Table 3. Corrected System Performance With Type-N Device Connectors, 85092C Electronic
Calibration Module
Applies to the E8801A, E8802A, and E8803A analyzer, 85092C (Type-N, 50Ω) electronic calibration (ECal) module, N6314A test port cable,
and a full 2-port calibration. Also applies to the following conditions:
• IF bandwidth = 10 Hz
• No averaging applied to data
• Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
• Isolation calibration not omitted
Specification (dB) Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 to
9 GHz
a
Directivity 52 54 52 47
Source Match 45 44 41 36
Load Match 47 47 44 39
Reflection Tracking ±0.040 ±0.040 ±0.060 ±0.070
Transmission
Tracking
±0.039 ±0.039 ±0.068 ±0.136
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)