User guide
6
Corrected System Performance with 3.5 mm Connectors
Table 4. Corrected System Performance With 3.5 mm Device Connector Type, 85033E
Calibration Kit
Applies to the E8801A, E8802A, and E8803A analyzer, 85033E (3.5 mm, 50Ω) calibration kit, N6314A test port cable, and a full 2-port
calibration. Also applies to the following conditions:
• IF bandwidth = 10 Hz
• No averaging applied to data
• Environmental temperature 25° ±5 °C, with < 1 °C deviation from calibration temperature
• Isolation calibration not omitted
Specification (dB) Description
300 kHz to
1.3 GHz
1.3 GHz to
3 GHz
3 GHz to
6 GHz
6 GHz to
9 GHz
Directivity 46 44 38 38
Source Match 43 40 37 36
Load Match 46 44 38 38
Reflection Tracking ±0.006 ±0.007 ±0.009 ±0.010
Transmission
Tracking
±0.012 ±0.021 ±0.057 ±0.075
Transmission Uncertainty (Specifications)
Reflection Uncertainty (Specifications)