Manual

5500A Black and White Extended Specifications Fluke Corporation 25
AC Current (Non-Sine Wave) Specifications
Triangle Wave &
Truncated Sine Wave
Ranges
[1]
Frequency
1-Year Absolute Uncertainty, tcal
± 5 °C,
± (% of output + % of range)
[2]
Maximum
Current Resolution
0.01 to 10 Hz 5.0 + 0.5 Two digits, e.g., 75 mA
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.25
2.9 to 92.999 mA
1 to 10 kHz 0.25 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5
10 to 45 Hz 0.25 + 0.5
Two digits
45 Hz to 1 kHz 0.25 + 0.5
93 to 929.999 mA
1 to 10 kHz 5.0 + 1.0
Six digits on each range
10 to 45 Hz 5.0 + 1.0 Two digits
45 Hz to 1 kHz 0.5 + 0.5
0.93 to 2.19 A
1 to 5 kHz 5.0 + 1.0
Six digits on each range
45 to 500 Hz 2.0 + 0.5 Two digits on each range
2.2 to 11 A
500 Hz to 1 kHz 5.0 + 1.0 Six digits on each range
0.01 to 10 Hz 5.0 + 0.5
10 to 45 Hz 0.25 + 0.5
Two digits, e.g., 50 mA
45 Hz to 1 kHz 0.25 +0.25
2.9 to 65.999 mA
1 to 10 kHz 0.25 + 0.5
Six digits on each range
0.01 to 10 Hz 5.0 + 0.5 Two digits
10 to 45 Hz 0.25 + 0.5
45 Hz to 1 kHz 0.25 + 0.5
66 to 659.999 mA
1 to 10 kHz 5.0 + 1.0
Six digits on each range
10 to 45 Hz 5.0 + 1.0 Two digits
45 Hz to 1 kHz 0.5 + 0.5
0.66 to 2.19 A
1 to 5 kHz 5.0 + 1.0
Six digits on each range
45 to 500 Hz 2.0 + 0.5 Two digits on each range
2.2 to 11 A
500 Hz to 1 kHz 5.0 + 1.0 Six digits on each range
[1] All waveforms are p-p output ranges.
[2] Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
AC Current, Square Wave Characteristics (typical)
Range Rise Time Settling Time Overshoot
I<4.4 A @ 400 Hz 25 µs 40 µs to 1 % of final value <10 % for loads <100
AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range