User`s manual
Chapter 3 Function
3-46
3
Reset
The function resets the measurement value.
At the time of making a peak measurement, valley measurement, or peak-to-peak
measurement with external multi input (MI), the measurement value kept on hold
at the desired timing can be reset.
Input a specified pulse width as multi input.
The pulse width can be input into the high-function type through a communications
command.
For details,
refer to "2-4 Timing Chart".
All judgment outputs will be turned OFF when
“
Reset
”
is executed. This
function can be applied to judge each measurement object for peak, valley, o
r
peak-to-peak measurements.
•
If “Reset” is executed, the system will be in a data unfixed status, analog output
will be set to a preset fix value (21.6 [mA] and +11.000 [V]), and all the
judgment outputs will be OFF.
•
If Reset ON is executed by a communications command, the memory contents
will be cleared to zero after Reset is executed.
•
The sensor head does not save setting changes made over the multi-input line.
Save the settings over the multi-input line, through the panel, with the
communications command, or with the operation of the console if the settings
need to be maintained after the system is turned OFF.
● Setting Procedure
1
Press the [UP]/[ENTER] keys together
while in Run mode.
Reset will be executed.