Datasheet

Operating principle
A galvanic dissolved oxygen probe consists of a PTFE membrane, an anode bathed in
an electrolyte and a cathode. Oxygen molecules defuse through the probes membrane
at a constant rate (without the membrane the reaction happens to quickly). Once the
oxygen molecules have crossed the membrane they are reduced at the cathode and
a small voltage is produced. If no oxygen molecules are present, the probe will output
0 mV. As the oxygen increases so does the mV output from the probe. Each probe will
output a different voltage in the presence of oxygen. The only thing that is constant is that
0mV = 0 Oxygen.
Electrolyte
Cathode (Pure silver rod)
Membrane
Anode (Zinc)
Time (sec)
90%
60%
5 10 15 20 25 300
Dissolved Oxygen
Stagnant water
Oxygen being
consumed by probe
20%
One of the drawbacks from using a galvanic
probe is that it consumes a VERY small amount
of the oxygen it reads. Therefore, a small amount
of water movement is necessary to take accurate
readings. Approximately 60 ml/min.
5
Copyright © Atlas Scientific LLC