motherboard transfers the vital sign signals from the analog boards to the DPU board. The motherboard transfers the control signals from the DPU board to the analog boards. The motherboard lets the two ECG RESP1 to 3 boards communicate.
4. BOARD/UNIT DESCRIPTION ECG RESP1, ECG RESP2 and ECG RESP3 Boards These three boards process the ECG and impedance method respiration signals. There are two ECG processing circuits on the ECG RESP1 board. One is ECG analog signals for arrhythmia analysis and screen display which are selectable with the lead select function. The other one is digital ECG data for 12 lead ECG display and analysis which is digitized by the A to D converter in the CPU on the ECG RESP3 board.
4. BOARD/UNIT DESCRIPTION The ECG RESP2 board has the high-pass filter (for 12 lead ECG), low-pass filter (for 12 lead ECG) and polarization voltage detection circuit. The lead for arrhythmia analysis is selected on this board. The ECG analog signal on the selected lead for arrhythmia analysis goes to the ECG RESP3 board.
4. BOARD/UNIT DESCRIPTION The ECG RESP3 board has the high-pass filter (for the selected lead ECG analog signal at the amplifier final stage), low-pass filter (for the selected lead ECG analog signal at the amplifier final stage) and A to D converter (for 12 lead ECG) in the CPU for primary data processing. There is a supply voltage control circuit to adjust the 12 lead ECG sensitivity. The NIBP cuff size (adult/child or neonate) sensor (hall element) is provided on this board.
4. BOARD/UNIT DESCRIPTION NK SpO2 Board and Nellcor SpO2 Module NK SpO2 Board This board consists of the following blocks. • Differential amplifier block for the electrical signals detected by the photodiode when the red LED and infrared LED in the SpO2 probe alternatively transmits the • • • • • • SpO2 probe Nellcor SpO2 Module 4.
4. BOARD/UNIT DESCRIPTION NIBP Measure Board and NIBP Safety Board The NIBP Measure board has the pressure sensor for NIBP measurement to detect the cuff pressure and convert it to an electrical signal. The NIBP Safety board consists of a microcontroller for safety control, pressure sensor, solenoid valve, pump and power supply cut-off circuit for the valve and pump. This board as well as the NIBP Measure board individually monitors the cuff pressure.
4. BOARD/UNIT DESCRIPTION MP1, MP2 and MP3 Boards The MP1 and MP3 boards have universal input sockets. The MP1 board has two input sockets but the MP3 board has one or no socket depending on the instrument model number.
4. BOARD/UNIT DESCRIPTION The MP2 board acquires the analog signals from the MP1 and MP3 boards and has the preamplifier for CO, temperatures and FiO2 input signals. These analog signals are converted to digital signals on the MP2 board. The communication control circuits for the EEPROM in the connection cord connector and CO2 sensor are also on this board. MP2 MP1 Exc. 1 A to D Conv. to DPU board via motherboard 2 CPU MP3 Exc. 3 Service Manual BSM-4100 4.
4. BOARD/UNIT DESCRIPTION Recorder Unit (for BSM-4101/4102/4103/4104 only) The recorder unit consists of thermal array head, paper drive motor, photo sensor, gears, recorder chassis, sensor board and sensor control board. The RACOON (ASIC) on the MAIN board allows the recorder unit to record the 3 channel waveforms, vital sign list and trendgraph. Since the thick membrane type thermal array head is used, the adjustment of the supply voltage to the head is not required.
4. BOARD/UNIT DESCRIPTION Power Supply Block The power supply block consists of the power supply unit (+15 V DC output), heat sink chassis, fuse holder built-in type AC inlet socket, ground terminal and line frequency detection coil. Refer to Section 8 REPLACEABLE PARTS LIST. This power supply unit same as in the BSM-1101/1102, BSM-2301 and WEP4204/4208 is used. LCD Unit 12.1 inch TFT color LCD which has 800 dots × 600 dots (SVGA) is used as the display. This LCD unit can display maximum 512 colors.
Section 5 Disassembly and Assembly Before You Begin ................................................................................................................. 5.1 Warnings and Caution ................................................................................................ 5.1 Required Tools ........................................................................................................... 5.1 Opening the Instrument ......................................................................
5. DISASSEMBLY AND ASSEMBLY The procedures in this section tell how to remove, replace and install the major components in the instrument. Before You Begin Removing, replacing and installing the major components should be done by qualified service personnel. Warnings and Caution WARNING • To avoid the possibility of injury to yourself or damage to the instrument, do not install or remove any component or change the internal switch settings while the power is on.
