Specifications

If there is a volume control on the amplifier it should have been already set
counter-clockwise (to minimum). Therefore leave the CLIO Output level at its
default 1.000V. In absence of a volume control on the amplifier set CLIO Output
level at 0.01V or below. Open the Generator Control Panel and generate Pink Noise
by clicking on Pink and then on Play. You can ignore the Size for now.
Either with the amplifier's volume control, or with the CLIO output level control
arrows, increase the level(you should hear a sound like a waterfall, which is Pink
Noise). Adjust the output to read around 80dBSPL on the Level Meter. Turn the
Generator Off. Click on the LogChirp button to invoke the LogChirp panel and
press the Frequency & Time Graph button so that you have Frequency and Time
Domains visible at the same time. Press the Go button to execute a LogChirp
Frequency Response measurement with exactly the same settings shown for the
Nearfield measurement.
44 4 ENJOY CLIO POCKET CLIO POCKET - SURVIVAL GUIDE