User guide

Model PRO-HPULSE
Pulse Magnetic Field for compliant testing per IEC 61000-4-9 and EN 61000-4-9
Field Pulse 8/20µs
Field Amplitude 100A/m - 1000A/m, ±10%
Resolution 5A/m
Coil Factor 0.65 to 1.00
Options Model CM-HMON: Measurement probe for power
frequency magnetic fields
Model CM-HCOIL: 1m x 1m magnetic field coil
Model PRO-PQF
Dips and Interrupts for compliant testing IEC 61000-4-11 Edition 2, and EN 61000-4-11
Dips 40%, 70%, 80%
Interrupts 0% (short and open)
Transition Time 1µs - 5µs
Inrush Minimum 250Amps @ 100 - 120V,
Minimum 500Amps @ 220 - 240V
AC Voltage 50 - 250VAC, 50/60Hz
AC Current 16A max.**
PQF Sync Output 5V signal occurs at each dip or interrupt transition
Options Model PQF-QUAL: Circuit per IEC 61000-4-11 for
testing PQF generator inrush capability
PQF Waveform Monitoring
Voltage Input Connection Fixed, L1 to L2
Voltage Attenuation 100:1 ±5%
Current Input Connection Fixed, L1
Peak Current Minimum 500A inrush into 1700µF
Current Attenuation 200:1 ±5%
OPTIONAL COUPLERS/DECOUPLERS
Model CM-3CD-16 & CM-3CD-32
*
Semi-automatic, stand alone, three-phase AC/DC mains coupler/decouplers for EFT &
Surge per IEC 61000-4-4, Edition 2 and IEC 61000-4-5
ELECTRICAL
Waveforms EFT: 5/50ns, per IEC 61000-4-4
Surge: Combination wave: 1.2/50µs open-circuit
voltage, 8/20µs short-circuit current, per IEC 61000-4-5
Maximum Surge 6.6kV, 3.3kA
Voltage & Current
Maximum EFT Voltage 4.4kV
Coupling Modes EFT: L1, L2, L3, N or PE
Surge Hi: L1, L2, L3 or N
Surge Lo: L1, L2, L3, N or PE
*
Not available for delivery until October 2004
COUPLER/DECOUPLERS
AC Voltage 50 to 250V, 50/60Hz line to ground, 50 to 433V
line to line
AC Current CM-3CD-16: 16A/phase continuous
CM-3CD-32: 32A/phase continuous
DC Current CM-3CD-16: 16A up to 48V
8A up to 110V
1.2A up to 220V
0.3A up to 440V
CM-3CD-32: 25A up to 48V
8A up to 220V
1.2A up to 220V
0.3A up to 440V
EUT Mains Safety Sockets
Output Connectors
POWER REQUIREMENTS
Input Voltage 90-250VAC, 50/60Hz
Input Current 1A at 120VAC; 0.5A at 240VAC
Model CM-I/OCD
I/O coupler/decoupler - provides the ability to couple surges from EMCPro PLUS or any
surge simulator, to I/O or data lines per IEC 61000-4-5
ELECTRICAL
Waveforms Designed to couple combination waves of 1.2/50µs
open-circuit voltage, 8/20µs short-circuit current
supplied by option PRO-SURGE with the KeyTek
EMCPro PLUS
Repetition Rate Up to 5 per minute at 4.4kV
Data Line Frequency To greater than 100kHz without significant degradation
when CM-I/OCD-HS is installed. Option CM-I/OCD-HS
is recommended for data line frequencies greater than
1kHz
Number of Lines Eight lines - any line can be surged to any other line or
ground
Maximum Surge Voltage 4.4kV
Maximum Signal Line Voltage 200V
Maximum Signal Line Current 1A AC or DC
Clamping Selectable built-in clamps of 20V and 220V; external
bias input for other clamp levels
Available Options CM-I/OCD-HS: Internally-installed option provides
selectable parallel resistors (400s, 200s, 100s) - highly
recommended for data line frequencies greater than
1kHz.
Model CM-TELCD
Telecom line coupler/decoupler - provides the ability to couple both the telecom wave and
combination wave per IEC 61000-4-5
ELECTRICAL
Waveforms Designed to couple 1.2/50µs combination or 10/700µs
telecom waves
Telecom Line Frequency To 100kHz without significant degradation
Number of Lines Up to four lines - one or two pairs of balanced Telecom
lines
Maximum Surge Voltage 4.4kV
Maximum Signal Line Voltage 200V
Maximum Signal Line Current 1A AC or DC
Clamping Selectable built-in clamps of 20V and 225V: external
bias input for other clamp levels