5. DISASSEMBLY AND ASSEMBLY Opening the Instrument 1. Turn off the instrument. 2. Disconnect the power cord from the instrument and remove the battery from the instrument. 3. Remove the rubber cap from each end of the handle by popping the cap out with a flat-blade screwdriver. A hexagonal head screw is under each cap. 4. Use a 5 mm Allen wrench to remove the 2 hexagonal head screws which secure the handle. Remove it from the instrument.
5. DISASSEMBLY AND ASSEMBLY 6. Remove the 5 pan head screws from the rear enclosure. Pan head screws Pan head screws 7. Remove the collar head screw with spacer from the bottom of the instrument. Release the latches on the bottom corners of the instrument and slowly and carefully lift the rear enclosure to open the instrument and expose the chassis block. Latches Collar head screw Service Manual BSM-4100 5.
5. DISASSEMBLY AND ASSEMBLY 8. Remove the binding head screw which secures the side panel to the upper chassis (power chassis) cover. Remove the side panel from the instrument. 2 screws are visible under the side panel. NOTE in Assembling the Instrument Check that the binding head screw which was removed in step 8 is used to secure the side panel. The binding head is slightly bigger than a pan head so more torque can be applied to the side panel with the binding head screw. 9.
5. DISASSEMBLY AND ASSEMBLY 10. Carefully lift the upper chassis cover up and disconnect the following cables from the specified sockets. Remove the upper chassis cover. • Disconnect the cable between the power supply unit and MAIN board from CN105 socket on the MAIN board. • Disconnect the cable between the AC inlet socket and DPU board from CNJ031 socket on the DPU board. This cable has a toroidal core for line frequency detection. CAUTION Pull the connector housing to disconnect the cable.
5. DISASSEMBLY AND ASSEMBLY Removing the Nellcor SpO2 Module (MP304*) * When the MP304 SpO2 module is built in, there is no indication of “NELLCOR OxiMax”. Refer to page 1.20 “On the BSM-4102/4104/4112/ 4114 Monitors”. Perform the following procedure after the “Opening the Instrument” procedure. 1. Disconnect the connection cable (MP304 J2) marked with an arrow from the SpO2 module. Open the shield sheet which covers the SpO2 module. 2.
5. DISASSEMBLY AND ASSEMBLY Removing the Nellcor SpO2 Module (MP506*) * When the MP506 SpO2 module is built in, “NELLCOR OxiMax” is indicated below the SpO2 socket on the front panel. Refer to page 1.20 “On the BSM4102/4104/4112/4114 Monitors”. Perform the following procedure after the “Opening the Instrument” procedure. 1. Disconnect the 2 connection cables at the J1 and J2 connectors on the MP506 marked with an arrow. 2. Remove the 3 screws which secure the SpO2 module.
5. DISASSEMBLY AND ASSEMBLY Removing the NIBP Safety Board Perform the following procedure after the “Opening the Instrument” procedure. 1. Disconnect the air tube from the joint on the NIBP Safety board. 2. Disconnect the cable between the NIBP Safety board and DPU board from the CNJ201 socket on the NIBP Safety board. 3. Remove the 4 screws which secure the NIBP Safety board. Remove the board.
5. DISASSEMBLY AND ASSEMBLY CAUTION in Assembling the Instrument When you fasten the NIBP Safety board and analog output cable at the ferrite core with the longer shaft screw, do not leave the analog output cable on the three ICs (U802, U096 and U097) of the DPU board. If the analog output cable is long and it has been left on the three ICs, bind the cable with a cable binder as shown below so that the cable is not left on the three ICs of the DPU board.
5. DISASSEMBLY AND ASSEMBLY Removing the Motherboard, NIBP Measure Board Perform the following procedure after the “Removing the NIBP Safety Board” procedure. 1. Remove the 2 screws and 2 spacer bolts on the motherboard holder as shown below. Slightly tilt the upper part (marked with a thick arrow in the picture) of the motherboard holder and slowly remove it and its insulation sheet to expose the motherboard.
5. DISASSEMBLY AND ASSEMBLY 3. Disconnect the air tube from the NIBP Measure board and remove the screw which secures the NIBP Measure board to the DPU board. Remove the NIBP Measure board from the DPU board.
