Manual

24 Fluke Corporation 5500A Black and White Extended Specifications
AC Voltage, Square Wave Characteristics
Rise Time
@ 1 kHz
Typical
Settling Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range Duty Cycle Uncertainty
[1]
<1 µs
<10 µs to 1 %
of final value
<2 %
1 % to 99 %, <3.3 V p-p,
0.01 Hz to 100 kHz
± (0.8 % of period +140 ns) for
frequencies >10 kHz; + (0.8 % of
period + 2 µs) for frequencies
10 kHz.
[1] For duty cycles of 10.00 % to 90.00 %.
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz Aberrations
0.3 % of p-p value, from 10 % to 90 % point <1 % of p-p value, with amplitude >50 % of range
AC Current (Sine Wave) Extended Bandwidth Specifications
1-Year Absolute Uncertainty,
tcal ± 5 °C,
± (% of output + % of range)
Range Frequency
% Output % Range
Maximum
Current Resolution
All current ranges, <330 mA 0.01 to 10 Hz 5.0 0.5 2 digits each range
10 to 10 kHz (See AC Current (Sine Wave) Specifications)