User`s manual
Chapter 4 Communications Control
4-16
4
● Error code
Error
code
Code type Description
01
Command error
The command is undefined.
02
Address error
The start address is larger than the end address or the
address is larger than 999999 when the RDD or WDD
command is executed.
The address length has not reached the prescribed
length when the RDD or WDD command is executed.
03
Data error
The data length does not correspond to the command.
The data length has not reached the prescribed length.
04
BCC error
BCC check was not conformable.
11
Communication
error
A parity error occurred during data reception.
A framing error occurred during data reception.
An overrun error occurred during data reception.
21
Control flow error The system is in setting mode.
22
Execution error
Calibration or analog scaling is not executable.
31
Buffering
condition error 1
An attempt was made to make a buffering setting
change without stopping buffering.
32
Buffering
condition error 2
An attempt was made to start buffering with an
inadequate buffering setting.
33
Buffering
condition error 3
Data was read after buffering operation started.
Data was read while the system was not in the
accumulation completed status.
Data in excess of the final data point was specified and
read.
• If an abnormal response is received, check if there are any errors such as
transmission data errors. Correct such errors, if any. If an error occurs
regardless of the absence of mistakes, it may be a temporal phenomenon
caused by noise or other external reasons or the external host device may
have a system error. Turn the sensor head or system turned OFF and ON.