User guide
Contact us on +64 3 943 0700, sales@commtest.com or visit www.commtest.com
SPECIFICATIONS vb8
®
DATA ANALYZER REMARKS
Sensors
Sensor input 4 channels Simultaneous sampling
Compatible Sensor Types Accelerometer, velocity,
displacement, current, voltage
output, 4 to 20 mA
AC coupled range 16 V peak-peak Allowsfor±8Vsensoroutput
swing(±80g)
DC coupled ranges 0 V to 20 V, -10 V to 10 V,
-20 V to 0 V
E.g. for reading prox-probe gap
Connectors 1xBNC(CH1)
1xLEMO(CH2/CH3/CH4)
Safetyfeature:Break-free
inline connector
Analog to digital conversion 24-bit ADC
Sensor excitation current 0 mA or 2.2 mA (configurable),
24 V maximum
2.2mArequiredforICP®-type
accelerometer
Sensor detection Warns if short circuit or not
connected
Tachometer
Sensor Laser sensor, reflective tape Sensor triggers on beam
reflection
Laser sensor range 10 cm to 2 m nominal Depends on size of reflective
tape
Other sensor types supported Contact, TTL pulse,
Keyphasor
®
Instrument has optically
isolated input
Power supply to sensor 5 V, 50 mA
TTL pulse rating 3.5V(4mA)min,28V(5mA)
max,off-state0.8V
Keyphasor® thresholds 7.7±0.5V,13.3±0.8V,
18.8±1V
Nominally8V,13V,18V
Speed range 10to300000RPM(0.2Hzto
5kHz)
Pulse width at least 0.1 ms
Accuracy +/-0.1%
Output to drive strobe Upto140Hz(8400CPM) Typical. Depends on strobe
type.Specialcablerequired.
Parameter Indication
Maximum levels
>1000g(10000m/s2)
>1000in/sec(25000mm/s)
> 100 in (2500 mm),
> 10 000 Amps
Effective limit is sensor
sensitivity and output voltage
Dynamic signal range
>95dB
Typical at 400 line resolution
Harmonicdistortion Lessthan-70dBtypical Other distortions and noise
are lower
Units gorm/s2oradB
in/sormm/sorvdB
mil or mm or µm
amps, user-defined
0-peak, peak-peak or RMS.
Auto-scale by 1000x when
required.US&SIoptionsfor
bothadB&vdB.
Magnitude&cursors Overall RMS value, waveform
true pk-pk, dual cursors,
harmonics
Digital readouts on chart
Baseaccuracy ±1%(approx.0.1dB) ForDClevel:%offullscale.
ForACsignal:%ofreading
Highfrequencyattenuation
≤0.1dB100Hzto10kHz
≤3dB>10kHzto40kHz
Attenuation tolerances are in
addition to base accuracy
AC coupling attenuation
≤0.1dB10Hzto<100Hz
≤3dB1Hzto<10Hz
Attenuation due to Integration
≤ 0.1dB1Hzto<100Hz
≤1.5dB0.2Hzto<1Hz
≤0.1dB10Hzto<100Hz
≤1.5dB1Hzto<10Hz
Lowfreq.mode:Whencoupling
=DC,Fmax≤100Hz.Normal
mode: Applicable in all other
cases. Values apply to single
integration (accel. to veloc.).
Double the values for double
integration (accel. to displ.).
Spectrum Display
Fmaxranges 25, 50, 100, 125, 150, 200, 300,
400,500,600,800,1000,1200,
1600, 2000, 2500, 3000, 4000,
5000,6000,8000,10000,
15 000, 20 000, 30 000, 40 000,
60000,80000Hz
OrequivalentCPMvalues
Or orders-based from 1X to
999X
Fminpossiblerange 0toFmax Instrument zeroes all spectral
linesbelowFmin
Resolution 400,800,1600,3200,6400,
12800lines
6400 lines max. for 2-channel
measurements.3200 lines max.
for 4-channel measurements.
