Specifications
2.5 INITIAL TEST MEASUREMENT
Let's now execute our first test measurement: to play and capture a 1kHz sinusoid.
The In-Out Loop button should be down, click on it if not; in this way CLIO
connects output with input with an internal relay. This connection is very important
as it lets you capture and analyze a signal generated by CLIO without the need for
an external connecting cable.
Then click on the generator icon to play a 1kHz sinusoid, it's the default signal.
The input meter should react as in Figure and display the level in Volts of the signal
at the input.
If everything is OK you should obtain a reading of circa 1V, variable between a
minimum of 0.95V and a maximum of 1.05V, which is the mean output level of a
sinusoidal signal when the system is not calibrated.
Now press the FFT button, the Frequency and Time Graphs button and finally
the Go Button.
2 CLIO POCKET MAP 27