Data Sheet

Settings that are retained if power is cut
Baud rate
Calibration
Continuous mode
Device name
Enable/disable parameters
Enable/disable response codes
Hardware switch to I
2
C mode
LED control
Protocol lock
Software switch to I
2
C mode
Find
Pressure compensation
Salinity compensation
Sleep mode
Temperature compensation
Settings that are NOT retained if power is cut
Settings that are retained if power is cut
Baud rate
Calibration
Continuous mode
Device name
Enable/disable parameters
Enable/disable response codes
Hardware switch to I
2
C mode
LED control
Protocol lock
Software switch to I
2
C mode
Find
Pressure compensation
Salinity compensation
Sleep mode
Temperature compensation
Settings that are NOT retained if power is cut
Pressure compensation
P,n
<cr>
<cr>
n = any value in kPa
P,? compensated pressure value?
Command syntax
<cr>
<cr>
<cr>
*OKP,90.25
P,?
Example Response
?,P,90.25
*OK
<cr>
<cr>
Default value = 101.3 kPa
This parameter can be omitted if the
water is less than 10 meters deep
8.01 mg/L 6.94 mg/L
P,90. 25 <cr>
28 Copyright © Atlas Scientic LLC