This galvanic dissolved oxygen probe is a passive device that generates a small voltage
from 0mv – 60+ mv depending on the oxygen saturation of the PTFE sensing membrane.
This voltage can easily be read by a multimeter or an analog to digital converter.
Because every use case is different, there is no set schedule for recalibration.
The dissolved oxygen probe reacts with oxygen in the water, the more oxygen it reacts
with the more the probe is depleted of its electrolyte solution. Typically a dissolved oxygen
probe will last ~6 months before the electrolyte is depleted (results will vary). When the
electrolyte is depleted, the probe will read very low numbers. Best practice is to replace
the electrolyte solution and PTFE membrane every 6 – 12 months.
The Mini Lab Grade Dissolved Oxygen probe comes with:
How often do you need to
recalibrate a dissolved oxygen probe?
V
V
0000
Dissolved Oxygen
Electrolyte Solution
30ml
30ml
Dissolved Oxygen
Electrolyte Solution
Syringe Replacement
Membrane Cap