19. ANNEX 1 EMC (Electromagnetic Compatibility) Information
Applicable Standards and Test Methods
5.2.1.1 a)
The EUT uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
5.2.1.1 b)
The EUT is suitable for use in ail establishments, including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes
5.2.1.1 c)
WARNING : Use of this equipment adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
5.2.1.1 d) e)
Not applicable
5.2.1.1 f)
WARNING : Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30cm (12inches) to any part of the [ME EQUIPMENT or ME SYSTEM],
including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could
result.
5.2.1.2
Not applicable(CLASS B)
5.2.2.1 a) ~ c)
Phenomenon
Basic EMC
standard or test
method
Operating
mode
Port tested
Test
Voltage
Test
level/require
ment
Mains terminal
disturbance voltage
CISPR 11:2015
N/A
N/A
N/A
N/A
Radiated disturbance
CISPR 11:2015
Continuous
Enclosure
Battery
Group 1,