Datasheet
r 0.1
Alarm
Command syntax
<cr>
<cr>
<cr>
<cr>
Alarm,n
Alarm,en,[1,0]
Alarm,tol,n
Alarm,?
enable / disable alarm
sets alarm
sets alarm tolerance (0 – 60)
alarm set?
<cr>
<cr>
<cr>
<cr>
<cr>
*OK
<cr>
*OK
<cr>
*OK
Alarm,en,1
Alarm,5.5
Alarm,tol,1
Alarm,?
if all are enabled
Enable alarm
O2 level must fall one percentage point
below set point for alarm to reset.
?,alarm,5.50,1.00,1
<cr>
Example Response
The alarm pin will = 1 when O2 levels are > alarm set point.
Alarm tolerance sets how far below the set point O2 levels
need to drop before the pin will = 0 again.
Alarm
(5.5%)
Alarm
Alarm on Alarm off
(Alarm set point - tolerance)
(5.5% - 1%)
4.5%
Tolerence
(1%)
O
2
19