5500A Black and White Extended Specifications Fluke Corporation 5
Resistance Specifications
Absolute Uncertainty, tcal ± 5 °C
± (% of output + )
[2]
Range
[1]
90 days 1 year
Resolution
Allowable
Current
[4]
0 to 10.99 0.009 0.008
[3]
0.012 0.008
[3]
0.001 1 to 125 mA
11 to 32.999 0.009 0.015
[3]
0.012 0.015
[3]
0.001 1 to 125 mA
33 to 109.999 0.007 0.015
[3]
0.009 0.015
[3]
0.001 1 to 70 mA
110 to 329.999 0.007 0.015
[3]
0.009 0.015
[3]
0.001 1 to 40 mA
330 to 1.09999 k 0.007 0.06 0.009 0.06 0.01 250 µA to 18 mA
1.1 to 3.29999 k 0.007 0.06 0.009 0.06 0.01 250 µA to 5 mA
3.3 to 10.9999 k 0.007 0.6 0.009 0.6 0.1 25 µA to 1.8 mA
11 to 32.9999 k 0.007 0.6 0.009 0.6 0.1 25 µA to 0.5 mA
33 to 109.999 k 0.008 6 0.011 6 1 2.5 µA to 0.18 mA
110 to 329.999 k 0.009 6 0.012 6 1 2.5 µA to 0.05 mA
330 k to 1.09999 M 0.011 55 0.015 55 10 250 nA to 0.018 mA
1.1 to 3.29999 M 0.011 55 0.015 55 10 250 nA to 5 µA
3.3 to 10.9999 M 0.045 550 0.06 550 100 25 nA to 1.8 µA
11 to 32.9999 M 0.075 550 0.1 550 100 25 nA to 0.5 µA
33 to 109.999 M 0.4 5500 0.5 5500 1000 2.5 nA to 0.18 µA
110 to 330 M 0.4 16500 0.5 16500 1000 2.5 nA to 0.06 µA
[1] Continuously variable from 0 to 330 M.
[2] Applies for COMP OFF (to the 5500A Calibrator front panel NORMAL terminals) and 2-wire and 4-wire compensation.
[3] The floor adder is improved to 0.006 (0 to 10.99 range) and 0.010 (11 to 329.999 ) if the 5500A Calibrator is zeroed
(ohms zero or instrument zero) within 8 hours and temperature is ±1 °C of zeroing ambient temperature.
[4] Do not exceed the largest current for each range. For currents lower than shown, the floor adder increases by Floor(new) =
Floor(old) x Imin/Iactual. For example, a 100 µA stimulus measuring 100 has a floor uncertainty of 0.01 x 1 mA/100 µA
= 0.1 .
Range Maximum Voltage
[1]
Maximum Lead Resistance
[2]
0 to 10.99 1.37 <3.2
11 to 32.999 4.12 <3.2
33 to 109.999 7.7 <3.2
110 to 329.999 13.2 <3.2
330 to 1.09999 k 19.8 <6
1.1 to 3.29999 k 16.5 <6
3.3 to 10.9999 k 19.8 <6
11 to 32.9999 k 16.5 <6
33 to 109.999 k 19.8 <6
110 to 329.999 k 16.5
330 k to 1.09999 M 19.8
1.1 to 3.29999 M 16.5
3.3 to 10.9999 M 19.8
11 to 32.9999 M 16.5
33 to 109.999 M 19.8
110 to 330 M 19.8
(n/a 110 k and above)
[1] This is for the largest resistance for each range. The maximum voltage for other values is Imax (highest value of Allowable Current
above) multiplied by Rout.
[2] Maximum lead resistance for no additional error in 2-wire COMP.