Agilent PNA Microwave Network Analyzers Data Sheet This document describes the performance and features of the Agilent Technologies PNA microwave network analyzers: E8362C 10 MHz to 20 GHz E8363C 10 MHz to 40 GHz E8364C 10 MHz to 50 GHz E8361C 10 MHz to 67 GHz Note: For the complete and most current instrument, calibration kit and connector specifications, refer to the online Help file in the “manuals” library on our web site: http://na.tm.agilent.
Some Definitions All specifications and characteristics apply over a 25 °C ±5 °C range (unless otherwise stated) and 90 minutes after the instrument has been turned on. Nominal (nom.): A general, descriptive term that does not imply a level of performance. It is not covered by the product warranty. Calibration: The process of measuring known standards to characterize a network analyzer’s systematic (repeatable) errors. Specification (spec.): Warranted performance.
Table of Contents E8362/3/4C Corrected system performance . . . . . . . . . . . . . . . . 4 System dynamic range . . . . . . . . . . . . . . . . . . . . 4 Receiver dynamic range . . . . . . . . . . . . . . . . . . 6 Corrected system performance with 2.4 mm connectors. . . . . . . . . . . . . . . . . 7 Corrected system performance with 3.5 mm connectors. . . . . . . . . . . . . . . . . 9 Uncorrected system performance . . . . . . . . . . . . . 11 Test port output . . . . . . . . . . . . . . . . . . . . . . .
E8362/3/4C Corrected system performance The specifications in this section apply for measurements made with the Agilent E8362/3/4C PNA Series microwave network analyzer with the following conditions: • 10 Hz IF bandwidth • no averaging applied to data • isolation calibration with an averaging factor of 8 Note: Samples of uncertainty curves are included in this Data Sheet. Please download our free uncertainty calculator (www.agilent .com/find/na_calculator) to generate the curves for your setup.
E8362/3/4C Corrected system performance continued System dynamic range 1 Description Specification (dB) at test port 2 Typical (dB) at direct receiver access input 3 Supplemental information Dynamic range Standard configuration and extended power range and bias-tees (E8362/3/4C-Option UNL) 10 to 45 MHz 4 79 N/A 45 to 500 MHz 5 92 N/A 500 MHz to 2 GHz 117 N/A 2 to 10 GHz 120 N/A 10 to 20 GHz 121 N/A 20 to 30 GHz 112 N/A 30 to 40 GHz 108 N/A Option 016 degrades 40 to 45 GHz 105 N/A
E8362/3/4C Receiver dynamic range1 Description Specification (dB) at test port2 Typical (dB) at direct receiver access input3 Supplemental information Dynamic range Standard configuration and standard power range (E8362/3/4C) or standard configuration and extended power range and bias-tees (E8362/3/4C-Option UNL) 10 to 45 MHz4 82 N/A 45 to 500 MHz5 94 N/A 500 MHz to 2 GHz 119 N/A 2 to 10 GHz 122 N/A 10 to 20 GHz 125 N/A 20 to 30 GHz 114 N/A Option 016 degrades performance by 2 dB 30 t
E8362/3/4C Corrected system performance with 2.4 mm connectors Standard configuration and standard power range (E8363/4C) Applies to E8363/4C PNA Series analyzer, 85056A (2.4 mm) calibration kit, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 45 MHz to 2 GHz 42 41 42 ±0.001 (+0.02/°C) ±0.010 (+0.
E8362/3/4C Corrected system performance with 2.4 mm connectors continued Fully Optioned (E8363/4C-Option 014/UNL/080/081/016) Applies to E8363/4C PNA Series analyzer, 85056A (2.4 mm) calibration kit, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.
E8362/3/4C Corrected system performance with 3.5 mm connectors Standard configuration and standard power range (E836xC) Applies to E836xC PNA Series analyzer, 85052B (3.5 mm) calibration kit, 85131F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Data and traces above 20 GHz applicable to E8363C and E8364C network analyzers, when used with 85052B 26.
E8362/3/4C Corrected system performance with 3.5 mm connectors continued Fully Optioned (E836xC-Option 014/UNL/080/081/016) Applies to E836xC PNA Series analyzer, 85052B (3.5 mm) calibration kit, 85131F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Data and traces above 20 GHz applicable to E8363C and E8364C network analyzers, when used with 85052B 26.
