2 Fluke Corporation 5500A Black and White Extended Specifications
General Specifications
Warmup Time Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time Less than 5 seconds for all functions and ranges except as noted.
Standard Interfaces IEEE-488 (GPIB), RS-232, 5725A Amplifier
Temperature Performance
Operating: 0 °C to 50 °C
Calibration (tcal): 15 °C to 35 °C
Storage: -20 °C to 70 °C
Temperature Coefficient Temperature Coefficient for temperatures outside tcal +5 °C is 0.1X/ °C of the 90-day
specification (or 1-year, as applicable) per °C.
Relative Humidity
[1]
Operating: <80 % to 30 °C, <70 % to 40 °C, <40 % to 50 °C
Storage: <95 %, non-condensing
Altitude
Operating: 3,050 m (10,000 ft) maximum
Non-operating: 12,200 m (40,000 ft) maximum
Safety Complies with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation 20 V
EMC Designed to comply with FCC Rules Part 15; VFG 243/1991. If used in areas with
Electromagnetic fields of 1 to 3 V/m, resistance outputs have a floor adder of
0.508 . Performance not specified above 3 V/m. This instrument may be susceptible
to electro-static discharge (ESD) from direct contact to the binding posts. Good static
aware practices should be followed when handling this and other pieces of
electronic equipment.
Line Power
Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
Line Frequency: 47 Hz to 63 Hz
Line Voltage Variation: ±10 % about line voltage setting
Power Consumption 5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA
Dimensions 5500A Calibrator:
Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on
bottom of unit
Width, 43.2 cm (17 in), standard rack width
Depth: 47.3 cm (18.6 in) overall 5725A Amplifier:
Height, 13.3 cm (5.25 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on
bottom of unit
Width, 43.2 cm (17 in), standard rack width
Depth, 63.0 cm (24.8 in) overall.
Weight (without options) 5500A Calibrator, 22 kg (49 lb); 5725A Amplifier 32 kg (70 lb)
Absolute Uncertainty Definition The 5500A specifications include stability, temperature coefficient, linearity, line and
load regulation, and the traceability of the external standards used for calibration.
You do not need to add anything to determine the total specification of the 5500A for
the temperature range indicated.
Specification Confidence
Interval
99 %
[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be required.