Manual
5500A Black and White Extended Specifications Fluke Corporation 5
Resistance Specifications
Absolute Uncertainty, tcal ± 5 °C
± (% of output + Ω)
[2]
Range
[1]
90 days 1 year
Resolution
Ω
Allowable
Current
[4]
0 to 10.99 Ω 0.009 0.008
[3]
0.012 0.008
[3]
0.001 1 to 125 mA
11 to 32.999 Ω 0.009 0.015
[3]
0.012 0.015
[3]
0.001 1 to 125 mA
33 to 109.999 Ω 0.007 0.015
[3]
0.009 0.015
[3]
0.001 1 to 70 mA
110 to 329.999 Ω 0.007 0.015
[3]
0.009 0.015
[3]
0.001 1 to 40 mA
330 Ω to 1.09999 kΩ 0.007 0.06 0.009 0.06 0.01 250 µA to 18 mA
1.1 to 3.29999 kΩ 0.007 0.06 0.009 0.06 0.01 250 µA to 5 mA
3.3 to 10.9999 kΩ 0.007 0.6 0.009 0.6 0.1 25 µA to 1.8 mA
11 to 32.9999 kΩ 0.007 0.6 0.009 0.6 0.1 25 µA to 0.5 mA
33 to 109.999 kΩ 0.008 6 0.011 6 1 2.5 µA to 0.18 mA
110 to 329.999 kΩ 0.009 6 0.012 6 1 2.5 µA to 0.05 mA
330 kΩ to 1.09999 MΩ 0.011 55 0.015 55 10 250 nA to 0.018 mA
1.1 to 3.29999 MΩ 0.011 55 0.015 55 10 250 nA to 5 µA
3.3 to 10.9999 MΩ 0.045 550 0.06 550 100 25 nA to 1.8 µA
11 to 32.9999 MΩ 0.075 550 0.1 550 100 25 nA to 0.5 µA
33 to 109.999 MΩ 0.4 5500 0.5 5500 1000 2.5 nA to 0.18 µA
110 to 330 MΩ 0.4 16500 0.5 16500 1000 2.5 nA to 0.06 µA
[1] Continuously variable from 0 to 330 MΩ.
[2] Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3] The floor adder is improved to 0.006 Ω (0 to 10.99 Ω range) and 0.010 Ω (11 to 329.999 Ω) if the 5500A Calibrator is zeroed
(ohms zero or instrument zero) within 8 hours and temperature is ±1 °C of zeroing ambient temperature.
[4] Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases by Floor(new) =
Floor(old) x Imin/Iactual. For example, a 100 µA stimulus measuring 100 Ω has a floor uncertainty of 0.01 Ω x 1 mA/100 µA
= 0.1 Ω.
Range Maximum Voltage
[1]
Maximum Lead Resistance
[2]
0 to 10.99 Ω 1.37 <3.2
11 to 32.999 Ω 4.12 <3.2
33 to 109.999 Ω 7.7 <3.2
110 to 329.999 Ω 13.2 <3.2
330 Ω to 1.09999 kΩ 19.8 <6
1.1 to 3.29999 kΩ 16.5 <6
3.3 to 10.9999 kΩ 19.8 <6
11 to 32.9999 kΩ 16.5 <6
33 to 109.999 kΩ 19.8 <6
110 to 329.999 kΩ 16.5
330 kΩ to 1.09999 MΩ 19.8
1.1 to 3.29999 MΩ 16.5
3.3 to 10.9999 MΩ 19.8
11 to 32.9999 MΩ 16.5
33 to 109.999 MΩ 19.8
110 to 330 MΩ 19.8
(n/a 110 kΩ and above)
[1] This is for the largest resistance for each range. The maximum voltage for other values is Imax (highest value of Allowable Current
above) multiplied by Rout.
[2] Maximum lead resistance for no additional error in 2-wire COMP.