5. DISASSEMBLY AND ASSEMBLY Removing the DPU Board Perform the following procedure after the “Removing the Motherboard, NIBP Measure Board” procedure. 1. Remove the 4 screws and 2 spacer bolts from the DPU board as shown below. 2. Disconnect the analog output cable between the DPU board and ECG RESP2 board from the CNJ013 socket on the DPU board. 3. Remove the DPU board from the DPU board holder.
5. DISASSEMBLY AND ASSEMBLY Removing the MAIN Board Perform the following procedure after the “Removing the DPU Board” procedure. 1. Remove the 2 small screws (PS2.6 x 8) which secure the AUX socket of the MAIN board to the DPU board holder. Remove the 4 screws which secure the DPU board holder and remove it. 2. Disconnect the 6 cables from the CN103, CN108, CN006 and CN121 to CN123 sockets on the MAIN board. Remove the 4 screws which secure the MAIN board and remove it.
5. DISASSEMBLY AND ASSEMBLY Removing the Analog Boards Perform the following procedure after the “Removing the Motherboard, NIBP Measure Board” procedure step 2. The analog boards are coupled as shown below. • ECG RESP1 and ECG RESP2 boards • ECG RESP3 and SpO2 boards • MP1 and MP2 boards • MP3 and MP4 boards 1. Remove the coupled boards from the lower chassis. 2. Remove the pan head screw which combines the two boards as shown below. 3. Separate the two boards. 5.
5. DISASSEMBLY AND ASSEMBLY Removing the Recorder Unit Perform the following procedure after the “Opening the Instrument” procedure. 1. Disconnect the 3 cables between the recorder unit and MAIN board from the CN121 to CN123 sockets on the MAIN board. 2. Remove the 2 screws marked with arrows in the above picture. 3. Place the chassis block face up. Pull the paper magazine of the recorder unit upward and remove it.
5. DISASSEMBLY AND ASSEMBLY Replacing the Touch Screen Perform the following procedure after the “Opening the Instrument” procedure step 8. 1. Remove the 4 screws which secure the main chassis to the upper front enclosure. 2. Disconnect the touch screen film cable from the connection board. CAUTION To disconnect the film cable, pull up the cable stopper of the socket on the MAIN board so that the film cable is set free. 5.
5. DISASSEMBLY AND ASSEMBLY 3. Disconnect the cable connecting to the Operation board, Power SW board and alarm indicator board from the MAIN board. 4. Separate the upper front enclosure with the touch screen from the chassis block by slowly setting the chassis block upright. CAUTION • Since the touch screen is covered with glass, do not apply excessive mechanical shock such as dropping or pressure to the touch screen. • Pay attention to the sharp glass parts such as the touch screen edges.
5. DISASSEMBLY AND ASSEMBLY 5. Replace the touch screen and the two long and short packings with new ones. Refer to Section 8 Replaceable Parts List. CAUTION in Assembling the Instrument Fit the touch screen onto the original position of the front enclosure. Otherwise, the touch screen will be broken when the chassis block is attached to the front enclosure. 6. Assemble the instrument by reversing the above procedure. 5.
5. DISASSEMBLY AND ASSEMBLY Replacing the Backlight Lamps Perform the following procedure after the “Replacing the Touch Screen” procedure step 4. 1. Disconnect the cable connecting to the backlight lamps of the LCD unit from the inverter board as shown below. CAUTION Be careful not to strain the cable. 2. Disconnect the cable connecting to the LCD unit from the Connection board as shown below. Service Manual BSM-4100 5.
5. DISASSEMBLY AND ASSEMBLY 3. Place the chassis block face up on the table. Remove the 4 screws from the LCD unit as shown below and separate the LCD unit from the chassis block. CAUTION • Since the LCD unit is covered with glass, do not apply a mechanical shock such as dropping or pressure to the LCD unit. • Before placing the LCD unit face down on the table, check that the table is covered with a clean, soft and smooth material to avoid making the screen dirty or damaging the screen. 4.
5. DISASSEMBLY AND ASSEMBLY • Carefully handle the edges of the backlight lamp and cover because their edges are sharp and can cause cuts and injuries. • Discard used backlight lamps (fluorescent tubes) according to your local laws. For monitors with revision number before CA i) Remove the 2 small screws which secure the backlight lamp cover. Slide it outward. 2 backlight lamps are in the holder of the LCD unit. Backlight lamp cover 2 small screws ii) Replace the 2 backlight lamps with new ones.
