Technical Data Sheet High Performance, Broadband Network Analysis Solutions MS4640B Series Microwave Vector Network Analyzers Introduction This document provides detailed specifications for the MS4640B series microwave Vector Network Analyzers (VNAs) listed below, including all related options, and accessories.
Table of Contents VectorStar™ MS4640B Series Description Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . System Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Receiver Dynamic Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . Receiver Compression . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . High Level Noise . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Noise Floor . . . . . . . . . . . . . . .
VectorStar™ MS4640B Series System Dynamic Range System Dynamic Range System dynamic range is calculated as the difference between the maximum rated source power and the specified noise floor at the specified reference plane. Option 031 System Dynamic Range is listed in alternating tables. MS4642B 20 GHz Model, System Dynamic Range (dB) at Ports 1 or 2 at b1 or b2 Frequency Range Standard Option 051 Option 061a or 062 Option 051 0.07 to 0.3 MHz 85 83 81 114 Option 061a or 062 112 0.
Receiver Dynamic Range VectorStar™ MS4640B Series Receiver Dynamic Range Calculated as the difference between the maximum receiver input level for 0.1 dB compression and the specified noise floor at the specified reference plane. Characteristic Performance. All Models, Receiver Dynamic Range (dB) at b1 or b2 at Ports 1 or 2 Standard Frequency Range Option 051 Option 061a or 062 Option 061a or 062 Option 051 0.07 to 0.3 MHz 80 79 78 90 89 0.
VectorStar™ MS4640B Series High Level Noise High Level Noise Measured at 1 kHz IF bandwidth, at default power, with either full reflects or through transmission. RMS. Characteristic performance on MS4645B and MS4647B with either Option 051, 061, or 062. High level noise magnitude may be degraded to 20 mdB RMS (typical) at particular frequencies due to receiver residuals. Frequency (GHz) Magnitude (dB) Phase (degree) 70 kHz to 500 kHz < 0.04 < 0.4 500 kHz to 2.5 < 0.0045 < 0.05 2.5 to 5 < 0.
Noise Floor VectorStar™ MS4640B Series MS4647B Example Noise Floor (standard configuration) 0.0 -20.0 Mag (dBrms) -40.0 -60.0 -80.0 -100.0 -120.0 -140.0 0.0 10.0 20.0 30.0 40.0 50.0 60.0 70.0 Frequency (GHz) Example System Dynamic Range (as measured, not RMS) 10 Hz IFBW, No Averaging, at Max Power Ports Terminated after Transmission Cal Example System Dynamic Range 6 PN: 11410-00611 Rev.
VectorStar™ MS4640B Series Power Range Power Range Maximum Rated Power to minimum level. The difference reflects the ALC range for standard models or with Option 051, and the ALC + Attenuator Range for models with Options 061 or 062. Maximum Rated Power is typical from 2.4 - 2.7 GHz. MS4642B, 20 GHz Model, Power Range (dBm) Frequency Standard Option 051 Option 061a or 062 70 kHz to 0.01 GHz +10 to –25 +9 to –25 +8 to –95 0.01 to 2.5 GHz +12 to –25 +11 to –25 +10 to –95 2.
Power Range VectorStar™ MS4640B Series MS4642B Example Maximum Power (standard configuration) 25.0 Option 031 20.0 Max Output Power (dBm) 15.0 Non-Option 031 10.0 5.0 0.0 -5.0 -10.0 0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 16.0 18.0 20.0 Frequency (GHz) MS4644B Example Maximum Power (standard configuration) 25.0 20.0 Option 031 Max Output Power (dBm) 15.0 Non-Option 031 10.0 5.0 0.0 -5.0 -10.0 0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.
VectorStar™ MS4640B Series Output Default Power Output Default Power Instrument default power. For maximum rated power, refer to “Power Range” above. Model Standard (No Options) Option 051, 061 or 062 MS4642B, 20 GHz +5 dBm MS4644B, 40 GHz +5 dBm +5 dBm +5 dBm MS4645B, 50 GHz –3 dBm –10 dBm MS4647B, 70 GHz –3 dBm –10 dBm Power Accuracy, Linearity, and Resolution Frequency (GHz) Accuracya (dB) Linearityb (dB) Resolution (dB) 70 kHz to 0.01 ± 1.5 ± 1.5 0.01 0.01 to 40 ± 1.5 ± 1.0 0.