E8362/3/4C Uncorrected system performance 1 Description Specification Directivity 10 to 45 MHz2 23 dB 45 MHz to 2 GHz 24 dB 2 to 10 GHz 22 dB 10 to 20 GHz 16 dB 20 to 40 GHz 16 dB 40 to 45 GHz 15 dB 45 to 50 GHz 13 dB Source match - standard 10 to 45 MHz2 11 dB 45 MHz to 2 GHz 23 dB 2 to 10 GHz 16 dB 10 to 20 GHz 14 dB 20 to 40 GHz 10 dB 40 to 45 GHz 9 dB 45 to 50 GHz 7.
E8362/3/4C Uncorrected system performance 1 continued Description Crosstalk1 - standard 10 to 45 MHz2 45 MHz to 1 GHz 1 to 2 GHz 2 to 20 GHz 20 to 40 GHz 40 to 45 GHz 45 to 50 GHz Crosstalk1 - Option UNL or 014 10 to 45 MHz2 45 MHz to 1 GHz 1 to 2 GHz 2 to 20 GHz 20 to 40 GHz 40 to 45 GHz 45 to 50 GHz Crosstalk1 - Option UNL and 014 10 to 45 MHz2 45 MHz to 1 GHz 1 to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 40 GHz 40 to 45 GHz 45 to 50 GHz Crosstalk - Option 080 enabled 3 10 to 45 MHz 45 MHz to 1 GHz 1 to 2 GH
E8362/3/4C Test port output Description Standard Frequency range E8362C E8363C E8364C Nominal power 2 E8362C 0 dBm E8363/4C -12 dBm Frequency resolution 1 Hz CW accuracy ± 1ppm Frequency stability Power level accuracy 1 10 to 45 MHz 3 ±2.0 dB 45 MHz to 10 GHz ±1.5 dB 10 to 20 GHz ±2.0 dB 20 to 40 GHz ±3.0 dB 40 to 45 GHz ±3.0 dB 45 to 50 GHz ±3.0 dB Power level linearity 6 10 to 45 MHz 3 ±1.0 dB 4 45 MHz to 20 GHz ±1.0 dB 4 20 to 40 GHz ±1.0 dB 4 40 to 50 GHz ±1.
E8362/3/4C Test port output continued Description Specification Supplemental information Phase noise (1 kHz offset from center frequency, nominal power at test port) 10 MHz to 10 GHz -60 dBc typical 10 to 20 GHz -55 dBc typical 20 to 50 GHz -50 dBc typical Phase noise (1 kHz offset from center frequency, nominal power at test port) – Option 080 enabled 10 MHz to 10 GHz -60 dBc typical 10 to 20 GHz -60 dBc typical 20 to 50 GHz -50 dBc typical Phase noise (10 kHz offset from center frequency, nominal power
E8362/3/4C Test port input Description Standard 014 Specification UNL Test port noise floor 1 10 Hz IF bandwidth 10 to 45 MHz 2 < -77 dBm < -77 dBm 45 to 500 MHz 3 < -89 dBm < -89 dBm 500 MHz to 2 GHz < -114 dBm < -114 dBm 2 to 10 GHz < -117 dBm < -117 dBm 10 to 20 GHz < -120 dBm < -119 dBm 20 to 40 GHz < -114 dBm < -113 dBm 40 to 50 GHz < -114 dBm < -112 dBm 1 kHz IF bandwidth 10 to 45 MHz 2 < -57 dBm < -57 dBm 45 to 500 MHz 3 < -69 dBm < -69 dBm 500 MHz to 2 GHz < -94 dBm < -94 dBm 2 to 10 GHz < -97 dB
E8362/3/4C Test port input continued Description Specification Standard, 014, UNL UNL and 014 Direct receiver access input noise floor1,2 - Option 080 enabled 4 10 Hz IF bandwidth 10 to 45 MHz < -127 dBm < -127 dBm 45 to 500 MHz3 < -126 dBm < -126 dBm 500 MHz to 2 GHz < -132 dBm < -132 dBm 2 to 10 GHz < -131 dBm < -131 dBm 10 to 20 GHz < -133 dBm < -133 dBm 20 to 40 GHz < -124 dBm < -124 dBm 40 to 50 GHz < -122 dBm < -122 dBm 1 kHz IF bandwidth 10 to 45 MHz < -107 dBm < -107 dBm 45 to 500 MHz3 < -106 dBm <
E8362/3/4C Test port input continued Description Standard Specification 014 Trace noise magnitude 10 to 45 MHz1 < 0.050 dB rms 45 to 500 MHz2 < 0.010 dB rms 500 MHz to 20 GHz < 0.006 dB rms 20 to 40 GHz < 0.006 dB rms 40 to 50 GHz < 0.006 dB rms Trace noise magnitude – Option 080 enabled 1, 4 10 to 45 MHz1 < 0.060 dB rms 45 to 500 MHz2 < 0.010 dB rms 500 MHz to 20 GHz < 0.006 dB rms 20 to 40 GHz < 0.007 dB rms 40 to 50 GHz < 0.008 dB rms Trace noise phase 10 to 45 MHz1 < 0.350° rms 45 to 500 MHz2 < 0.