5. DISASSEMBLY AND ASSEMBLY ii) Pull out the backlight lamp by pulling the backlight cable in the direction of the arrow. Replace the 2 backlight lamps with new ones. 7. Carefully assemble the instrument by reversing the above procedure. 5.
5. DISASSEMBLY AND ASSEMBLY Installing the Options Installing QM-421P Flash Disk Card into the Slot on the MAIN Board Perform the following procedure after the “Opening the Instrument” procedure. 1. Remove the 2 screws (marked with arrow 2 in the following picture) which secure the inverter board. 2. Pull out the inverter board and put it beside the alarm indicator. 3. Remove the screw (marked with arrow 4 in the step 1 picture) which secure the MAIN board. This screw prevents disconnection of the card.
5. DISASSEMBLY AND ASSEMBLY 4. Insert the card into the card slot on the MAIN board. CAUTION The slot is hard to see. To correctly install the card into the slot, keep the card parallel to the slot while slowly inserting it. Otherwise, pin(s) of the slot may be bent or broken. DPU board holder QM-421P Slot MAIN board 5. Carefully assemble the instrument by reversing the above procedure. 6. Perform the ATA CARD check in the manual check to check the card function.
5. DISASSEMBLY AND ASSEMBLY Service Manual BSM-4100 5.
5. DISASSEMBLY AND ASSEMBLY Replacing the Lithium Battery Perform the following procedure after the “Removing the DPU Board” procedure. CAUTION • Use a glove to handle the new lithium battery. If you touch it with your bare hand, it may cause an increase of resistance at the battery terminals and shorten the lifetime. • Never charge, short-circuit, disassemble, deform, heat, or throw the battery into fire. This may cause overheating, explosion, or fire.
Section 6 Maintenance To Be Replaced Periodically ................................................................................................ 6.1 Required Tools ..................................................................................................................... 6.1 Measuring and Test Equipment ........................................................................................... 6.2 YS-073P8 Board/Unit Maintenance Kit ..............................................................
6. MAINTENANCE To Be Replaced Periodically We recommend the periodic replacement of the following components according to the expected life span. NK Code No.
6. MAINTENANCE Measuring and Test Equipment To repair, check, or adjust the instrument, the following measuring and test equipment or equipment with similar function and performance is required. Digital Voltmeter A digital tester with at least 3-1/2 digits LCD display that can measure voltage, current, and resistance. Cathode-ray Oscilloscope An oscilloscope with a bandwidth of more than 1 MHz and a sensitivity of more than 10 mV/cm.
6. MAINTENANCE YS-073P8 Board/Unit Maintenance Kit Composition The kit consists of the following. • Extension boards: 11 pcs. • Extension cables: 11 pcs. • 80-pin connectors: 2 pcs. Connection Refer to Connection Diagram as shown below. • Extension boards DEBUG1: Connects the MAIN board and DPU board. The two 80-pin connectors of this kit let you align the DPU board either vertical or horizontal to the MAIN board. The desired connector (vertical or horizontal) must be soldered on the DEBUG1 board.
6. MAINTENANCE EXT8: CNA9 cable. The EXT7 board has one connector each for the analog board and cable. Connects the motherboard to the EXT7 board with the CNA9 cable. The EXT8 board has one connector each for the motherboard and cable. • Extension cables CNA1 (50 cm long): Extends the distance between the MAIN and Inverter boards. CNA2 (30 cm long): Extends the distance between the DEBUG2 and DEBUG3 boards. The cable is used for touch screen signals.
Service Manual BSM-4100 CNA5 EXT3 ECG RESP1 MP1 EXT5 EXT6 EXT8 ECG RESP2 ECG RESP3 CNA9 EXT4 Motherboard CNA4 DPU DPU Speaker EXT7 SpO2 MP2 MP3 CNA8 MAIN Recorder Unit EXT2 CNA7 CNA10 DEBUG1 DEBUG1' AC/DC CNA6 DEBUG2 CNA2 CNA11 CNA3 DEBUG3 Inverter Power SW Connection CNA1 EXT1 Operation Alarm Indicator 6. MAINTENANCE Connection Diagram 6.
6. MAINTENANCE Maintenance Check Items and Schedule Perform this maintenance check once every six months. A maintenance check sheet is provided at the end of this section. Make a copy of this check sheet before using it. The check items are grouped as follows: • External • Input conditions • Operation • Display • Recorder • Vital sign parameters • Power • Data backup • Safety • Others Following are the procedures for each check item.
