User's Manual
User Manual
Document No.
UM-C-002-EN
Date of Revision
2020.08.30
Revision No.
0
QF-001
ATsens Co., Ltd.
69/123
Figure 10.78 Save function in Class tab
2) Date Drop box
3) The current working date information of the Class Tab is displayed in the Data Drop Box. Click (▼) in the Data
drop box to select and change the date the user wants.
Hotkeys for date change are as follows
However, even if an arbitrary Group Class Label Tab is selected in the Class Tab when the date is changed, the
Normal Class Label Tab is always selected and displayed after the change.
10.6.3.4 Group Class Card - [Figure 10.77 Class Tab ③]
It is a card with QRS label of each group. The number of labels belonging to each Group Class is arranged in the
order of the greatest number. At the top of each Group Class Card, the type of QRS label and the percentage of
that type (%, the ratio of the number of each group class to the total number of heartbeats per day) are displayed,
and at the bottom, the serial number of the group class card and this class label The number of QRS labels that
belong to is displayed.
Figure 10.79 Group Class Label Card
1) QRS label type: Group Class Label Shows which group the QRS Label of the card belongs to.
2) Index: This is the order in which the Group Class Label Card is displayed while grouping the QRS Label
pattern in the Class Tab
3) QRS Label Number: This indicates the number of ECG patterns in the Group Class Label Card.
4) Label rate: It represents the ratio (%) of the number of QRS labels in each N/S / V/A group to the total QRS
labels during the day.
5) QRS Peak: This indicates the location of the QRS peak of the ECG pattern currently being viewed by the user.