User's Manual
User Manual
Document No.
UM-C-002-EN
Date of Revision
2020.08.30
Revision No.
0
QF-001
ATsens Co., Ltd.
46/123
Figure 10.27 AT-Patch Import Process
7) Error message according to user actions in ECG Data USB download and import
Error Message
Detail
USB download error
Failed open a FT4222 device.
please check USB cable connection.
When the dedicated cable connection is unstable during USB
download
USB download error
Not found a FT4222 device.
please check whether USB cable is connected.
When the dedicated cable is not connected during USB
download
USB download error
Communication error has occurred.
please retry again.
When communication fails for an unknown reason during
USB download (reconnect AT-Patch, turn on the power, and
try again.)
USB download error
PSN is empty or invalid format.
please enter PSN again.
When clicking USB download, when the SN (Serial
Number) is empty or an invalid SN is entered.
USB download error
This model does not support USB download.
When clicking on USB download When the device does not
support USB download
USB download error
File open error.
When the disk to be downloaded is insufficient for USB
download
USB download error
AT-Patch is powered off
please turn on AT-Patch.
When AT-Patch is powered off during USB download
USB download error
AT-Patch is not matched
please check SN(serial number) of AT-Patch.
When the SN (Serial Number) entered during USB
download and the SN of AT-Patch are different
USB download error
AT-Patch USB download is failed
When it is not possible to get all the stored data of AT-Patch
when downloading via USB