Datasheet

Calibration
r 0.1
Cal,nn.nn
<cr>
<cr>
<cr>
<cr>
calibration to O2 levels at your altitude.
nn.nn =%o2
Cal,0 calibrate device to 0 oxygen
Cal,clear
Cal,?
delete calibration data
device calibrated?
Command syntax
<cr>
<cr>
<cr>
<cr>
<cr>
*OK
<cr>
*OK
<cr>
*OK
Calibrated to O2 concentration at sea level
Cal,20.95
Cal,0
Cal,clear
Cal,?
Altitude
(feet)
Altitude
(meters)
%
Example Response
?Cal,0 or ?Cal,1 or ?Cal,2
*OK
<cr>
<cr>
<cr> <cr>
single point two point
After ~1 year the sensor may need re-calibration.
A single point calibration to atmospheric O2 levels
is all thats needed. 0 point calibration can also be
done if accuracy at low O2 levels is needed.
1,000
5,000
10,000
305
1,524
3,048
20.1
17.3
14.3
20
Copyright © Atlas Scientific LLC