Data Sheet
Operating principle
A galvanic dissolved oxygen probe consists of a Polytetrauoroethylene membrane, an
anode bathed in an electrolyte and a cathode. Oxygen molecules defuse through the
probes membrane at a constant rate (without the membrane the reaction happens to
quickly). Once the oxygen molecules have crossed the membrane they are reduced
at the cathode and a small voltage is produced. If no oxygen molecules are present, the
probe will output 0 mV. As the oxygen increases so does the mV output from the probe.
Each probe will output a different voltage in the presence of oxygen. The only thing that
is constant is that 0mV = 0 Oxygen. (A galvanic dissolved oxygen probe can also be used
to detect the Oxygen content in gases).
The Atlas Scientic
™
EZO
™
Dissolved Oxygen circuit works with:
Slow response, requires external power, expensive.
Requires external power, output in μA.
Requires no external power, output in mV.
Optical probe
Polar Graphic probe
Galvanic probe
Electrolyte
Cathode (Pure silver rod)
Membrane
Anode (Zinc)
Flow Dependence
Time (sec)
90%
60%
5 10 15 20 25 300
Dissolved Oxygen
Stagnant water
Oxygen being
consumed by probe
20%
One of the drawbacks from using a galvanic
probe is that it consumes a VERY small amount
of the oxygen it reads. Therefore, a small amount
of water movement is necessary to take accurate
readings. Approximately 60 ml/min.
6 Copyright © Atlas Scientic LLC