E8362/3/4C Test port input continued Group delay 1 Description Aperture (selectable) Maximum aperture Range Maximum delay Specification Supplemental information (typical) (frequency span)/(number of points – 1) 20% of frequency span 0.5 x (1/minimum aperture) Limited to measuring no more than 180° of phase change within the minimum aperture. The following graph shows characteristic group delay accuracy with type-N full 2-port calibration and a 10 Hz IF bandwidth.
E8362/3/4C Test port input continued Dynamic accuracy (specifications) 1 Applies to input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. Also applies to the following conditions: • IF bandwidth = 10 Hz Magnitude Phase E836xC E836xC 100 -10 dBm (10 MHz-20 GHz) -20 dBm (10 MHz-20 GHz) -30 dBm (10 MHz-20 GHz) -40 dBm (10 MHz-20 GHz) 1 Accuracy (degrees) Accuracy (dB) 10 0.
E8361C Corrected system performance The specifications in this section apply for measurements made with the Agilent E8361C PNA Series microwave network analyzer with the following conditions: • 10 Hz IF bandwidth • no averaging applied to data System dynamic range 1 Description Specification (dB) Typical (dB) at direct receiver access input 3 at test port 2 Supplemental information Dynamic range Standard configuration (E8361C) 61 N/A 10 to 45 MHz 4 87 N/A 45 to 500 MHz 5 500 to 750 MHz 112 N/A 750 MHz t
E8361C Corrected system performance with 1.85 mm connectors Standard configuration and standard power range Applies to E8361C PNA Series analyzer, N4694A (1.85 mm) ECal electronic calibration module, N4697E/F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.
E8361C Corrected system performance with 1.85 mm connectors continued Standard configuration and standard power range (E8361C) Applies to E8361C PNA Series analyzer, N4694A (1.85 mm) ECal electronic calibration module, N4697E/F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.
E8361C Corrected system performance with 1.85 mm connectors continued Fully optioned (E8361C with options 014/UNL/080/081/016) Applies to E8361C PNA Series analyzer, N4694A (1.85 mm) ECal electronic calibration module, N4697F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 10 to 200 MHz1 33 25 25 ±0.050 (+0.02/°C) ±0.146 (+0.
E8361C Corrected system performance with 1.85 mm connectors continued Fully optioned (E8361C with options 014/UNL/080/081/016) Applies to E8361C PNA Series analyzer, N4694A (1.85 mm) ECal electronic calibration module, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 30 to 40 GHz 44 34 33 ±0.060 (+0.02/°C) ±0.082 (+0.
E8361C Corrected system performance with 1.85 mm connectors continued Standard configuration and standard power range (E8361C) Applies to E8361C PNA Series analyzer, 85058B (1.85 mm) calibration kit, N4697F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 10 to 45 MHz1 35 34 34 ±0.019 (+0.02/°C) ±0.164 (+0.
