User guide

8
Table 6. Corrected System Performance With 7-16 Device Connector Type, 85038A Calibration
Kit
Applies to the E8801A, E8802A, and E8803A analyzer, 85038A (7-16, 50) calibration kit, N6314A test port cable, and a full 2-port
calibration. Also applies to the following conditions:
IF bandwidth = 10 Hz
No averaging applied to data
Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
Isolation calibration not omitted
Specification (dB) Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 to
6 GHz
6 to
9 GHz
a
Directivity 40 40 36 36
Source Match 37 37 34 34
Load Match 39 39 35 35
Reflection Tracking ±0.089 ±0.089 ±0.115 ±0.115
Transmission
Tracking
±0.024 ±0.033 ±0.082 ±0.103
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)