User guide
Model PRO-ESD
ESD per IEC 61000-4-2 and EN 61000-4-2
Trigger Modes One shot manual, multi-shot tripod
Repetition Rate Single shot, 1pps or 20pps
Air Discharge Voltage 500V - 8.8kV ±10%
Contact Discharge Voltage 500V - 4.4kV ±10%
Discharge Capacitor 150pF ±10%
Discharge Resistance 330Ω ±10%
Charging Resistance 50MΩ - 100MΩ
Polarity Front panel or software controlled
Shot Counter 1 - 999 discharges
Energy Storage 5.8mJ @ 8.8kV
Model PRO-EFT
EFT per IEC 61000-4-4 Edition 2, EN 61000-4-4 and ANSI C62.41
Voltage Waveform 5/50ns ±30%
Peak Voltage 250V - 4.4kV ±5%
Burst Period 300ms ±10%
Burst Duration 15ms ±20%, for pulse frequencies uo to 5kHz, 0.75ms
above 5kHz
Frequency 1-100kHz, in 0.5kHz steps, ±10%
DC Blocking Capacitor 10nF (internal)
Options Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT &
surge coupler/decoupler
Model CM-CCL: Capacitive coupling clamp
Model CM-CCLC: Coupling clamp cover
Model EFT-ATTN: EFT attenuator for oscilloscope
monitoring
Model PRO-SURGE
Surge for compliant testing per IEC 61000-4-5, EN 61000-4-5, ANSI C62.41 Category B and
UL 1449
Voltage Waveform 1.2/50µs
Peak Voltage 250V - 6.6kV ±5%, 12Ω mode
250V - 6.0kV ±5%, 2Ω mode
Peak Current 125A - 3.3kA ±10%
Additional 10Ω Resistor Software selectable
Repetition Rate Up to 4 per minute
Open-circuit Voltage Front time: 1.2µs ±30%
Duration: 50 µs ±20%
1
Undershoot: ≤ 30%
Short-circuit Current Front time: 8.0µs ±20%
Duration
*
50µs ±20%
Undershoot ≤ 30%
Line sync accuracy ±15%, 50 - 277VAC
Options Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT
& surge coupler/decoupler
Model CM-I/OCD: External 8 line coupler/decoupler
for I/O signal lines
Model CM-I/OCD-HS: High speed I/OCD option for
testing data rates to >100kHz
*
Durations are reduced in 12Ω mode and when coupling multiple lines to PE
Model PRO-RING
**
Ring Wave Surge per ANSI C62.41 Cat. A, B, and UL 864
Voltage Waveform 100kHz damped cosine
Peak Voltage 250 - 6.6kV ±5%
Repetition Rate <4/minute at 6kV, faster at lower voltages
Open-circuit Voltage Rise Time: 0.5µs ±30%
Short-circuit Current Vp/Ip: 12Ω ±3Ωor 30Ω ±8Ωsoftware selectable
Options Model CM-3CD-16/32: 16 or 32 Amp, 3 phase EFT
& surge coupler/decoupler
Model PRO-TELECOM
**
Surge Telecom compliant testing per IEC 61000-4-5, EN 61000-4-5, FCC Part 68, ITU K.17,
K.20, K.21 and ETSI
Voltage Waveform 10/700µs (9/720µs FCC Part 68)
Peak Voltage 250V - 6.6kV ±5%
Peak Current 6.25 - 165A +10/-0%, 40Ω mode
Repetition Rate Up to 4 per minute
Open-circuit Voltage Front time: 7.0µs to 11.7µs
Duration: 576µs to 840µs
Short-circuit Current Front time: 3.5µs to 6.5µs
Duration: 256µs to 384µs
Options Model CM-TELCD: External coupler for telecom lines
Surge Waveform Monitoring
Lines Monitored Monitors are automatically switched to match
generator coupling mode
Open-circuit Voltage 1000:1 ±10%
Short-circuit 200:1 ±7%
Current Attenuation
Model PRO-HPOWER
Power Frequency Magnetic Field for compliant testing per IEC 61000-4-8 and EN 61000-4-8
Field Frequency 50Hz/60Hz
Field Amplitude 0.5 - 4A/m, in 0.25A steps, ±10% (with CM-HCOIL)
up to 100A/m with optional external HPOWER-EXT
AC Source Internal
Resolution 0.25A minimum
Coil Factor 0.65 to 1.00
Coil Resistance 0.05Ω maximum
Options Model CM-HMON: Measurement probe for power
frequency magnetic fields
Model CM-HCOIL: 1m x1m magnetic field coil
Model HPOWER - EXT: External generator for power
frequency magnetic field to 30A/m
**
PRO-TELECOM and PRO-RING can not be installed in same unit.