Manual
14 Fluke Corporation 5500A Black and White Extended Specifications
DC Power Specification Summary
Absolute Uncertainty, tcal ± 5 °C, ± (% of Watts output)
[1]
5500A Calibrator Current Range
Voltage Range
3.3 to 8.999 mA 9 to 32.999 mA 33 to 89.99 mA 90 to 329.99 mA
90 days 33 mV to 1020 V 0.03 0.02 0.03 0.02
1 year 33 mV to 1020 V 0.04 0.03 0.04 0.03
Voltage Range 0.33 to 0.8999 A 0.9 to 2.1999 A 2.2 to 4.4999 A 4.5 to 11 A
90 days 33 mV to 1020 V 0.07 0.05 0.08 0.06
1 year 33 mV to 1020 V 0.08 0.06 0.12 0.09
5725A Amplifier Current Range
Voltage Range
1.5 to 4.4999 A 4.5 to 11 A
90 days 33 mV to 1020 V 0.09 0.07
1 year 33 mV to 1020 V 0.10 0.08
[1] To determine dc power uncertainty with more precision, see the individual “DC Voltage Specifications” and “DC Current
Specifications” and “Calculating Power Uncertainty.”