Agilent E4412A and E4413A Power Sensor Operating and Service Guide Agilent Technologies
Notices © Agilent Technologies, Inc. 1999, 2009 Warranty No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions.
Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standard and Technology (formerly National Bureau of Standards), to the extent allowed by that organization’s calibration facility, and to the calibration facilities of other International Standards Organization members.
Limitation of Warranty The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by the Buyer, Buyer-supplied products or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance. The design and implementation of any circuit on this product is the sole responsibility of the Buyer.
Restricted Rights Legend The Software and Documentation have been developed entirely at private expense. They are delivered and licensed as “commercial computer software” as defined in DFARS 252.227-7013 (Oct 1988), DFARS 252.211-7015 (May 1991), or DFARS 252.227-7014 (Jun 1995), as a “commercial item” as defined in FAR 2.101(a), or as “restricted computer software” as defined in FAR 52.227-19 (Jun 1987) (or any equivalent agency regulation or contract clause), whichever is applicable.
Safety Summary The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies, Inc. assumes no liability for the customer’s failure to comply with these requirements. Safety Notices vi WARN IN G A WARNING notice denotes a hazard.
Safety Symbols The following symbol on the instrument and in the documentation indicates precautions that must be taken to maintain safe operation of the instrument. Caution, risk of danger. The Instruction Documentation Symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the supplied documentation.
PROTECTIVE CONDUCTOR TERMINAL. Frame or chasis TERMINAL. Equipotentiality. On (Supply). Off (Supply). Equipment protected throughout by DOUBLE INSULATION or REINFORCED INSULATION. Caution, risk of electric shock. Caution, hot surface. In position of bi-stable push control.
Out position of bi-stable push control.
Regulatory Markings The CE mark shows that the product complies with all the relevant European legal Directives (if accompanied by a year, it signifies when the design was proven). ISM 1- A N10149 The C-tick mark is a registered trademark of the Spectrum management Agency of Australia. This signifies compliance with the Australian EMC Framework regulations under the terms of the Radio Communications Act of 1992. This product complies with the WEEE Directive (2002/96/EC) marking equipment.
General Safety Information The following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements.
In This Guide ... 1 Operation and Service Guide This chapter contains information about initial inspection, performance tests, specifications, operations, troubleshooting and service of the Agilent E4412A and E4413A power sensors. 2 Appendix This Appendix contains specifications that applies to E4412A and E4413A sensors (formerly EXCP-E18A and ECP-E26A, respectively) with a serial prefix below US3848/MYxxxxxxxx/SGxxxxxxxx.
Contents Notices ii General Warranty iii Warranty Service iii Limitation of Warranty iv Exclusive Remedies iv Technology Licenses v Safety Summary vi Safety Notices vi Safety Symbols vii Regulatory Markings x General Safety Information xi In This Guide ...
Cleaning 21 Disassembly Procedure 22 Reassembly Procedure 23 Agilent Sales and Service Offices 24 A Appendix E4412A and E4413A Specifications (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-4 E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-5 xiv
List of Figures Figure 1 E4412A and E4413A Power Sensors (Formerly ECP-E18A and EXCP-E26A, respectively) 2 Figure 2 Relative Mode Power Measurement Linearity with Power Meter/Sensor at 25 ºC (Typical) 7 Figure 3 N8485A power sensor with adapter 14 Figure 4 Illustrated Parts Break down 18 Figure 5 Removing Power Sensor Shell 22 xv
List of Tables Table 1 E4412A and E4413A Specifications A-5 Table 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) A-9 Table 3 E4413A Calibration Factor Uncertainty at 1 mW (0 dBm) A-10 Table 4 Power Sensor SWR and Reflection Coefficient for the E4412Aa A-15 Table 5 Power Sensor SWR and Reflection Coef ficient for the E4413Aa A-16 Table 5 Power Sensor SWR and Reflection Coef ficient for the E4413Aa A-16 Table 6 Replaceable Parts A-19 Table 7 Agilent Sales and Service Offices A-24 Table A-1 E4412A and
Agilent E4412A and E4413A Power Sensors Operating and Service Guide 1 Operation and Service Guide General Information 3 Performance Test 15 Replaceable Parts 17 Service 20 This chapter contains information about initial inspection, performance tests, specifications, operations, troubleshooting and service of the Agilent E4412A and E4413A power sensors.
