Datasheet

Electrical isolation
The Atlas Scientific Gravity
Analog Dissolved Oxygen Meter is a very sensitive device and
is susceptible to electrical interference from external sources. This interference is caused
by micro-voltages entering the water from unnatural sources such as pumps, solenoid
valves, or other probes / sensors. When electrical interference is affecting the D.O. readings,
it is common to see rapidly fluctuating readings or readings that are consistently off.
To verify that electrical noise is causing inaccurate readings, place the D.O. probe in a cup
of water by itself. The readings should stabilize quickly, confirming that electrical noise
was the issue.
To stop electrical interference from affecting the readings, Atlas Scientific strongly
recommends using our Gravity
Analog Isolator. It is 100% effective at preventing
electrical currents in the water from entering the probe and interfering with the readings.
Refer to the Gravity
Analog Isolator datasheet for more information about how the
isolator works and how to use it.