6. MAINTENANCE Operation Item Key function on the operation panel Touch screen key function Check Procedure Check that the function of each key on the control panel works correctly. Check that the function of each key on the screen works correctly. Action Remove the cause if the key function is wrong. Remove the cause if the key function is wrong. Display Item Crack, scratch or dirt Check Procedure Check that there is no crack, scratch or dirt on the touch screen.
6. MAINTENANCE Vital Sign Parameters Item Waveform and numeric data SpO2 and NIBP NIBP check Multi-parameter socket function Zero balance of IBP Alarm function Transducer Check Procedure Check that ECG, respiration, blood pressure waveforms, their parameters and temperature data are displayed correctly and QRS synchronous sound is generated when connecting AX-800P vital sign simulator to the instrument.
6. MAINTENANCE Data Backup Item System Setup data backup Clock function backup Check Procedure Check that the System Setup data is saved for a long time after the power is turned off. Check that the clock function works correctly when the power is turned on. Action If the System Setup data disappears within 30 minutes after the power off, replace the super capacitor on the MAIN board with a new one.
6. MAINTENANCE Others Item Line voltage Connection to other equipment Vibration or resonant sound Screws Option check Installation condition check 6.10 Check Procedure Check that the line voltage is within the range of nominal voltage ±10%. Check that the instrument is correctly connected to other equipment according to the operator’s manual. Check that the instrument has no abnormal vibration or resonant sound. Check that there are no loose screws.
6. MAINTENANCE Maintenance Check Sheet (Refer to the Maintenance section of the service manual for details.
6. MAINTENANCE Power Power cord (and ground lead) are not damaged. Instrument is firmly grounded to a dedicated grounding terminal. Fuse is the correct rating. Yes Yes Yes No No No Safety Protective earth impedance is less than prescribed limit. Earth leakage current is less than prescribed limits. Enclosure leakage current is less than prescribed limits. Patient leakage current is less than prescribed limits. Withstand voltages of instrument are prescribed voltage and time limits or more.
Section 7 Adjustment Sensors in the Recorder Unit .............................................................................................. 7.1 Adjusting the Output Voltages with Digital or Analog Multimeter ............................... 7.1 Adjusting the Output Voltages with the YS-073P8 Board/Unit Maintenance Kit and Digital or Analog Multimeter ................................................................................. 7.
7. ADJUSTMENT Sensors in the Recorder Unit After you replace the sensor board of the recorder unit with a new one, adjust the paper and mark detection output voltages on the sensor control board as shown below. Adjusting the Output Voltages with Digital or Analog Multimeter 1. Connect 150 Ω resistors to pins 9 and 11 of the 12-pin connector cable from the sensor control board as shown below. 2.
7. ADJUSTMENT Adjusting the Output Voltages with the YS-073P8 Board/Unit Maintenance Kit and Digital or Analog Multimeter 1. Connect the EXT2 board of the kit to the recorder unit. Refer to “YS-073P8 Board/Unit Maintenance Kit” in Section 6. 2. Short pins 1 and 2 at the jumper setting terminal on the EXT2 board. 3. Connect +5 V DC of a +5 V DC power supply to the +5 V input terminal on the EXT2 board. 4. Connect the ground terminal of the +5 V DC power supply to the ground terminal on the EXT2 board. 5.
7. ADJUSTMENT 650 µs 3.5 to 4.0 V 0V 16 ms 8. Adjust the PAPER VR and MARK VR so that the PAPER and MARK test pins on the EXT2 board output the rectangular amplitude within the range of 3.5 V DC to 4.0 V DC when the two sensors detect the white part of the paper. Service Manual BSM-4100 7.
Section 8 Replaceable Parts List Bedside Monitor BSM-4101/4103/4111/4113 ....................................................................... 8.2 Bedside Monitor BSM-4102/4104/4112/4114 (MP304* included) ....................................... 8.10 Bedside Monitor BSM-4102/4104/4112/4114 (MP506* included) ....................................... 8.18 RG-921P Recorder Unit ..................................................................................................... 8.
8. REPLACEABLE PARTS LIST When ordering parts or accessories from your nearest Nihon Kohden Corporation distributor, please quote the NK code number and part name which are listed in this service manual, and the name or model of the unit in which the required part is located. This will help us to promptly attend to your needs. Always use Nihon Kohden parts and accessories to assure maximum performance from your instrument. Service Manual BSM-4100 8.