Frequencyscale Hz,CPM,Orders Linear scale with zooming
Amplitude scale Acceleration, velocity,
displacement, current, or
user-defined
Linear or log scales, auto or
manual scaling
Window shapes Hanning,rectangular
Overlap (0, 12.5, 25, 37.5, 50, 62.5, 75,
87.5)%
DependentonFmaxand
number of lines
Numberofaverages 1,2,4,8,16,32,64,128 Increases sampling time
proportionally
Averaging types Linear, exponential,
peak hold, synchronous
SPECIFICATIONS vb8
®
DATA ANALYZER REMARKS
Spectrum Display Continued
Demodulation bandwidths 23 bandwidth options From125Hzto1250Hzupto
16kHzto20kHz
6Pack Upto40kHz&3200lines1
channel.Upto20kHz&1600
lines 3 channel.
Spectrum and waveform for
lowfreq,highfreq,demod.
Order tracking Upto6kHzFmax,orders-
based
Tachometerrequired,mounted
on high-speed shaft
Order tracking - Distortion Lessthan-65dB Within50%to200%speed
variation during recording
Waveform Display
Numberofsamples 1024,2048,4096,8192,
16384,32768
Time scale 10 ms to 512 seconds Or orders based from 1 to
999 revs
Time synchronous averages 1,2,4,8,16,32,64,128 Only available when tach.
triggered
LongtimewaveformFmax 25Hzto40kHzFmax 20kHz2-or4-channel
Long time waveform duration 14.7 million samples (total over
channels)
E.g.forFmax1kHz,Fsample=
2.56kHzandDuration=1.6hrs
Logging & Analysis
Output formats Instrument screen, transfer to
Ascent
®
,XML,UFF
Data storage Dual1GBnon-volatileflash
memories
Database mirror copy on
second flash memory
Data storage structure Folders/machines/points/
locations/routes
Nolimitsapplied,50character
names
Max folder size 10 000 measurement locations
Modal analysis CH1forhammer,upto3
response channels, ≤10kHz
Coherence&FRF
(Accelerance/Mobility/
Compliance)
Cross channel spectrum 1 reference, up to 3 other
sensors
Coherence&FRFforimporting
into ODS software
Balancing
Planes Up to 2 planes, 4 sensors
Speed range 30 to 60 000 RPM
Measurement type Acceleration, velocity,
displacement
Weight modes Angle 0° to 360°, fixed position,
circumference arc
E.g. weights on fan blades,
linear distance around
circumference
Remove trial weights Yes/No Automatic recalculation
Manual data entry Yes Allows re-entry of previous
balance jobs
Storage Against machines in data
structure
Nolimitsapplied
Display & Communication
Display GraphicGrayscaleLCD WhiteLEDBacklight
Resolution&size 480x320(HVGA),5.5”(140
mm)
Readable in direct sunlight
Supported languages ENG,FRE,SPA,POR,RUS,CHI FirmwarereleasesinEnglish,
translations follow
Communication with PC USBandEthernet UsePROFLASHtoupgrade
USBhostport USB2.0,supplying5V,250mA SavefolderstoUSBflashdrive
UFFexport Spectra,Coherence,FRFmag.
&phase
UniversalFileFormatfor
Modal&FRFdata
Battery & Charger
Batterytype Custom Lithium Ion pack,
7.4 V, 4500 mAh
Operating time 10 hours Backlighton(60second
time-out)
Charger type Internal charging,
automatic control
External Power pack 12 V DC,
3 A output
Charge rate 3 A nominal 3 hours for complete charge
Mechanical
Size 9.9”Wx5.8”Lx2.4”H
(252x148x60)mm
Weight 2.7 lb (1.2 kg) Including battery and strap
Environmental
Operating temperature 14°Fto122°F(-10to50)°C
Storagetemperature&
humidity
-4°Fto140°F(-20to60)°C,
95%RH
If storage exceeds 1 month: Up
to95F(35C),85%RH
EMC EN61326
Ruggedness 4’ (1.2m) drop on concrete, IP65 Procedure: 26 drops following
MIL-STD-810F-516.5-IV
Hazardouslocations CSAClassI,Division2(Groups
A,B,C,D)
Certification