20
Measurement Throughput Summary
Table 21. Typical Cycle Time
a,b
(ms)
Number of Points
101 201 401 1601
Start 1.8 GHz, Stop 2 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
7 10 16 52
2-Port cal 27 36 55 164
Start 300 kHz, Stop 3 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
48 54 64 104
2-Port cal 103 119 145 254
Start 300 kHz, Stop 9 GHz, 35 kHz IF bandwidth
Uncorrected,
1-port cal
51 57 64 103
2-Port cal 112 124 138 220
a Typical performance.
b Includes sweep time, retrace time and band-crossing time. Analyzer display turned off with
DISPLAY:ENABLE OFF. Add 21 ms for display on. Data for one trace (S11) measurement..
Table 22. Cycle Time vs. IF Bandwidth
a
Applies to the Preset condition (201 points, correction off) except for the following changes:
• CF = 1 GHz
• Span = 100 MHz
• Display off (add 21 ms for display on)
IF Bandwidth
(Hz)
Cycle Time (ms)
b
40,000 8
35,000 9
30,000 11
20,000 13
10,000 28
7000 36
5000 48
3000 72
1000 196
300 620
100 1875
30 8062
10 17877