1 Introduction Figure 1 2 E4412A and E4413A Power Sensors (Formerly ECP-E18A and EXCP-E26A, respectively) Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Introduction 1 General Information Warranty The power sensors are warranted and certified as indicated on the inside cover of this manual. Instruments Covered by Manual These instruments have a two- part serial number: the prefix (two letters and the first four numbers), and the suffix (the last four numbers). The two letters identify the country in which the unit was manufactured.
1 Introduction These power sensors measure CW power, displayed on a compatible power meter in logarithmic (dBm or dB) or linear (Watts or %) measurement units. The E4413A is shipped with a 3.5- mm to Type- N adapter, part number 08485- 60005. N O TE N O TE The E4412A and E4413A power sensors are compatible ONLY with the newer E44XX-Series power meters. They are NOT compatible with the earlier 430-Series, E1416A, or 70100A power meters.
Introduction Table 1 1 E4412A and E4413A Specifications Limit Frequency Range E4412A: 10 MHz to 18 GHz E4413A: 50 MHz to 26.5 GHz Power Range –70 dBm to +20 dBm (100 pW to 100 mW) Impedance 50 ohm Connector Type E4412A: Type-N (male) Comments nominal E4413A: 3.5 mm (male) SWR Rho 10 MHz to <30 MHz 1.22 30 MHz to <2 GHz 1.15 0.099 0.070 20.08 23.13 2 GHz to <6 GHz 1.17 0.078 22.12 6 GHz to <11 GHz 1.20 0.091 20.82 1.27 0.119 18.42 50 MHz to <100 MHz 1.21 0.
1 Introduction Limit Comments Power Linearityab 25 ± 5°C 0 to 55°C (After calibration at 0 dBm at ambient temperature) 100 pW to 10 mW ±3% ±7% 10 mW to 100 mW ±4.5% ±10% –70 dBm to +10 dBm +10 dBm to +20 dBm Operating Temperature 0 to 55°C Net Weight E4412A 0.47 kg (1.04 lb) E4413A 0.45 kg (1.00 lb) Dimensions E4412A Length: 130 mm (5.1 in) E4413A Length: 102 mm (4.0 in) Width: 38 mm (1.5 in) Same for both models Height: 30 mm (1.2 in) Same for both models a.
Power Level Being Measured (dBm) Introduction +20 +10 +6 0 1 A ±6 % ±3 % –20 B –35 ±1 % –70 –70 –35 –25–20 0 +20 Power Level Used as Reference (dBm) Figure 2 Relative Mode Power Measurement Linearity with Power Meter/Sensor at 25 ºC (Typical) The chart in Figure 2 shows the typical uncertainty in making relative power measurements with the E44XX- Series power meter in the Rel (relative) mode. In a relative measurement, two power levels are compared using a single sensor.
1 Introduction EXAMPLE A (Amplifier Gain Measurement): Input power (reference) = –35 dBm; power measured = +6 dBm’ so the relative gain is +41 dB. the chart indicates a typical error of ±6 %, which corresponds to +0.25 dB/–0.27 dB uncertainty in the gain measurement. EXAMPLE B (Insertion Loss Measurement): Reference power = –25 dBm; power measured = –35 dBm, for the case of an insertion loss of 10 dB. the chart indicates a typical error of ±1 %, which corresponds to ±0.
Introduction Table 2 1 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) Frequency Uncertainty (%)a 10 MHz 1.8 30 MHz 1.8 50 MHz Reference 100 MHz 1.8 1.0 GHz 1.8 2.0 GHz 2.4 4.0 GHz 2.4 6.0 GHz 2.4 8.0 GHz 2.4 10.0 GHz 2.4 11.0 GHz 2.4 12.0 GHz 2.4 14.0 GHz 2.4 16.0 GHz 2.6 18.0 GHz 2.6 a. Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above. For earlier instruments, refer to Appendix A.
1 Introduction Table 3 E4413A Calibration Factor Uncertainty at 1 mW (0 dBm) Frequency Uncertainty (%)a 50 MHz Reference 100 MHz 1.8 1.0 GHz 1.8 2.0 GHz 2.4 4.0 GHz 2.4 6.0 GHz 2.4 8.0 GHz 2.4 10.0 GHz 2.6 11.0 GHz 2.6 12.0 GHz 2.8 14.0 GHz 2.8 16.0 GHz 2.8 17.0 GHz 2.8 18.0 GHz 2.8 19.0 GHz 3.0 20.0 GHz 3.0 22.0 GHz 3.0 24.0 GHz 3.0 26.0 GHz 3.0 26.5GHz 3.0 a. Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above.
