User`s manual
Table Of Contents
- CONTENTS
- 1 INTRODUCTION
- 2 THE CLIO SYSTEM
- 3 CLIO INSTALLATION
- 3.1 MINIMUM PC CONFIGURATION
- 3.2 FW-01 DRIVERS INSTALLATION UNDER WINDOWS XP
- 3.3 FW-01 DRIVERS INSTALLATION UNDER WINDOWS VISTA AND 7
- 3.4 SOFTWARE INSTALLATION
- 3.5 THE 'CLIO BOX'
- 3.6 RUNNING CLIO FOR THE FIRST TIME
- 3.7 SYSTEM CALIBRATION
- 3.8 CLIO SERIAL NUMBER AND DEMO MODE
- 3.9 TROUBLESHOOTING CLIO INSTALLATION
- 4 CLIO BASICS
- 5 SYSTEM OPERATIONS AND SETTINGS
- 6 COMMON MEASUREMENT INTERFACE
- 7 SIGNAL GENERATOR
- 8 MULTI-METER
- 9 FFT, RTA AND “LIVE” TRANSFER FUNCTION
- 10 MLS & LOG CHIRP
- 11 SINUSOIDAL
- 12 WATERFALL , DIRECTIVITY & 3D
- 12.1 INTRODUCTION
- 12.2 WATERFALL, DIRECTIVITY & 3D CONTROL PANEL
- 12.3 WATERFALL SPECIFIC CONTROLS
- 12.4 MAKING A CUMULATIVE SPECTRAL DECAY
- 12.5 DIRECTIVITY SPECIFIC CONTROLS
- 12.6 MEASURING LOUDSPEAKER SINGLE POLAR DATA (1D MODE)
- 12.7 REPRESENTING SINGLE POLAR DATA
- 12.8 3D SPECIFIC CONTROLS
- 12.9 MEASURING LOUDSPEAKER SINGLE POLAR DATA (3D MODE)
- 12.10 MEASURING FULL SPHERE LOUDSPEAKER POLAR DATA (3D MODE)
- 12.11 REPRESENTING 3D BALLOON DATA
- 12.12 EXPORT 3D BALLOON DATA
- 13 MEASURING IMPEDANCE AND T&S PARAMETERS
- 14 LINEARITY & DISTORTION
- 15 ACOUSTICAL PARAMETERS
- 16 Leq LEVEL ANALYSIS
- 17 WOW AND FLUTTER
- 18 WAVELET ANALYSIS
- BIBLIOGRAPHY
- NORMS

5.5 DESKTOP MANAGEMENT
Desktop management is a powerful feature that lets you save your work at a
certain point and reload exactly as it was.
It is possible to do this automatically when exiting CLIO; at successive startup the
program will automatically reload from where you left; to do this activate the Save
measurement session from CLIO General Options.
Load a previously saved measurement session (*.sna files).
Takes a snapshot of current measurement session and saves it to disk (*.sna
files).
If pressed clears current measurement desktop i.e. closes and resets all
measurement menu. It is also possible to clear one single measurement
selectively opening and choosing from the associated drop down menu.
5.6 CALIBRATION
File > Calibration
This option will perform a calibration of your CLIO hardware. Please refer to
paragraph 3.7 and follow the procedure described.
In order to determine, at any given time, if it is necessary to calibrate CLIO do
the following:
- Let the system warm up
- Proceed to perform the verification described in 3.7.1
- Consequently decide whether or not to calibrate
The result of the measurement may vary in some way from the time we
calibrated because of many small changes in measurement conditions, including
changes in the atmospheric conditions, the season, and the mains voltage.
Note: the CLIO hardware is highly precise and stable and, under normal
operating conditions, does not require frequent calibrations.
Always perform a calibration if:
- CLIO asks for it showing the “System Not Calibrated” message
- You reinstalled CLIO in a different computer
- You installed a software upgrade