11. How to Use the Device (ATP-C130)
11.1 How to Use and Attach the Device (ATP-C130)
11.1.1 How to use Device (ATP-C130)
1) Initial devices (Device, ATP-C130) are packaged in “Sleep Mode” and delivered to customers. For patients
who want to attach the device, press the button in device for more than 3 seconds. After this, the green LED blinks
twice, and after the sound “beep” is generated once, the device is normally changed to “Active Mode”.
2) After attaching the device (Device, ATP-C130) and initially connecting to the App (ATN-C130), the device
operates in the normal “Active Mode”, and the device is “Sleep Mode ”, it operates in“ Active Mode ”, and after
the device is operated for the duration of use, it automatically enters“ Sleep Mode ”, the red LED flashes once,
and the sound“ "beeps" (beep--- beep beep---) “It occurs 3 times.
3) The device (Device, ATP-C130) and AT-Report (PC SW) Hook-up or App connection must be connected only
once for the first time, otherwise the connection is terminated as shown in 2). The device does not work properly.
After the initial connection, even if AT-Report (PC SW) Hook-up or App Bluetooth is terminated due to reasons,
ECG measurement is performed normally.
※ You cannot use the device in case of a PC or mobile phone that does not support the Bluetooth function.
4) Whether the device (Device, ATP-C130) is operating normally can be checked only through real-time live
monitoring of the App or AT-Report (PC SW) Hook-up.
5) AT-Patch is not terminated for the period of use (14 days) when the connection is confirmed after the first
connection (Connecting).
※ Physical Power-Off is not possible, and it is automatically powered off when the use period (14 days) ends.
※ If you cannot check the buzzer sound and Red LED after the use period of AT-Patch is over, you can check it
only with App and AT-Report (PC SW) Hook-up
11.1.2 How to Attach the Device (ATP-C130) to the Human Body
1) Skin Condition before Attaching to the Human Body
The skin care area of the patient should be larger than the area where the Device (ATP-C130) is attached.
If there is a lot of hair in the attachment area, shaving is necessary. However, if there is a wound or bleeding
during hair removal, it should be attached after the bleeding has stopped.