8. REPLACEABLE PARTS LIST Bedside Monitor BSM-4101/4103/4111/4113 Index NK Code No. Qty Description 1 289848 2 Hexagon socket head bolt 2 515542B 1 NIBP socket 3 531346 1 NIBP air filter 4 577288 1 Collar head screw with spacer 5 579375 1 Washer 6 1114-139714 6 Spacer bolt (7.2 mm long) 7 1114-164116 2 Spacer bolt (27.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 80 34 2 7 z 79 53 3 6 WEP -420 4 w x w 34 81 T M Y N U b a 36 45 26 78 56 50 60 52 c 85 6 W Q 15 34 34 41 76 G x V 72 R 64 23 H a y J I b a 4111 4113 a 4103 b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4101 65 71 74 P 54 70 O 73 65 75 69 63 45 y 18 25 59 48
8. REPLACEABLE PARTS LIST Index 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 *1 *2 *3 *4 *5 *6 *7 NK Code No.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 80 34 2 7 z 79 53 3 6 WEP -420 4 w x w 34 81 T M Y N U b a 36 45 26 78 56 50 60 52 c 85 6 W Q 15 34 34 41 76 G x V 72 R 64 23 H a y J I b a 4111 4113 a 4103 b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4101 65 71 74 P 54 70 O 73 65 75 69 63 45 y 18 25 59 48
8. REPLACEABLE PARTS LIST Index 73 74 75 76 77 78-a 78-b 79 80 81 NK Code No.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 80 34 2 7 z 79 53 3 6 WEP -420 4 w x w 34 81 T M Y N U b a 36 45 26 78 56 50 60 52 c 85 6 W Q 15 34 34 41 76 G x V 72 R 64 23 H a y J I b a 4111 4113 a 4103 b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4101 65 71 74 P 54 70 O 73 65 75 69 63 45 y 18 25 59 48
8. REPLACEABLE PARTS LIST Index Q R S T U V W X Y NK Code No. Qty Description UR-3559 1 MP3 board for BSM-4103/4113 UR-35591 1 MP3 dummy board for BSM-4101/4111 6190-020774C 1 MP4 board UR-3563 1 DPU board UR-3564 1 NIBP Measure board UR-3566* 1 Motherboard UR-3567 1 NIBP Safety board 532149 1 NIBP pump SC-039R 1 Power supply block Consists of the following components.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 80 34 2 7 z 79 53 3 6 WEP -420 4 w x w 34 81 T M Y N U b a 36 45 26 78 56 50 60 52 c 85 6 W Q 15 34 34 41 76 G x V 72 R 64 23 H a y J I b a 4111 4113 a 4103 b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4101 65 71 74 P 54 70 O 73 65 75 69 63 45 y 18 25 59 48
8. REPLACEABLE PARTS LIST Bedside Monitor BSM-4102/4104/4112/4114 (MP304* included) * When the MP304 is built in the instrument, there is no indication of “NELLCOR OxiMax”. Refer to page 1.20 “On the BSM-4102/ 4104/4112/4114 Monitors”. Index NK Code No. Qty Description 1 289848 2 Hexagon socket head bolt 2 515542B 1 NIBP socket 3 531346 1 NIBP air filter 4 577288 1 Collar head screw with spacer 5 579375 1 Washer 6 1114-139714 7 Spacer bolt (7.2 mm long) 7 1114-164116 2 Spacer bolt (27.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 82 80 34 2 7 z 53 3 T w β x w WEP -420 4 81 Z 6 34 M Y β N U b a 36 45 26 78 56 50 60 52 c 85 34 Q 41 76 15 69 34 G 6 W 79 x V 72 R 64 23 H a y J 4114 4112 4104 I a b a b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4102 65 71 74 P 54 70 O 73 65 75 α 63 45
8. REPLACEABLE PARTS LIST Index 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 *1 *2 *3 *4 *5 *6 *7 NK Code No.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 82 80 34 2 7 z 53 3 T w β x w WEP -420 4 81 Z 6 34 M Y β N U b a 36 45 26 78 56 50 60 52 c 85 34 Q 41 76 15 69 34 G 6 W 79 x V 72 R 64 23 H a y J 4114 4112 4104 I a b a b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4102 65 71 74 P 54 70 O 73 65 75 α 63 45
8. REPLACEABLE PARTS LIST Index 73 74 75 76 77 78-a 78-b 79 80 81 82 85 86 A B C D E F or G H I J K L M N O P Q R S T NK Code No.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 82 80 34 2 7 z 53 3 T w β x w WEP -420 4 81 Z 6 34 M Y β N U b a 36 45 26 78 56 50 60 52 c 85 34 Q 41 76 15 69 34 G 6 W 79 x V 72 R 64 23 H a y J 4114 4112 4104 I a b a b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4102 65 71 74 P 54 70 O 73 65 75 α 63 45
8. REPLACEABLE PARTS LIST Index U V W X Y Z α β NK Code No. Qty Description 1 UR-35661* 1 Motherboard UR-3567 1 NIBP Safety board 532149 1 NIBP pump SC-039R 1 Power supply block Consists of the following components.