Introduction 1 Installation Initial Inspection Inspect the shipping container for damage. If the shipping container or packaging material is damaged, it should be kept until the contents of the shipment have been checked mechanically and electrically. If there is mechanical damage or if the instrument does not pass the performance tests, notify the nearest Agilent Technologies office. Keep the damaged shipping materials (if any) for inspection by the carrier and a Agilent Technologies representative.
1 Introduction Storage and Shipment Environment The instruments should be stored in a clean, dry environment. The following limitations apply to both storage and shipment: Temperature –55 to +75 ºC Relative Humidity <95 % at 40 ºC Altitude <15,240 metres (50,000 feet) Original Packaging Containers and materials identical to those used in factory packaging are available through Agilent Technologies offices.
Introduction 1 Operation Precautions WARN IN G BEFORE CONNECTING THE POWER SENSOR TO OTHER INSTRUMENTS, ensure that all instruments are connected to the protective (earth) ground. Any interruption of the protective earth grounding will cause a potential shock hazard that could result in personal injury and cause damage to the power sensor. If the following energy and power levels are exceeded, the power meter system may be damaged.
1 Introduction Power Meter Calibrations Follow the calibration procedures given in your power meter manual. The E4413A power sensor is fitted with 3.5 mm (m) connectors as standard. To convert the 3.5 mm (m) connector for calibration, an adapter (3.5 mm (f) to Type- N (m)) is included with the power sensor. See Figure 3. Figure 3 N O TE E4413A power sensor with adapter The 3.5 mm to Type-N adapter is intended for the use of 1 mW, 50 MHz power reference of the power meter only.
Introduction 1 Performance Test Standing Wave Ratio (SWR) and Reflection Coefficient (Rho)Performance Test This section does not establish preset SWR test procedures since there are several test methods and different equipm ent available for testing the SWR or reflection coefficient. Therefore, the actual accuracy of the test equipment must be accounted for when measuring against instrument specifications to determine a pass or fail condition.
1 Introduction Table 5 Power Sensor SWR and Reflection Coef ficient for the E4413Aa Frequency System Rho Uncertainty Actual Measurement Maximum Rho 50 MHz to < 100 MHz ± 0.010 0.095 100 MHz to < 8GHz ± 0.010 0.087 8 GHz to < 18MHz ± 0.010 0.095 18 GHz to < 26.5 GHz ± 0.015 0.115 a.Specifications apply to instruments with serial prefix US3848/MYxxxxxxxx/SGxxxxxxxx and above.
Introduction 1 Replaceable Parts Table 6 is a list of replaceable parts. Figure 4 is the illustrated parts breakdown (IPB) that identifies all of the replaceable parts. To order a part, quote the Agilent Technologies part number, specify the quantity required, and address the order to the nearest Agilent Technologies office. N O TE Within the USA, it is better to order directly from the Agilent Parts Center in Roseville, California.
1 Introduction Figure 4 18 Illustrated Parts Break down Agilent E4412A and E4413A Power Sensors Operating and Service Guide
Introduction Table 6 1 Replaceable Parts Reference Designation Part Number Qty Description A1/A2 E4412A E4412-60006 1 SENSOR MODULE (E4412A) A1/A2 E4412A E4412-69006 1 RESTORED SENSOR MODULE (E4412A) A1/A2 E4413A E4413-60003 1 SENSOR MODULE (E4413A) A1/A2 E4413A E4413-69003 1 RESTORED SENSOR MODULE (E4413A) 08485-60005 1 ADAPTOR, 3.
1 Introduction Service Service instructions consist of principles of operation, troubleshooting, and repairs. Principles of Operation The A1 Bulkhead assembly provides a 50 ohm load to the RF signal applied to the power sensor. A diode assembly in the bulkhead rectifies the applied RF to produce a dc voltage which varies with the RF power across the 50 ohm load. Thus the voltage varies with the RF power dissipated in the load. With maximum specified RF power (100 mW) the dc voltage is approximately 1V.
