Specifications
Applications
Ryton
®
Scott Spray Chambers
The Ryton
®
Scott Spray Chamber is made from corrosion-resistant
Ryton
®
, which is inert to most mineral acids including HF, aqua regia
and all organic solvents normally used in ICP analyses. The Scott-type
design minimizes pulsations from the peristaltic pump. The Type II,
two-piece design ensures leak-free performance. The removable inner
tube allows the chamber to be easily disassembled for cleaning.
Ryton
®
Scott Spray Chambers
Baffled Cyclonic Spray Chambers
25
Baffled Cyclonic Spray Chambers
The Baffled Cyclonic Spray Chamber features a central transfer tube
which acts as a secondary particle separator, helping to separate larger
aerosol particles from the sample. This reduces solvent load in the
plasma without compromising detection limits.
Baffled Cyclonic Spray Chamber
Ryton
®
Scott Spray Chamber
The internal baffle design has two important characteristics:
■
The distance between the bottom of the baffle and the chamber wall is
carefully controlled so that droplets cannot form on the bottom of the baffle.
The advantage is smooth draining and improved precision
■
The baffle is cut at a sharp angle so that a large cross-section of the spray
chamber is sampled as opposed to a straight cut which would have a very
small sampling cross-section. The advantage is more representative sampling
and reduced matrix effects
ICP Model Type Part No.
Optima 2X00/4X00/5X00/7X00 DV Standard N0776053
Optima 2X00/4X00/5X00/7X00 DV O-Ring Free N0775352
Optima 2X00/4X00/5X00/7X00 DV Standard N0776090
for Oil Analysis (4 mm baffle)
Optima 2X00/4X00/5X00/7X00 DV O-Ring Free,Water-Cooled N0774100
Optima 2X00/4X00/5X00/7X00 DV O-Ring Free N0775355
for Oil Analysis (4 mm baffle)
Optima 4300V/5300V/7300V Standard N0771501
Optima 4300V/5300V/7300V O-Ring Free N0775356
Optima 4300V/5300V/7300V HF Resistant N0771550
Optima 5X00/7X00 DV, (7 mm baffle) N0777036
Drain line included without aux. gas port and it connects directly to ESI injectors.
ICP Model Part No.
Optima 2X00/4X00/5X00/7X00 N0770357
Optima 3000/3000 SCR/3X00 RL N0690640
Optima 3X00 DV/XL/3000 SCX N0690720
Plasma 400 (complete) N0580542
Plasma 400 (inner tube) N0582259
Plasma 400 (outer chamber) N0582258
Ryton
®
Scott External Spray Chamber Kit
Includes external Ryton
®
Scott Type II Spray Chamber, 1.2 mm
Quartz Injector and appropriate mounting hardware. For use with
high-volatility samples.
ICP Model Part No.
Optima 3X00 Series N0690289
Organic Sample Introduction Kit
Includes 4mm Baffled Cyclonic Spray Chamber, 3-slot Torch, GemCone
low-flow Nebulizer, and 1.2mm Alumina Injector.
Part No.
Optima 2X00/4X00/5X00/7X00 DV N0770420
ORDER TODAY
Order online at www.perkinelmer.com/supplies
Metal Ball Joint Clip
Used to connect spray chamber and torch.
Includes tension screw.
Part No.
N0773114
UniFit Drain Connector with Tubing
for Cyclonic Spray Chambers
Sample Tube Size Pkg. Part No.
1
⁄16 in. o.d. x 0.75 mm i.d. x 700 mm 3 N0774080
1.3 mm o.d. x 0.75 mm i.d. x 700 mm 3 N0774077