Service Manual BSM-4100 77 61 49 35 C 57 37 39 32 10 47 63 A a 86 55 D 20 12 33 E S B 68 b 21 43 42 L 66 40 46 44 67 K F 82 80 34 2 7 z T 53 3 WEP 81 Z 6 34 M -420 4 w β x w Y β N U b a 36 45 26 78 56 50 60 52 c 85 34 Q 41 76 15 69 34 G 6 W 79 x V 72 R 64 23 H a y J 4114 4112 4104 I a b a b 78 z Y 11 b NO YES NO YES YES NO YES NO 80 31 29 30 62 BSM-4102 65 71 74 P 54 70 O 73 65 75 α 63 4
8. REPLACEABLE PARTS LIST Bedside Monitor BSM-4102/4104/4112/4114 (MP506* included) * When the MP506 is built in the instrument, “NELLCOR OxiMax” is indicated below the SpO2 socket on the front panel. Refer to page 1.20 “On the BSM-4102/4104/4112/4114 Monitors”. Index NK Code No. Qty Description 1 289848 2 Hexagon socket head bolt 2 515542B 1 NIBP socket 3 531346 1 NIBP air filter 4 577288 1 Collar head screw with spacer 5 579375 1 Washer 6 1114-139714 5 Spacer bolt (7.
Service Manual BSM-4100 86 55 85 8. REPLACEABLE PARTS LIST 8.
8. REPLACEABLE PARTS LIST Index 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 *1 *2 *3 *4 *5 *6 *7 NK Code No.
Service Manual BSM-4100 86 55 85 8. REPLACEABLE PARTS LIST 8.
8. REPLACEABLE PARTS LIST Index 73 74 75 76 77 78-a 78-b 79 NK Code No.
Service Manual BSM-4100 86 55 85 8. REPLACEABLE PARTS LIST 8.
8. REPLACEABLE PARTS LIST Index V W X Y Z α β NK Code No. Qty Description UR-3567 1 NIBP Safety board 532149 1 NIBP pump SC-039R 1 Power supply block Consists of the following components.
Service Manual BSM-4100 86 55 85 8. REPLACEABLE PARTS LIST 8.
8. REPLACEABLE PARTS LIST RG-921P Recorder Unit Index 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 NK Code No. 6111-002327E 6113-017726C 6114-067901A 6114-053132A 515284 6114-053141B 6114-053177C 515506A 6114-053168C 6114-053159B 6112-008444C 6114-053114C 492503 6114-059474B 6114-053186A 6113-023951A 107002 A B C D 445074 445109B 448053 UR-3570 578447 525656A 449132A UR-3571 E F G 8.
8. REPLACEABLE PARTS LIST Service Manual BSM-4100 8.
Section 9 Connector Pin Assignment MAIN Board ........................................................................................................................ 9.1 CN006 (for Speaker) ........................................................................................ 9.1 CN101 (for DPU board) .................................................................................... 9.2 CN102 (for Connection board) ......................................................................... 9.
CN101 (for Function Dial board) .................................................................... 9.19 CN102 (for MAIN board) ................................................................................ 9.19 Alarm Indicator Board ........................................................................................................ 9.20 CN1 (for MAIN board) .................................................................................... 9.20 Power SW Board .......................................
9. CONNECTOR PIN ASSIGNMENT MAIN Board CN006 (for Speaker) Pin Number Signal Name Signal Description 1 SND+ Speaker positive 2 SND− Speaker negative (Ground) 3 NC Not connected 4 NC Not connected Service Manual BSM-4100 9.
