Manual
Page 5 of 20
Spectrum Master™ MS2723B Spectrum Analyzer Specifications
Interference Analyzer (Option 0025)
Measurements Spectrum
Field Strength
Occupied Bandwidth
Channel Power
Adjacent Channel Power (ACPR)
AM/FM/SSB Demodulation (Wide/Narrow FM, Upper/Lower SSB), (audio out only)
Carrier-to-Interference ratio (C/I)
Spectrogram (Collect data up to one week)
Signal Strength (Gives visual and aural indication of signal strength)
Received Signal Strength Indicator (RSSI) (collect data up to one week)
Gives visual and aural indication of signal strength
Signal ID (up to 12 signals)
Center Frequency
Bandwidth
Signal Type (FM, GSM, W-CDMA, CDMA, Wi-Fi)
Closest Channel Number
Number of Carriers
Application Options Signal-to-Nose Ratio (SNR) > 10 dB
Impedance (50 Ω, 75 Ω, Other)
Channel Scanner (Option 0027)
Number of Channels 1 to 20 Channels (Power Levels)
Measurements Graph/Table, Max Hold (On/5 sec/Off), Frequency/Channel, Current/Maximum, Dual Color
Scanner Scan Channels, Scan Frequencies, Scan Customer List, Scan Script Master™
Amplitude Reference Level, Scale
Custom Scan Signal Standard, Channel, # of Channels, Channel Step Size, Custom Scan
Application Options Impedance (50 Ω, 75 Ω, Other)
Gated Sweep (Option 0090)
Mode Spectrum Analyzer, Sweep
Trigger External TTL
Setup Gated Sweep (On/Off)
Gate Polarity (Rising, Falling)
Gate Delay (0 to 65 ms typical)
Gate Length (1 µs to 65 ms typical)
Zero Span Time
Zero Span IF Output (Option 0089)
Mode Spectrum Analyzer/Span/Zero Span
Center Frequency 37.8 MHz
IF Bandwidth 7 MHz, 10 MHz, 16 MHz, typical
Output Power Level –20 dBm to –45 dBm Typical given:
RF Input = +30 dBm to –43 dBm with Preamp OFF
–40 dBm to –60 dBm with Preamp ON
Reference Level set at RF Input Level
Auto RF Attenuation
RF Attenuation Auto
Connector BNC female