8 Fluke Corporation 5500A Black and White Extended Specifications
AC Voltage (Sine Wave) Specifications (cont.)
Range Frequency
Maximum Distortion and Noise
10 Hz to 5 MHz Bandwidth
± (% output + µV)
10 to 45 Hz 0.15 % + 90 µV
45 Hz to 10 kHz 0.035 + 90 µV
10 to 20 kHz 0.06 + 90 µV
20 to 50 kHz 0.15 + 90 µV
50 to 100 kHz 0.25 + 90 µV
1.0 to 32.999 mV
100 to 500 kHz 0.3 + 90 µV
10 to 45 Hz 0.15 % + 90 µV
45 Hz to 10 kHz 0.035 + 90 µV
10 to 20 kHz 0.06 + 90 µV
20 to 50 kHz 0.15 + 90 µV
50 to 100 kHz 0.20 + 90 µV
33 to 329.999 mV
100 to 500 kHz 0.20 + 90 µV
10 to 45 Hz 0.15 % + 200 µV
45 Hz to 10 kHz 0.035 + 200 µV
10 to 20 kHz 0.06 + 200 µV
20 to 50 kHz 0.15 + 200 µV
50 to 100 kHz 0.20 + 200 µV
0.33 to 3.29999 V
100 to 500 kHz 0.20 + 200 µV
10 to 45 Hz 0.15 % + 2 mV
45 Hz to 10 kHz 0.035 + 2 mV
10 to 20 kHz 0.08 + 2 mV
20 to 50 kHz 0.2 + 2 mV
3.3 to 32.9999 V
50 to 100 kHz 0.5 + 2 mV
45 Hz to 1 kHz 0.15 % + 10 mV
1 to 10 kHz 0.05 + 10 mV
33 to 329.999 V
10 to 20 kHz 0.6 + 10 mV
45 Hz to 1 kHz 0.15 % + 30 mV
330 to 1000 V
1 to 10 kHz 0.07 + 30 mV
5725A Amplifier
45 Hz to 1 kHz 0.07 %
1 to 20 kHz 0.15 %
100 to 1000 V
20 to 30 kHz 0.3 %
100 to 750 V 30 to 100 kHz 0.4 %
Auxiliary Output (dual output mode only) 10 Hz to 100 kHz Bandwidth
10 to 20 Hz 0.2 % + 200 µV
20 to 45 Hz 0.06 + 200 µV
45 Hz to 1 kHz 0.08 + 200 µV
1 to 5 kHz 0.3 + 200 µV
10 to 329.999 mV
5 to 10 kHz 0.6 + 200 µV
10 to 20 Hz 0.2 % + 200 µV
20 to 45 Hz 0.06 + 200 µV
45 Hz to 1 kHz 0.08 + 200 µV
1 to 5 kHz 0.3 + 200 µV
0.33 to 3.29999 V
5 to 10 kHz 0.6 + 200 µV