User`s manual

Table Of Contents
3.7 SYSTEM CALIBRATION
This section describes how to perform the system calibration.
Be sure that, any time you perform a calibration, the system has warmed up for, at
least 15-20 minutes.
Select Calibration from the File menu (5.6);
Leave the CLIO Box front plugs unconnected.
Answer OK to the initial prompt; this will run an automatic procedure that will last
several minutes. The calibration procedure is completely automatic and several
progress indicators will accompany all the executed measurements. At the end of it
your CLIO system should be calibrated and ready to carry out measurements.
At the end of the calibration process it is always mandatory to verify the calibration
itself; this is done by two simple measurements as described in the following
section.
3.7.1 CALIBRATION VALIDATION
Figure 3.29
32 3 CLIO INSTALLATION