MS4642B 20 GHz VNA System Performance VectorStar™ MS4640B Series MS4642B 20 GHz VNA System Performance MS4642B – 12-Term SOLT – Sliding Load – 3652A-1 K Calibration Kit MS4642B 20 GHz Model, with 12-term SOLT with Sliding Load Calibration, using the 3652A-1 K Calibration Kit. Frequency Range Directivity Source Match Load Matcha Reflection Tracking (dB) (GHz) (dB) (dB) (dB) Transmission Tracking (dB) 70 kHz to 0.01 > 38 > 36 > 38 ± 0.02 0.01 to 2.5 > 42 > 41 > 42 ± 0.005 ± 0.05 ± 0.03 2.
VectorStar™ MS4640B Series MS4642B 20 GHz VNA System Performance MS4642B – 12-Term SOLT – 3652A or 3652A-1 K Calibration Kit MS4642B 20 GHz Model, with 12-term SOLT Calibration, using 3652A K or 3652A-1 K Cal Kit. Frequency Range Directivity Source Match Load Match a Reflection Tracking (dB) (GHz) (dB) (dB) (dB) Transmission Tracking (dB) 70 kHz to 0.01 > 38 > 36 > 38 ± 0.02 ± 0.05 0.01 to 2.5 > 37 > 41 > 37 ± 0.005 ± 0.03 2.5 to 20 > 34 > 39 > 35 ± 0.006 ± 0.07 a.
MS4642B 20 GHz VNA System Performance VectorStar™ MS4640B Series MS4642B – 12-Term SOLT – Sliding Load – 3650A-1 3.5 mm Calibration Kit MS4642B 20 GHz Model, with 12-term SOLT Calibration with Sliding Load Calibration, using the 3650A-1 3.5 mm Cal Kit. Reflection Tracking Transmission Tracking Frequency Range Directivity Source Match Load Matcha (dB) (dB) (dB) (GHz) (dB) (dB) 70 kHz to 0.01 > 40 > 37 > 40 ± 0.02 ± 0.05 0.01 to 2.5 > 42 > 41 > 42 ± 0.005 ± 0.03 2.5 to 10 > 43 > 39 > 43 ± 0.
VectorStar™ MS4640B Series MS4642B 20 GHz VNA System Performance MS4642B – 12-Term SOLT – 3650A or 3650A-1 3.5 mm Calibration Kit MS4642B 20 GHz Model, with 12-term SOLT Calibration, using the 3650A or 3650A-1 3.5 mm Cal Kit. Frequency Range Directivity Source Match Load Matcha Reflection Tracking (dB) (GHz) (dB) (dB) (dB) Transmission Tracking (dB) 70 kHz to 0.01 > 40 > 37 > 40 ± 0.02 0.01 to 2.5 > 42 > 40 > 42 ± 0.005 ± 0.05 ± 0.03 2.5 to 10 > 40 > 34 > 40 ± 0.005 ± 0.
MS4642B 20 GHz VNA System Performance VectorStar™ MS4640B Series MS4642B – 12-Term – 36585K K AutoCal™ MS4642B 20 GHz Model, with 12-term Calibration, using the 36585K K Automatic Calibrator (AutoCal) Frequency Range Directivity Source Match Load Matcha Reflection Tracking (dB) (GHz) (dB) (dB) (dB) Transmission Tracking (dB) 70 kHz to 0.01 b > 40 > 40 > 43 ± 0.10 ± 0.10 0.01 to 2.5 > 43 > 47 > 43 ± 0.05 ± 0.03 2.5 to 20 > 50 > 47 > 50 ± 0.09 ± 0.03 a.
VectorStar™ MS4640B Series MS4644B 40 GHz VNA System Performance MS4644B 40 GHz VNA System Performance MS4644B – 12-Term SOLT – Sliding Load – 3652A-1 K Calibration Kit MS4644B 40 GHz Model, with 12-term SOLT with Sliding Load Calibration, using the 3652A-1 K Calibration Kit. Frequency Range (GHz) Directivity (dB) Load Matcha (dB) Source Match (dB) Transmission Tracking (dB) Reflection Tracking (dB) 70 kHz to 0.01 > 38 > 36 > 38 ± 0.02 ± 0.05 0.01 to 2.5 > 42 > 41 > 42 ± 0.005 ± 0.03 2.
