User guide

4
Corrected System Performance with Type-N Connectors
Table 2. Corrected System Performance With Type-N Device Connectors, 85032F Calibration
Kit
Applies to the E8801A, E8802A, and E8803A analyzer, 85032F (Type-N, 50) calibration kit, N6314A test port cable, and a full 2-port
calibration. Also applies to the following conditions:
IF bandwidth = 10 Hz
No averaging applied to data
Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
Isolation calibration not omitted
Specification (dB) Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6GHz to
9 GHz
Directivity 49 46 40 38
Source Match 41 40 36 35
Load Match 49 45 39 37
Reflection Tracking ±0.011 ±0.021 ±0.032 ±0.054
Transmission
Tracking
±0.012 ±0.020 ±0.055 ±0.083
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)