Unique behavior
Atlas Scientific’s microprobe technology has some unusual properties.
If your Atlas Scientific micro ORP or reference probe have been allowed to dry out for an
extended period of time, it can be reactivated by placing the probe in a small amount
ORP storage solution. Let the micro probe sit for 1 hour.
Unaffected by drying
Reactivation of a dried probe
ORP probes need to be kept wet at all times. If an ORP probe is allowed to dry out, the
probe could be permanently damaged.
Atlas Scientific’s micro probes seem to be unaffected by drying. We have found that if a
micro ORP or micro reference probe is allowed to dry, the probe still works normally when
it's put back into water. We have seen that the probes can be stored dry for many months
without any change in performance.
We believe this is related to the unusually small amount of glass used at the tip of the
probe.
*Atlas Scientific does not recommend storing the probes dry.