MS4644B 40 GHz VNA System Performance VectorStar™ MS4640B Series MS4644B – 12-Term SOLT – 3652A or 3652A-1 K Calibration Kit MS4644B 40 GHz Model, with 12-term SOLT Calibration, using the 3652A or 3652A-1 K Calibration Kit. Reflection Tracking Frequency Range Directivity Source Match Load Matcha (dB) (dB) (GHz) (dB) (dB) Transmission Tracking (dB) 70 kHz to 0.01 > 38 > 36 > 38 ± 0.02 ± 0.05 0.01 to 2.5 > 37 > 41 > 37 ± 0.005 ± 0.03 2.5 to 20 > 34 > 39 > 35 ± 0.006 ± 0.
VectorStar™ MS4640B Series MS4644B 40 GHz VNA System Performance MS4644B – 12-Term – 36585K K AutoCal MS4644B 40 GHz Model, with 12-term Calibration, using the 36585K K AutoCal. Frequency Range Directivity Source Match Load Matcha (dB) (GHz) (dB) (dB) Reflection Tracking (dB) Transmission Tracking (dB) 70 kHz to 0.01 b > 40 > 40 > 43 ± 0.10 ± 0.10 0.01 to 2.5 > 43 > 47 > 43 ± 0.05 ± 0.03 2.5 to 20 > 50 > 47 > 50 ± 0.09 ± 0.03 20 to 40 > 48 > 47 > 48 ± 0.14 ± 0.07 a.
MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance VectorStar™ MS4640B Series MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance MS4645B/MS4647B VNAs – 12-Term SOLT Sliding Load – 3654D-1 V Calibration Kit Frequency Range (GHz) Directivity (dB) Source Match (dB) Load Matcha (dB) Reflection Tracking (dB) Transmission Tracking (dB) 70 kHz to 0.01 > 38 > 36 > 38 ± 0.02 ± 0.05 0.01 to 2.5 > 41 > 39 > 41 ± 0.02 ± 0.05 2.5 to 20 > 41 > 37 > 41 ± 0.02 ± 0.
VectorStar™ MS4640B Series MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance MS4645B/MS4647B Measurement Uncertainties (Reflection) The graphs give measurement uncertainties after the above calibration. The errors are a worst-case contribution of residual directivity, load and source match, frequency response and isolation, network analyzer dynamic accuracy, and connector repeatability. 10 Hz IF Bandwidth is used. For transmission uncertainties, it is assumed that S11 = S22 = 0.
MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance VectorStar™ MS4640B Series MS4645B/MS4647B VNAs – 12-Term SOLT – 3654D or 3654D-1 V Calibration Kit Frequency Range (GHz) Directivity (dB) Source Match (dB) Load Matcha (dB) Reflection Tracking (dB) Transmission Tracking (dB) 70 kHz to 0.01 > 38 > 36 > 38 ± 0.02 ± 0.05 0.01 to 2.5 > 40 > 39 > 40 ± 0.02 ± 0.05 2.5 to 20 > 40 > 37 > 40 ± 0.02 ± 0.07 20 to 40 > 35 > 32 > 35 ± 0.02 ± 0.08 40 to 65 > 32 > 28 > 32 ± 0.
VectorStar™ MS4640B Series MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance MS4645B/MS4647B Measurement Uncertainties (Reflection) The graphs give measurement uncertainties after the above calibration. The errors are a worst-case contribution of residual directivity, load and source match, frequency response and isolation, network analyzer dynamic accuracy, and connector repeatability. 10 Hz IF Bandwidth is used. For transmission uncertainties, it is assumed that S11 = S22 = 0.
MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance VectorStar™ MS4640B Series MS4645B/MS4647B VNAs – LRL – 3657-1 V Multi-Line Calibration Kit MS4645B 50 GHz and MS4647B 70 GHz VNAs, with an LRL Calibration, using the 3657-1 V Multi-Line Calibration Kit, with symmetric reflects. Frequency Range Directivity Source Match Load Match a Reflection Tracking Transmission Tracking (dB) (GHz) (dB) (dB) (dB) (dB) 0.24 b to 2.5 > 50 > 50 > 50 ± 0.005 2.5 to 20 > 50 > 50 > 50 ± 0.005 ± 0.
VectorStar™ MS4640B Series MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance MS4645B/MS4647B Measurement Uncertainties (Reflection) The graphs give measurement uncertainties after the above calibration. The errors are a worst-case contribution of residual directivity, load and source match, frequency response and isolation, network analyzer dynamic accuracy, and connector repeatability. 10 Hz IF Bandwidth is used. For transmission uncertainties, it is assumed that S11 = S22 = 0.
MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance VectorStar™ MS4640B Series MS4645B/MS4647B VNAs – 12-Term – 36585V V AutoCal MS4645B 50 GHz and MS4647B 70 GHz VNAs, with 12-term Calibration, using the 36585V V AutoCal. Reflection Tracking Frequency Range Directivity Source Match Load Match a (dB) (dB) (GHz) (dB) (dB) Transmission Tracking (dB) 70 kHz to 0.01 b > 40 > 40 > 40 ± 0.10 ± 0.10 0.01 to 2.5 > 43 > 47 > 43 ± 0.05 ± 0.03 2.5 to 20 > 50 > 47 > 50 ± 0.09 ± 0.
VectorStar™ MS4640B Series MS4645B 50 GHz / MS4647B 70 GHz VNA System Performance MS4645B/MS4647B Measurement Uncertainties (Reflection) The graphs give measurement uncertainties after the above calibration. The errors are a worst-case contribution of residual directivity, load and source match, frequency response and isolation, network analyzer dynamic accuracy, and connector repeatability. 10 Hz IF Bandwidth is used. For transmission uncertainties, it is assumed that S11 = S22 = 0.
Measurement Times VectorStar™ MS4640B Series Measurement Times Measurement times include sweep time, and band-switching time, in single channel mode. Typical.
VectorStar™ MS4640B Series Standard Capabilities Standard Capabilities Operating Frequency MS4642B 10 MHz to 20.2 GHz MS4644B 10 MHz to 40.5 GHz MS4645B 10 MHz to 50.5 GHz MS4647B 10 MHz to 70 GHz MS4640B-070 Optional for all MS4640B Series VNAs. Provides 40 kHz to 10 MHz Coverage Extension. Provides a lower limit specified to 70 kHz, which is allowed to extend to 40 kHz.
Standard Capabilities VectorStar™ MS4640B Series Channels, Display, and Traces Channels and Traces Display Display Colors Trace Memory and Math Intra-trace Math Scale Resolution 16 channels, each with up to 16 traces Color touch screen LCD, 26.4 cm (10.4”) diagonal Unlimited colors for data traces, memory, text, markers, graticules and limit lines.
VectorStar™ MS4640B Series Calibration and Correction Capabilities Calibration and Correction Capabilities Calibration Methods Correction Models Short-Open-Load-Through (SOLT) with Fixed or Sliding Load Offset-Short-Offset-Short-Load-Through (SSLT) with Fixed or Sliding Load Triple-Offset-Short-Through (SSST) Short-Open-Load-Reciprocal (SOLR) or Unknown Through Method (SSLR, SSSR) Line-Reflect-Line (LRL) / Line-Reflect-Match (LRM) Advanced-LRM (A-LRM™) for improved on-wafer calibrations AutoCal Thru Upd
Remote Operability VectorStar™ MS4640B Series Remote Operability VectorStar supports several remote operability options. Communication Type Data Format Via GPIB Via LAN Via USB Drivers for GPIB, LAN, or USB Performance Using IEEE 488.2 1 MB/s Data Transfer Speed Using VXI-11 Protocol 2.5 MB/s Data Transfer Speed Using USBTMC Protocol 5.5 MB/s Data Transfer Speed Description Use SCPI or previous generation Lightning VNA commands.
VectorStar™ MS4640B Series Optional Capabilities Optional Capabilities Time Domain Measurements – Option 002 Displays all S-parameters and overlays with Frequency Domain, Low-pass Mode with added harmonics frequency list flexibility, Band-pass Mode, Phasor Impulse Mode, Windowing, Gating (pass-band or reject-band), and Frequency with Time Gate.
Optional Capabilities VectorStar™ MS4640B Series IF Digitizer — Option 035 Description Required Options Compatible Options Incompatible Options When combined with Option 042 PulseView™, adds the capability to generate and measure pulsed signals. Four internal signal generators are included enabling singlet, doublet, triplet, quadruplet, and/or burst signal generation. Pulse measurements include pulse profile, point-in-pulse, and pulse-to-pulse capability.
VectorStar™ MS4640B Series Optional Capabilities PulseView™ — Option 042 Description Required Options Compatible Options Incompatible Options When combined with Option 035 IF Digitizer, adds the capability to generate and measure pulsed signals. Four internal signal generators are included enabling singlet, doublet, triplet, quadruplet, and/or burst signal generation. Pulse measurements include pulse profile, point-in-pulse, and pulse-to-pulse capability.
