Manual
• Field validation of vibration sensors, proximity probes and
related vibration monitoring equipment
• Battery powered for up to 18 hours
• Rugged and portable
• Wide frequency range: 7 Hz to 10 kHz (420 to 600000 CPM)
• Internal memory and USB flash drive output
• Easy certificate of calibration with supplied Excel template
• Complete turnkey system
• Calculates and displays test sensor sensitivity automatically
• Digital closed-loop control for enhanced stability and accuracy
• Built-in ICP
®
input for common piezoelectric accelerometers
• Traceable calibration of accelerometers per ISO 16063-21
• Check alarm & alert trip points in condition monitoring systems
• Ideal for “in situ” and laboratory work
“Simplifying with Smart Sensing Solutions”
info@modalshop.com
Phone 513-351-9919
MODEL 9110D
PORTABLE VIBRATION CALIBRATOR
The 9110D Portable Vibration Calibrator is the ideal tool
for checking accelerometers, velocity transducers and
proximity probes over a wide operating frequency and
amplitude range. The unit is a compact, battery-powered,
and completely self-contained vibration reference source
which can be conveniently used to calibrate individual
sensors, vibration switches and data collectors, as well as
to validate the entire measurement channel of a condition
monitoring or recording system. An integral precision
quartz reference accelerometer and closed-loop level
control gives the 9110D enhanced stability and superior
vibration calibration over an extended 7 Hz to 10 kHz
frequency range. Packaged in a rugged Pelican
®
Storm
case, the 9110D is always ready for travel to test sites
bringing laboratory accuracy to the field.
Additional features include an ICP
®
or voltage test sensor
input for direct connection and readout of the most com-
mon types of accelerometers and velocity transducers.
The test sensor sensitivity is calculated and displayed on
the screen in real time. The unit’s internal memory capa-
bility can store up to 500 calibration records and data can
be easily transferred to a computer through a USB flash
drive. This allows for the creation and printing of ISO
17025-compliant customizable calibration certificates
and reports using the supplied Excel worksheet template.