5500A Black and White Extended Specifications Fluke Corporation 19
Additional Specifications
The following paragraphs provide additional specifications for the 5500A Calibrator ac voltage and ac current functions. These
specifications are valid after allowing a warm-up period of 30 minutes, or twice the time the 5500A has been turned off. All extended range
specifications are based on performing the internal zero-cal function at weekly intervals, or when the ambient temperature changes by more
than 5 °C. (See Chapter 4, Front Panel Operations in the 5500A Operator Manual.)
Frequency Specifications
Frequency
Range
Resolution
1-Year Absolute Uncertainty,
tcal ± 5 °C
Jitter
0.01 - 119.99 Hz 0.01 Hz 25 ppm, ± 1 mHz 2 µs
120.0 - 1199.9 Hz 0.1 Hz 25 ppm, ± 1 mHz 2 µs
1.200 - 11.999 kHz 1.0 Hz 25 ppm, ± 1 mHz
[1]
2 µs
12.00 - 119.99 kHz 10 Hz 25 ppm, ± 15 mHz 140 ns
120.0 - 1199.9 kHz 100 Hz 25 ppm, ± 15 mHz 140 ns
1.200 - 2.000 MHz 1 kHz 25 ppm, ± 15 mHz 140 ns
[1] ± (25 ppm + 15 mHz) above 10 kHz