Recirculating chillers provide clean
and reliable temperature controlled fluid
Chillers/Recirculating Water Cooling Systems
WhisperCool
™
Chiller
for ICP-OES and ICP-MS
Water cooling systems (chillers) provide clean, reliable temperature
controlled fluid for open tanks or closed loop cooling. The PolyScience
WhisperCool Refrigerated Chiller is designed to deliver performance
over a broad range of operating temperatures and conditions.
For added operational safety, they also feature audible and visual
alarms that alert you instantly when operating parameters exceed
pre-set limits. There are even user adjustable high and low temperature
limits to help prevent unauthorized set point changes, plus built-in
safeties to alert you to low fluid levels and thermal runaway. Extremely
dependable and energy efficient, it features a -10° to 40 °C range,
built-in process and equipment protection, and highly intuitive.
Heat Exchanger System
Blower and Vent Assembly for AA and ICP
A venting system is required to remove fumes and vapors from the torch
of ICP emission spectrometers. A vent is recommended for use over the
power supply unit of most ICP spectrometers for removal of dissipated
heat. Use exhaust venting to:
■
Protect lab personnel from toxic
vapors
■
Protect your instrument from
corrosive vapors
■
Improve stability of the ICP torch
Includes Exhaust Hood, Adapter and
Blower. Does not include ducting.
PerkinElmer service engineers are not
permitted to install this unit.
■
Extra-large digital displays
■
Temperature and pressure/
flow rate readouts
■
Low flow alarm
■
High and low temperature alarms
■
One-touch temperature control
■
Quiet operation
■
Simple maintenance
Chiller Coolant Mix
Non-glycol coolant, which is made up of five half gallon bottles of
distilled water and includes an additive to control black algae and
other resistant strains.
Part No.
N0776099
ELAN Coolant Fluid
1 liter bottle. It is essential that this chiller coolant be used with ELAN
9000/6X00/DRC series instruments. Also suitable for organics chillers.
Part No.
WE016558
Cooling systems for the ELAN 9000/DRC II/e systems. Air cooled recirculator
without refrigeration. Not to be placed in areas with temperatures above 30 °C
(86 °F). Requires the use of ELAN Coolant Fluid (WE016558).
ICP-MS Power Requirement Part No.
120 V, 60 Hz N8122248
220/250 V, 50/60 Hz N8122247
Description Part No.
110V 03030447
230V 03030448
ICP Model Power Requirement Operating Temp. Flow Rate Part No.
ELAN 5000/6X00/DRCs 208–230 V, 60 Hz, 8 A -15° to 40 °C 60 psi: 4.3 gpm/16.3 Lpm N0772036
ELAN 5000/6X00/DRCs 240 V, 50 Hz, 8.5 A -15° to 40 °C 60 psi: 4.3 gpm/16.3 Lpm N0772035
Optima 2X00/4X00/5X00/7X00 208–230 V, 60 Hz, 8 A -15° to 40 °C 60 psi: 4.3 gpm/16.3 Lpm N0772036
Optima 2X00/4X00/5X00/7X00 240 V, 50 Hz, 8.5 A -15° to 40 °C 60 psi: 4.3 gpm/16.3 Lpm N0772035
Optima 3X00 120 V, 60 Hz, 13.1 A -5° to 40 °C 60 psi: 1 gpm/3.8 Lpm N0691883
Optima 3X00 240 V, 50 Hz, 7.3 A -5° to 40 °C 60 psi: 1 gpm/3.8 Lpm N0691884
ORDER TODAY