Optional Capabilities VectorStar™ MS4640B Series DifferentialView™ — Option 043 Description Required Options Compatible Options Incompatible Options Multiport Systems When combined with Option 031 Dual Source Architecture, provides dual source control and calibrations required for stimulating and measuring differential devices. Allows true differential and common mode device drives. Corrects mismatch introduced error of the DUT to VNA interface via real and time calibration.
VectorStar™ MS4640B Series CPU, Memory, and Security Features CPU, Memory, and Security Features CPU Intel Core™ i5 Display 26.4 cm (10.4”) Color XGA Touch-Screen Display Storage Serial-ATA (SATA) Solid State Drive (SSD), for OS, Programs, and Data. (> 30 GB) Security Features Display Blanking Removable Internal Drive Option 004 Spare SSD Virus Protection, Best Practices VectorStar™ software can obscure frequency and power levels on the system display for security.
Rear Panel Connections VectorStar™ MS4640B Series Rear Panel Connections MS4640B Series Rear Panel (with Option 035) AC Power Input AC Input connector, with On/Off switch, and fuses 350 VA maximum, 90 to 264 VAC, 47 to 63 Hz (power factor controlled) USB, PS/2, and LAN USB Control Port USB Ports Type B USB 2.0 port for controlling the instrument externally, for remote operation Two Type A USB 2.0 Ports for peripherals such as keyboard, mouse, memory stick, hardware key, etc.
VectorStar™ MS4640B Series Rear Panel Connections Bias Inputs Optional Requires Active Measurement Suite, Option 061 or 062 Bias Inputs BNC (female), one per port Bias Fuses 0.5 A, 250 V, one per port Direct Access Loops Optional Connector Type Description Damage Input Levels IF Inputs/Outputs Requires Option 051, 062, or 062 SMA (female) For Source, Test, and Receive paths, 3 per port, for < 2.
Rear Panel Connections VectorStar™ MS4640B Series Trigger Out Connector Type BNC (female) Voltage Output 0 to 3.3 V pulse output 1 µs positive pulse Voltage Impedance V(output high) = 2 V min @ –12 mA V(output low) = 0.8 V max @ +12 mA Low impedance (approximately 50 Ω) Pulse Generator Outputs All values listed are nominal. Optional Connector Type Pulse Generator Outputs Voltage Requires Options 035 and 042 PulseView™ SMA (female) P GEN 1, P GEN 2, P GEN 3, and P GEN 4 High: 3.
VectorStar™ MS4640B Series Mechanical and Environmental Mechanical and Environmental Dimensions Dimensions listed are for the instrument without rack mount option (MS4640B-001) attached.
Block Diagram VectorStar™ MS4640B Series Block Diagram > 2.5 GHz High Band A Optional < 2.5 GHz Low Band a1 A a1 a2 a2 B b1 b2 b1 Port 1 b2 Bias 1 Bias 2 Port 2 MS4640B Series VNA Block Diagram – Fully Loaded Configuration A. With Option 031 Dual Source Architecture, second low-band and high-band sources are added and the two switches are removed. One set of sources is dedicated to each of the VNA test port paths. B.
VectorStar™ MS4640B Series 36585-Series Automatic Calibrators (AutoCal) 36585-Series Automatic Calibrators (AutoCal) The 36585-Series Precision Automatic Calibrator (AutoCal) Module provides industry-leading performance in corrected characteristics using over-determined algorithms, and transferring characteristics from a highly accurate LRL type calibration. The resulting accuracies will even out perform a Sliding Load SOLT calibration.
Mechanical Calibration Kits VectorStar™ MS4640B Series Mechanical Calibration Kits SMA/3.5 mm Calibration Kit, 3650A Series 3650A cal kit provides 50 Ω calibrations for 3.5 mm or SMA devices using 3.5 mm standards. 3650A-1 cal kit includes Sliding Loads. 3650A Cal Kit contains: Quantity Part Number Return Loss: > 37 dB (F ≤ 18.5 GHz) > 30 dB (F > 18.5 GHz) 2 28S50-2 2 28SF50-2 Open 3.5 mm (male) Offset: 5 mm 1 24S50 Open 3.5 mm (female) Offset: 5 mm 1 24SF50 Short 3.