E8361C Corrected system performance with 1.85 mm connectors continued Standard configuration and standard power range (E8361C) Applies to E8361C PNA Series analyzer, 85058B (1.85 mm) calibration kit, N4697F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 20 to 35 GHz 37 41 36 ±0.033 (+0.02/°C) ±0.097 (+0.
E8361C Corrected system performance with 1.85 mm connectors continued Fully optioned (E8361C with options 014/ UNL/080/081/016) Applies to E8361C PNA Series analyzer, 85058B (1.85 mm) calibration kit, N4697F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 10 to 45 MHz1 35 34 34 ±0.019 (+0.02/°C) ±0.177 (+0.
E8361C Corrected system performance with 1.85 mm connectors continued Fully optioned (E8361C with options 014/ UNL/080/081/016) Applies to E8361C PNA Series analyzer, 85058B (1.85 mm) calibration kit, N4697F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 20 to 35 GHz 37 41 36 ±0.033 (+0.02/°C) ±0.084 (+0.
E8361C Corrected system performance with 2.4 mm connectors Standard configuration and standard power range (E8361C) Applies to E8361C PNA Series analyzer, N4693A (2.4 mm) ECal electronic calibration module, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.) Description Specification (dB) 10 to 200 MHz1 32 25 24 ±0.050 (+0.02/°C) ±0.100 (+0.
E8361C Corrected system performance with 2.4 mm connectors continued Fully optioned (E8361C with options 014/UNL/080/081/016) Applies to E8361C PNA Series analyzer, N4693A (2.4 mm) ECal electronic calibration module, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.
E8361C Corrected system performance with 2.4 mm connectors continued Standard configuration and standard power range (E8361C) Applies to E8361C PNA Series analyzer, 85056A (2.4 mm) calibration kit, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.
E8361C Corrected system performance with 2.4 mm connectors continued Fully optioned (E8361C with options 014/ UNL/080/081/016) Applies to E8361C PNA Series analyzer, 85056A (2.4 mm) calibration kit, 85133F flexible test port cable set, and a full two-port calibration. (Specifications apply over environmental temperature of 23 °C ±3 °C, with less than 1 °C deviation from calibration temperature.
E8361C Uncorrected system performance 1 Description Directivity 10 to 45 MHz 2 45 MHz to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 30 GHz 30 to 50 GHz 50 to 60 GHz 60 to 67 GHz 67 to 70 GHz2 Source match - standard 10 to 45 MHz 2 45 MHz to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to 67 GHz 67 to 70 GHz2 Source match - Option 014 10 to 45 MHz 2 45 MHz to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to
E8361C Uncorrected system performance 1 continued Description Load match - Option 014 10 to 45 MHz 2 45 MHz to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to 67 GHZ 67 to 70 GHz2 Load match - Option 014 and UNL 10 to 45 MHz 2 45 MHz to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to 67 GHZ 67 to 70 GHz2 Specification Typical 5.5 dB 8.5 dB 8 dB 8 dB 7.5 dB 7 dB 7.5 dB 7.5 dB 6 dB 6 dB 5 dB 5.