9.
9.
9.
9.
9.
9. CONNECTOR PIN ASSIGNMENT CN108 (for Inverter board) Pin Number Signal Name Signal Description 1 +15V +15 V supply voltage for inverter 2 +15V +15 V supply voltage for inverter 3 E2 Digital ground 4 E2 Digital ground 5 BCNT Backlight ON/OFF 6 BCNT_PWM Brightness control signal 7 NC Not connected 8 NC Not connected 9 NC Not connected 10 NC Not connected Service Manual BSM-4100 9.
9.
9.
9. CONNECTOR PIN ASSIGNMENT CN122 (for RG-921P recorder unit) Pin Number Signal Name Signal Description 1 MA1 Motor phase control A 2 MB1 Motor phase control B 3 MA2 Motor phase control A 4 MB2 Motor phase control B 5 +3.3V +3.
9. CONNECTOR PIN ASSIGNMENT DPU Board CN701 (for MAIN board) Refer to “CN101 (for DPU board)” in the “MAIN Board” section. CNJ022 (for Motherboard) Signal Description Signal Name Pin Number Signal Name Signal Description Address bus bit 2 CLK A1 B1 VBUS +15 V supply voltage Analog ground AGND A2 B2 VBUS +15 V supply voltage SpO2 digitized data SPO2 A3 B3 VBUS +15 V supply voltage +3.3 V supply voltage +3.
9. CONNECTOR PIN ASSIGNMENT CNJ031 (for power supply unit) Pin Number Signal Name Signal Description 1 AC_CLK Line frequency detection (50/60 Hz) 2 AGND Analog ground CNJ041 (for NIBP Measure board) Pin Number 9.
9.
9. CONNECTOR PIN ASSIGNMENT Motherboard CNJ701 (for DPU board) Refer to “CNJ022 (for Motherboard)” in the “DPU Board” section.
9.
9.
9.
9. CONNECTOR PIN ASSIGNMENT CNJ501 (for Nellcor SpO2 board) BSM-4102/4104/4112/4114 only Signal Description Signal Name Pin Number Signal Name Signal Description Not connected N.C. 1 2 EF Floating ground Floating ground EF 3 4 RESET Reset output Floating ground EF 5 6 N.C. Not connected −5 V for analog circuit −5F 7 8 NEL_TX MP304/MP506 serial data (RXD) MP304/MP506 serial data (TXD) NEL_RX 9 10 +5F +5 V for analog circuit CTS output to MP506 N.C.
9.
9. CONNECTOR PIN ASSIGNMENT Alarm Indicator Board CN1 (for MAIN board) Pin Number Signal Name Signal Description 1 +15V +15 V supply voltage 2 XRED Red LED cluster control 3 XYELLOW Yellow LED cluster control 4 XGREEN Green LED cluster control Power SW Board CN101 (for MAIN board) Pin Number 9.
9. CONNECTOR PIN ASSIGNMENT Connection Board CN101 (for LCD unit) Pin Number Signal Name Signal Description 1 +3.3V +3.3 V supply voltage 2 +3.3V +3.
9.
9.
Section 10 Internal Switch Setting and Board Compatibility Internal Switch Setting ..................................................................................................... 10.1 MAIN Board .................................................................................................. 10.1 DPU Board .................................................................................................... 10.1 Board Compatibility .......................................................................
10. INTERNAL SWITCH SETTING AND BOARD COMPATIBILITY Internal Switch Setting MAIN Board 1 When you replace the MAIN board with a new one, set the bit switches of DIP switch SW001 on the new MAIN board. 2 3 4 ON OFF SW001 DPU Board 1 Bit 1: ON: for BSM-5100 series OFF: for BSM-4100 series Bit 2: Reserved. Set to OFF. Bit 3: ON: for BSM-4100 series without recorder OFF: for BSM-4100 series with recorder or BSM-5100 series Bit 4: Reserved. Set to OFF.
10. INTERNAL SWITCH SETTING AND BOARD COMPATIBILITY Board Compatibility MAIN Board Since the MAIN board has had a design change to be commonly used between the BSM-4100 series and BSM-5100 series, specify the appropriate hardware revision (indicated on the label on the board) when you issue an order for a replacement MAIN board. If there is a MAIN board in your stock, check the hardware revision to determine if it is for BSM-4100 series only or for both BSM-5100 and BSM4100 series.