Introduction 1 If error message 241 or 310 is indicated on the power meter, suspect failed power sensor. If no error message is displayed, but a problem occurs when making a measurement, try replacing the cable from the power meter to the power sensor. If the problem still exists, try using a different power sensor to determine if the problem is in the power meter or in the power sensor. CAU T ION Electrostatic discharge will render the power sensor inoperative.
1 Introduction Clean the connector face using a cotton swab dipped in isopropyl alcohol. If the swab is too big use a round wooden toothpick wrapped in a lint free cotton cloth dipped in isopropyl alcohol. Refer to Agilent Application Note 326, Principals of Microwave Connector Care (5954- 1566) or Microwave Connector Care (08510- 90064) for proper cleaning methods. Disassembly Procedure Figure 5 CAU T ION Removing Power Sensor Shell Diassemble the power sensor only in a static free workstation.
Introduction 1 Reassembly Procedure 1 Replace the magnetic shields and the plastic shells as shown in Figure 5. Snap the plastic shells together.
1 Introduction Agilent Sales and Service Offices Table 7 Agilent Sales and Service Offices United States: (tel) 800 829 4444 (fax) 800 829 4433 Canada: (tel) 877 894 4414 (fax) 800 746 4866 China: (tel) 800 810 0189 (fax) 800 820 2816 Europe: (tel) 31 20 547 2111 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (080) 769 0800 (fax) (080) 769 0900 Latin America: (tel) (305) 269 7500 Taiwan: (tel) 0800 047 866 (fax) 0800 286 331 Other Asia Pacific Countries: (tel) (65) 6375 8
Agilent E4412A and E4413A Power Sensors Operating and Service Guide A Appendix E4412A and E4413A Specifications (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) 2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) 4 E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) 5 The following specifications apply to E4412A and E4413A sensors (formerly EXCP-E18A and ECP-E26A, respectively) with a seria
A Appendix Table A-1 E4412A and E4413A Specifications (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) Limit Frequency Range E4412A: 10 MHz to 18 GHz E4413A: 50 MHz to 26.5 GHz Power Range –70 dBm to +20 dBm (100 pW to 100 mW) Impedance 50 ohm Connector Type E4412A: Type-N (male) Comments nominal E4413A: 3.5 mm (male) SWR Rho 10 MHz to <30 MHz 1.34 30 MHz to <10 GHz 1.22 0.145 0.100 16.8 20.0 10 GHz to <18 GHz 1.27 0.120 18.4 50 MHz to <2 GHz 1.25 0.110 2 GHz to <18 GHz 1.
Appendix Limit Operating Temperature A Comments 0 to 55°C Net Weight E4412A 0.47 kg (1.04 lb) E4413A 0.45 kg (1.00 lb) Dimensions E4412A Length: 130 mm (5.1 in) E4413A Length: 102 mm (4.0 in) Width: 38 mm (1.5 in) Same for both models Height: 30 mm (1.2 in) Same for both models a. Limits are in percent of power in Watts. See Figure 2 for relative power measurement.
A Appendix Table A-2 E4412A Calibration Factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) A-4 Frequency Uncertainty (%) 10 MHz 2.7 30 MHz 2.7 50 MHz Reference 100 MHz 2.7 1.0 GHz 2.7 2.0 GHz 3.1 4.0 GHz 3.1 6.0 GHz 3.1 8.0 GHz 3.1 10.0 GHz 3.1 11.0 GHz 3.1 12.0 GHz 3.3 14.0 GHz 3.3 16.0 GHz 3.3 18.0 GHz 3.
Appendix A Table A-3 E4413A Calibration factor Uncertainty at 1 mW (0 dBm) (Serial Prefixes Below US3848/MYxxxxxxxx/SGxxxxxxxx) Frequency Uncertainty (%) 50 MHz Reference 100 MHz 2.7 1.0 GHz 2.7 2.0 GHz 3.1 4.0 GHz 3.1 6.0 GHz 3.1 8.0 GHz 3.1 10.0 GHz 3.1 12.0 GHz 3.3 14.0 GHz 3.3 16.0 GHz 3.3 17.0 GHz 3.3 18.0 GHz 3.5 20.0 GHz 3.5 22.0 GHz 3.5 24.0 GHz 3.5 26.0 GHz 3.5 26.5 GHz 3.
A A-6 Appendix Agilent E4412A and E4413A Power Sensors Operating and Service Guide
www.agilent.