VectorStar™ MS4640B Series Mechanical Calibration Kits V (1.85 mm) Calibration Kit, 3654D Series 3654D cal kit provides 50 Ω calibrations for V devices. 3654D-1 cal kit includes Sliding Loads. 3654D Cal Kit contains: Termination V (male) Additional Information (typical) Quantity Part Number Return Loss: > 40 dB (F ≤ 20 GHz); > 35 dB (F ≤ 40 GHz) > 32 dB (F ≤ 67 GHz); > 28 dB (F ≤ 70 GHz) 2 28V50D 2 28VF50D Open V (male) Offset: 4.75 mm 1 24V50C Open V (female) Offset: 4.
Mechanical Calibration Kits VectorStar™ MS4640B Series V (1.85 mm) Multi-Line Calibration Kit, 3657 Series The 3657 Calibration Kit provides 50 Ω beadless V (male to male) lines for metrology applications. The 3657-1 Calibration Kit includes Shorts for LRL-type coaxial calibrations.
VectorStar™ MS4640B Series Verification Kits Verification Kits Verification kits include characterized traceable standards (two attenuators, an airline, and a stepped impedance airline Beatty Standard) that can be used with the provided Performance Verification Software (PVS) and data to verify the calibration and resulting performance of your VNA. The applicable calibrations are Short-Open-Load-Through (SOLT) with and without Sliding Loads for the 3666-1, 3668-1, and 3669B-1 Verification Kits.
Test Port Cables VectorStar™ MS4640B Series Test Port Cables 3670-Series Test Port Cables, Ruggedized Semi-Rigid, up to 70 GHz Description K (female) to K (male) V (female) to V (male) Frequency Range Nominal Impedance DC to 40 GHz DC to 70 GHz Insertion Loss (dB, typical) Return Loss (dB, typical) 50 Ω 2.3 dB/m @ 20 GHz 4.7 dB/m @ 40 GHz 50 Ω 3.6 dB/m @ 20 GHz 5.2 dB/m @ 40 GHz 7.2 dB/m @ 70 GHz 70 GHz Phase Stable Flexible Test Port Cables, 3671-Series ≥ 16 ≥ 16 Length Part Number 30.
VectorStar™ MS4640B Series Universal Test Fixture (UTF) Universal Test Fixture (UTF) The 3680-series UTF provide an accurate, repeatable solution for measuring microstrip and coplanar substrate devices. • Input and output connections are made to the substrate device by two spring-loaded jaws that include coax-to-microstrip/coplanar launchers. • One jaw is movable in two dimensions to accommodate substrates of different lengths and offsets. • Right angle launchers are available for right angle devices.
Ordering Information VectorStar™ MS4640B Series Ordering Information Instrument Models The VectorStar MS4640B Series VNAs are available in four models to meet different frequency range requirements. Refer to “Standard Capabilities” on page 2-27 for extended operational frequency ranges.
VectorStar™ MS4640B Series Ordering Information Broadband/Millimeter-Wave Options Provides the MS4647B VNA with the capability to connect with the ME7838A Modular Broadband/Millimeter-Wave VNA System. Option 007 Receiver Offset (described above) is also required. MS4647B-080 Modular Broadband/Millimeter-Wave Capability. For MS4647B VNAs without Option 051, 061, or 062. If equipped on a MS4647B VNA, Option 082 below cannot be added. MS4647B-081 Modular Broadband/Millimeter-Wave Capability.
Ordering Information VectorStar™ MS4640B Series Test Port Cables, Flexible, Ruggedized-Style Female Connectors, Phase Stable Ruggedized style female connectors for VNA test ports. 3671KFS50-60 K (female) to 3.5 mm (male), 2 each 63.5 cm (25 in) 3671KFK50-60 K (female) to K (male), 2 each, 63.5 cm (25 in) 3671KFK50-100 K (female) to K (male), 1 each, 96.5 cm (38 in) 3671KFKF50-60 and 3671KFK50-60 3671VFV50-60 3671VFV50-100 K (female) to K (male), 1 each, 63.
VectorStar™ MS4640B Series Extended Service Options Extended Service Options Use the table below to select the service location, service period, type of service, and the VectorStar instrument model number.
Find Drivers, Utilities, Software Updates, and other Helpful Tools at the VectorStar Users Site. Visit: www.anritsu.com/en-us/Products-Solutions/Solution/Welcome-to-the-VectorStar-Users-Site-.aspx To receive a quote to purchase a product or order accessories visit our online ordering site: www.ShopAnritsu.com Training at Anritsu Anritsu has designed courses to help you stay up to date with technologies important to your job. For available training courses visit: www.anritsu.