E8361C Uncorrected system performance 1 continued Description Reflection tracking 10 to 45 MHz 45 MHz to 20 GHz 20 to 40 GHz 40 to 50 GHz 50 to 67 GHz 67 to 70 GHz Transmission tracking 3 10 to 45 MHz 45 MHz to 20 GHz 20 to 40 GHz 40 to 50 GHz 50 to 67 GHz 67 to 70 GHz Crosstalk 4 - standard 10 to 45 MHz 2 45 to 500 MHz 500 MHz to 2 GHz 2 to 10 GHz 10 to 24 GHz 24 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to 67 GHz 67 to 70 GHz 2 Crosstalk 4 - Option 014 10 to 45 MHz 2 45 to 500 MHz
E8361C Test port output Description Frequency range E8361C Nominal power1 Frequency resolution CW accuracy Frequency stability Power level accuracy2 10 to 45 MHz 3 45 MHz to 10 GHz 10 to 20 GHz 20 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to 67 GHz 67 to 70 GHz 3 Power level linearity 4 10 to 45 MHz 3 45 MHz to 67 GHz 67 to 70 GHz 3 Specification Standard -15 dBm 1 Hz ± 1 ppm Option 014 Option UNL and Option 014 Supplemental information 10 MHz to 67 GHz (Operation up to 70 GHz)
E8361C Test port output continued Description Specification Supplemental information Phase noise (10 kHz offset from center frequency, nominal power at test port) 10 to 45 MHz 80 dBc typical 45 MHz to 10 GHz 70 dBc typical 10 to 24 GHz 60 dBc typical 24 to 70 GHz 55 dBc typical Phase noise (10 kHz from center frequency, nominal power at test port) – Option 080 enabled 10 to 45 MHz 80 dBc, typical 45 MHz to 10 GHz 70 dBc, typical 10 to 24 GHz 60 dBc, typical 24 to 70 GHz 55 dBc, typical Phase noise (100 kH
E8361C Test port input Description Specification Standard Option 014 or Option 016 Option UNL and 014 Test port noise floor 1 10 Hz IF bandwidth 10 to 45 MHz 3 45 to 500 MHz 2, 4 500 MHz to 2 GHz 2 to 10 GHz 10 to 24 GHz 24 to 30 GHz 30 to 40 GHz 40 to 45 GHz 45 to 50 GHz 50 to 60 GHz 60 to 67 GHz 67 to 70 GHz 3 Supplemental information 080 enabled,5 typical < -70 dBm < -90 dBm < -112 dBm < -112 dBm < -116 dBm < -105 dBm < -105 dBm < -103 dBm < -101 dBm < -100 dBm < -99 dBm < -99 dBm < -70 dBm < -90 d
E8361C Test port input Description continued Specification Standard Option 014 or Option 014 and UNL (typ.) Supplemental information < -106 dBm < -105 dBm Online Help also includes the 500 MHz to 2 GHz < -125.
E8361C Test port input continued Description Specification Standard or Option 014 Third Order Intercept1 – Tone spacing from 100 kHz to 5 MHz Option UNL and Option 014 Typical: +30 dBm +24 dBm +23 dBm +24 dBm +26 dBm 10 to 500 MHz 500 MHz to 24 GHz 24 to 40 GHz 40 to 50 GHz 50 to 67 GHz Third Order Intercept1 – Tone spacing from 5 to 20 MHz Typical: Not applicable +20 dBm +20 dBm +22 dBm +24 dBm 10 to 500 MHz 500 MHz to 24 GHz 24 to 40 GHz 40 to 50 GHz 50 to 67 GHz Third Order Intercept1 – Tone spac
E8361C Test port input continued Description Reference level magnitude Range Resolution Reference level phase Range Resolution Stability magnitude 1 Specification Standard Option 014 ±500 dB 0.001 dB ±500 dB 0.001 dB ±500° 0.01° ±500° 0.
E8361C Test port input continued Group delay 1 Description Aperture (selectable) Maximum aperture Range Maximum delay Supplemental information (typical) (frequency span)/(number of points – 1) 20% of frequency span 0.5 x (1/minimum aperture) Limited to measuring no more than 180° of phase change within the minimum aperture. Specification The following graph shows characteristic group delay accuracy with type-N full 2-port calibration and a 10 Hz IF bandwidth.
E8361C Test port input continued Dynamic accuracy (specification)1 Applies to input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. Also applies to the following conditions: • IF bandwidth = 10 Hz Note: If the power is set above maximum specified leveled power, the test port output signal may show non-linear effects that are dependent on the DUT. Magnitude 1 E8361C typical 100 -10 dBm at 0.045 GHz -20 dBm at 0.045 GHz -30 dBm at 0.
E8361C Test port input continued Dynamic accuracy (specification)1 Applies to input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. Also applies to the following conditions: • IF bandwidth = 10 Hz Note: If the power is set above maximum specified leveled power, the test port output signal may show non-linear effects that are dependent on the DUT. Magnitude 1 0.1 0.
E8361C Test port input continued Dynamic accuracy (specification)1 Applies to input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. Also applies to the following conditions: • IF bandwidth = 10 Hz Note: If the power is set above maximum specified leveled power, the test port output signal may show non-linear effects that are dependent on the DUT. Magnitude E8361C 014 100 Accuracy (degrees) -10 dBm (0.500-5 GHz) -20 dBm (0.
