User guide

AC Output Voltage:
Range: 100V to 5000V AC
Resolution: 10 Volts / step
Frequency: 50/60 Hz selectable
Waveform: Sinusoidal
Regulation: +/-(1% of setting +5V)
Voltage Display: Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
AC Current Display: Total current
Range: 0.001 to 40mA AC
Resolution: 0.001mA
Accuracy:+/-(1.5% of reading + 5cnt)
High/Low Limit Test: 1mA to 40mA AC
Accuracy: +/-(1% of limit +1mA)
Low limit can be turned OFF
ArcDetection: Arc Level: >1mA to 40 mA & OFF in
0.001mA / Step
DC Output Voltage
Range: 100V to 6000V DC
Resolution: 10Volts / step
Regulation: +/-(1% of setting +5V)
Voltage Display: Accuracy: +/-(1% of reading + 5V)
Resolution: 1Volt
DC Current Display: Range:1mA to 20mA DC
Resolution: 0.001mA
Accuracy:+/-(1% of reading + 5cnt)
High/Low Limit Test: 0.001mA to 20mA DC
Low limit can be turned OFF
Arc Detection: Arc Level: >1mA to 40 mA & OFF in
0.001mA / Step
Insulation Resistance
Range: 10M - 10G
Accuracy: +/- 5% >= 500V 10M - 1G
+/-15% > 500V, 1G - 10G
+/-10% < 500V, 1M - 200M
+/- 15% < 200V, 200M - 1G
Voltage Range: 50V to 1000V DC
Voltage Accuracy: +/-(1% of setting +5V)
High/Low Limit Test: 10M - 10G
High limit can be turned OFF
IR Test Delay: 0.3 to 99.9 seconds programmable in
0.1 second steps
Ground Bond
Output Current: Range: 1.0 to 30.0A AC, setting
0.1A / step
Display: +/-(1% of setting + 0.3A )
Frequency: 50 or 60Hz Selectable
No Load Voltage: 6 to 15 V programmable
Resistance: Range: 0 - 500.0mV, 4 digits
Accuracy: +/- (1% of reading + 3 cnts)
Hi Limit: 10 mV to 510 mV
Offset Function: 0 to 100 mV, user selectable
Test Time: 0.5 - 999sec(+/-20ms)
Common Features
AC/DC Test Time: Ramp: 0.1 to 99.9s (+/- 20ms)
Test: 0.1 to 999s (+/- 20ms) and
Continuous
Remote Control: Inputs: Start,Stop
Characteristics: Optically Isolated with
+24V Active Low, Pulse Width >1ms.
Outputs: Pass/Fail/Under Test
Characteristics: Dry Contact relay
Ele. Characteristics: 120V 100mA max.
Logic: Closed if True
Connect: Terminal Strip & 9 pin D Serie
Test Setups: 50 Test Setups with 15 Steps each
Connectors: Front Connection and Rear
Connections
Front Panel Lockout: Password
Safety Features: Fast Cutoff (<0.4ms) and Fast
Discharge
Miscellaneous: Continuous Voltage on Fail
Scanner Delay: 0.1 to 99.9, 0.1s/step
Indication: Pass/Fail lights, audible sound
Buzzer Level: 1,2,3 and Off
Standard Interface: IEEE488
Optional Interface: RS232
Data Bits: 8, Parity: None, Stop Bits: 1
Default Baud Rate: 9.6k
EOS: CR + LF, Echo: Off
Dimensions: (w x h x d):17 x 6.8 x 17.7in
(430 x 175 x 450mm)
Weight: 53 lbs (24kg), Net
60 lbs (27kg), Ship
Environmental: Operating: 0° to + 4°C
Humidity: <75%
Storage: - 10° to + 60°C
Power: • 90 - 130V AC • 50 or 60Hz
• 200 - 250V AC • 300W max
Guardian 5000 Electrical Safety
Analyzer
Includes:
150206 Instruction Manual
700070 Power Cable
S02 Test Leads
G15 Ground Continuity Lead Set
Calibration Certificate Traceable to NIST
Optional Accessories:
Calibration Data
5000-01 Scanner, 8 Channel HV
5000-02 Scanner, 8 Channel HV/4GC
5000-03 Scanner, 8 Channel HV Rack
Mountable
5000-04 Scanner, 8 Channel HV/4GC
Rack Mountable
S04 HV Lead Set, 2m
S05 Foot Switch
S06 High Voltage Probe
S10 HV Lead, 2 meters, unterminated
S11 Gun Probe with remote start
S12 Load Box, resistive
S14 Load Box, custom resistors
S08 Gun Probe
S09 HV Lead, 1 meter, unterminated
G13 Corded Product Adapter
G14 Power Entry Adapter
G15 GC Lead Set, high &low
G16 International Power Strip
G22 Rear Panel Connection (factory
installed)
G23 Rack Mount Ears
G25 Corded Product Adapter (240V)
G26 RS232 Interface
Ordering Information
For more detailed specifications, visit www.quadtech.com
For more information about special purchase, rent & lease options, call
1-800-253-1230 Fax 1-978-461-4295 Intl. 1-978-461-2100
5 Clock Tower Place, 210 East, Maynard MA 01754
P/N 030072/A2
Guardian 5000