User guide
7
Table 5. Corrected System Performance With 3.5 mm Device Connector Type, 85093C
Electronic Calibration Module
Applies to the E8801A, E8802A, and E8803A analyzer, 85093C (3.5 mm, 50Ω) electronic calibration (ECal) module, N6314A test port cable,
and a full 2-port calibration. Also applies to the following conditions:
• IF bandwidth = 10 Hz
• No averaging applied to data
• Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
• Isolation calibration not omitted
Specification (dB) Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
9 GHz
a
Directivity 52 52 51 47
Source Match 44 44 39 34
Load Match 47 47 44 40
Reflection Tracking ±0.030 ±0.040 ±0.050 ±0.070
Transmission
Tracking
±0.039 ±0.049 ±0.068 ±0.117
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)