E8361C Test port input continued Dynamic accuracy (specification)1 Applies to input ports 1 and 2, accuracy of the test port input power reading relative to the reference input power level. Also applies to the following conditions: • IF bandwidth = 10 Hz Note: If the power is set above maximum specified leveled power, the test port output signal may show non-linear effects that are dependent on the DUT.
Microwave PNA Series General information Description System IF bandwidth range RF connectors E8362C E8363/4C E8361C Display Display range Magnitude Phase Polar Display resolution Magnitude Phase Marker resolution Magnitude Phase Polar CPU Line power (single phase) Frequency Max General environmental EMC Safety Operating environment Temperature Error-corrected temperature range Relative humidity Altitude Supplemental information 1 Hz to 40 kHz, nominal 3.
Microwave PNA Series General information continued Description Non-operating storage environment Temperature Cabinet dimensions Supplemental information -40 to +70 °C Excluding front and rear panel hardware and feet As shipped - includes front panel connectors, rear panel bumpers, and feet. As shipped plus handles As shipped plus rack mount flanges As shipped plus handles and rack mount flanges Weight Net Shipping 48 29 kg (64 lb), nom. 36 kg (80 lb), nom. Height 267 mm 10.50 in 280 mm 11.
Microwave PNA Series Rear panel Description Supplemental information External trigger rear panel I/O (typical) Trigger inputs/outputs BNC(f), TTL/CMOS compatible Trigger input Function Source Signal levels Input impedance Minimum trigger width Trigger modes Trigger delay range Trigger delay resolution Trigger output Function Source Signal levels Trigger polarity Pulse width Measurement of next point, next channel, or next group of channels Aux I/O (pin 19) or I/O 1 (BNC (f) connector) TTL-compatible 5
Microwave PNA Series Rear panel continued Description 10 MHz reference in Input frequency Input power Input impedance 10 MHz reference out Output frequency Signal type Output power Output impedance Harmonics Test set I/O Handler I/O Auxiliary I/O Bias tee inputs Connectors Maximum voltage Maximum current Fuse Supplemental information 10 MHz ±10 ppm, typ. –15 to +20 dBm, typ. 200 , nom. 10 MHz ±10 ppm, typ. Sine wave, typ. 10 dB ± 4 dB into 50 , typ. 50 , nom. < -40 dBc, typ.
Microwave PNA Series Measurement throughput summary Cycle time vs. IF bandwidth 1 Cycle time vs. number of points 1 Instrument state: preset condition, 201 points, CF = 28 GHz, Span = 100 MHz, correction off. Add 21 ms for display on. Cycle time includes sweep and re-trace time. Instrument state: preset condition, 35 kHz IF bandwidth, CF = 28 GHz, Span = 100 MHz, correction off. Add 21 ms for display on. Cycle time includes sweep and re-trace time.
Frequency Converter Application (Option 083) cycle time for fixed-IF measurements (s)1 101 Number of points 201 401 Hardware trigger 8.
Microwave PNA Series Measurement capabilities Source control Number of measurement channels Measured number of points per sweep Thirty-two independent measurement channels. A measurement channel is coupled to stimulus settings including frequency, IF bandwidth, power level, and number of points. User definable from 2 to 20,001. Number of display windows Segment sweep Sweep type Linear, CW (single frequency), power or segment sweep. Up to 24 active traces and 24 memory traces per window.
Microwave PNA Series Automation SCPI COM/DCOM GPIB LAN Internal X X X X X Methods Controlling via internal analyzer execution Write applications that can be executed from within the analyzer via COM (component object model) or SCPI standard-interface commands. These applications can be developed in a variety of languages, including Visual Basic, Visual C++, Agilent VEE, or LabView™ programming languages. Controlling via GPIB The GPIB interface operates to IEEE 488.
amount in degrees. calibration mixer is part of the calibration process. Microwave PNA Series Data accuracy enhancement Measurement calibration Measurement calibration significantly reduces measurement uncertainty due to errors caused by system directivity, source and load match, tracking and crosstalk. Full two-port calibration removes all the systematic errors to obtain the most accurate measurements.
Microwave PNA Series System capabilities Improved graphical user interface All PNA Series instruments (PNA-X, PNA-L, PNA) now employ the same graphical user interface and front panel layout. In addition, the new touch screen and softkeys make operating the PNA easier than ever. Built-in information system Embedded documentation provides measurement assistance in five different languages (English, Chinese, French, German, Japanese, and Spanish).
Frequency-offset (Option 080) Embedded LO measurements (Option 084) This option enables the PNA Series microwave network analyzers to set the source frequency independently from where the receivers are tuned. This ability is important for two general classes of devices: mixers (and converters) and amplifiers.
IF Access (Option H11) Commercial calibration certificate with test data (Option UK6) Provides hardware to enable antenna, point-in-pulse, and pulse-profile measurements, as well as broadband millimeter-wave measurements to 110 GHz, and banded millimeter-wave measurements to 325 GHz. For each of the microwave PNA’s measurement receivers, IF gates (enabled with pulsed-RF measurement capability Option H08) and external IF inputs are added.
Microwave PNA Series Simplified test set block diagram Standard power range R1 Extended power range and bias-tees (Option UNL) Source Source Switch Switch Reference receiver Reference receiver R1 R2 Source A attenuator1 R2 Source attenuator1 Bias-tee B Bias-tee A Measurement receivers Port 1 Reference receiver Reference receiver Port 1 Port 2 Measurement receivers B Port 2 Configuration test set (Option 014) Source Switch R1 Reference receiver Reference receiver A R2 B Mea
Fully optioned (Options 014, UNL, 016, 080, 081) Source Switch/splitter/leveler Source Reference receiver Reference receiver R1 attenuator1 R2 Source attenuator1 Option 081 Measurement receivers 10 dB steps Bias-tee 10 dB steps Bias-tee B A Receiver attenuator2 Source out SRC out Rcvr R1 in Cplr thru Cplr arm Rcvr B in Rcvr A in Cplr arm SRC Rcvr R2 Source out in out Cplr thru Fully optioned pulse, antenna, or mm-wave configuration (Options 014, UNL, 016, 080, 081, H11) Aux RF out (
Ordering guide for PNA series Network analyzers This guide is intended to assist you in the ordering process. For detailed ordering information, refer to the PNA Series Microwave Network Analyzer Configuration Guide (literature number 5989-7606EN). PNA Series microwave network analyzers1 E8362C E8363C E8364C E8361C 10 MHz to 20 GHz 10 MHz to 40 GHz 10 MHz to 50 GHz 10 MHz to 67 GHz Options To add options to a product, order the corresponding item number.
Test port cable specifications Connector Type Frequency (Test port to device) (GHz) Length2 cm (inch) Return loss Insertion loss (dB) (f in GHz) Stabillty1,2 ±magnitude ±Phase (degrees) DC to 26.5 81 (32) 17 dB 0.16 (f) +0.5 DC to 26.5 96.5 (38) 16 dB <0.22 dB 0.16 (f) +0.8 DC to 18 81 (32) 17 dB <0.06 dB 0.16 (f) +0.5 85132E flexible cable 3.5 mm3 to 7 mm DC to 18 97.2 (38.25) 17 dB 0.43 f +0.3 (2.5 dB at fmax) 0.35 f +0.3 (2.1 dB at fmax) 0.35 f +0.3 (1.8 dB at fmax) 0 35 f +0.
Test port cable specifications continued Connector Type Frequency (Test port to device) (GHz) Length2 cm (inch) Return loss Insertion loss (dB) (f in GHz) Stabillty1,2 ±magnitude ±Phase (degrees) DC to 50 53 (21 ) 15 dB 0.55 f +0.2 (3.7 dB at fmax) <0.06 dB 0.16 (f) DC to 50 72 (28) 12.5 dB 0.48 f +0.25 (3.64 dB at fmax) <0.17 dB 0.8 + 0.16 (f) DC to 26.5 53 (21) 16 dB 0.31 f +0.2 (1.8 dB at fmax) <0.06 dB 0.18 (f) DC to 26.5 53 (21) 16 dB 0.31 f +0.2 (1.8B dB at fmax) <0.
Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/find/agilentdirect Quickly choose and use your test equipment solutions with confidence. Agilent Open www.agilent.com/find/open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products.