Agilent E5071C ENA Network Analyzer • 9 kHz to 4.5/6.5/8.5 GHz • 100 kHz to 4.5/6.5/8.
Table of Contents Options ........................................................................................................................ 3 Definitions ................................................................................................................. 4 Boundary Conditions ............................................................................................. 4 Corrected System Performance........................................................................
Options This document provides technical specifications for the E5071C ENA network analyzer and the E5091A multiport test set. E5071C-230 E5071C-235 E5071C-240 E5071C-245 E5071C-260 E5071C-265 E5071C-280 E5071C-285 E5071C-430 E5071C-435 E5071C-440 E5071C-445 E5071C-460 E5071C-465 E5071C-480 E5071C-485 E5071C-2D5 E5071C-4D5 E5071C-2K5 E5071C-4K5 E5092A 2-port test set, 9 kHz to 3 GHz without bias tees (Discontinued) 2-port test set, 100 kHz to 3 GHz with bias tees (Discontinued) 2-port test set, 9 kHz to 4.
Definitions Specification (spec.): Warranted performance. All specifications apply at 23 ºC (±5 ºC), unless otherwise stated, and 90 minutes after the instrument has been turned on. Specifications include guard bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions.
Corrected System Performance The specifications in this section apply to measurements made with the Agilent E5071C network analyzer under the following conditions: • No averaging applied to data • Environmental temperature of 23 °C (±5 °C) with less than 1 °C deviation from the calibration temperature • Response and isolation calibration performed System dynamic range Table 1.
System dynamic range (continued) Table 2. Option 2D5/2K5/4D5/4K5 Description Specification SPD System dynamic range1, 2 300 kHz to 1 MHz 70 dB 1 MHz to 10 MHz 82 dB 10 MHz to 100 MHz 95 dB 100 MHz to 6 GHz IF bandwidth = 3 kHz 98 dB 6 GHz to 8.5 GHz 92 dB 8.5 GHz to 10.5 GHz 80 dB 10.5 GHz to 15 GHz 75 dB 15 GHz to 20 GHz 71 dB 300 kHz to 1 MHz 95 dB 105 dB 1 MHz to 10 MHz 107 dB 115 dB 10 MHz to 100 MHz 120 dB 129 dB 123 dB 130 dB 6 GHz to 8 GHz 117 dB 129 dB 8 GHz to 8.
Corrected system performance with calibration kit Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485 Table 3.
Table 4.
Table 5. Corrected system performance with 3.5 mm device connector type, 85033E calibration kit Network analyzer: Calibration kit: Calibration: E5071C 85033E (3.5 mm, 50 Ω) full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (±5 °C ) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) 9 kHz to 300 kHz Description 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.
Table 6. Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration (ECal) module Network analyzer: Calibration module: E5071C 85093C (3.
Option 2D5/2K5/4D5/4K5 Table 7. Corrected system performance with 3.5 mm device connectors, 85052D calibration kit Network analyzer: Calibration kit: Calibration: E5071C 85052D (3.
Table 8. Corrected system performance with 3.5 mm device connectors, N4691B electronic calibration (ECal) module Network analyzer: Calibration module: E5071C N4691B (3.
Uncorrected System Performance Table 9. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 User correction: OFF, system correction: ON Specification (dB) Description 9 kHz to 300 kHz 300 kHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.5 GHz Directivity 20 dB 25 dB 20 dB 15 dB Source match 20 dB 25 dB 20 dB 15 dB Load match 12 dB 17 dB 12 dB 10 dB Transmission tracking ±1.5 dB ±1.0 dB ±1.0 dB ±1.0 dB Reflection tracking ±1.5 dB ±1.0 dB ±1.0 dB ±1.0 dB Table 10.
Test Port Output (Source) Test port output frequency Table 11. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485/2D5/2K5/4D5/4K5 Description Specification Frequency range Option 230/430 Option 240/440 Option 260/460 Option 280/480 Option 235/435 Option 245/445 Option 265/465 Option 285/485 Option 2D5/4D5 Option 2K5/4K5 9 kHz to 3 GHz 9 kHz to 4.5 GHz 9 kHz to 6.5 GHz 9 kHz to 8.5 GHz 100 kHz to 3 GHz 100 kHz to 4.5 GHz 100 kHz to 6.5 GHz 100 kHz to 8.
Test port output power1 Table 12. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Nominal power (preset power) 0 dBm Level accuracy2, 6 (stepped sweep mode) ±0.650 dB (at 0 dBm, 50 MHz absolute) ±1.0 dB Level accuracy2 (swept sweep mode) Level linearity3, 5, 6 (stepped sweep mode) 9 kHz to 5 GHz 5 GHz to 6 GHz 6 GHz to 7 GHz 7 GHz to 8.5 GHz Typical ±2.5 dB ±0.75 dB (–20 to 10 dBm) ±0.75 dB (–20 to 9 dBm ) ±0.75 dB (–20 to 8 dBm ) ±0.
Test port output power7 (continued) Table 13. Option 2D5/2K5/4D5/4K5 Description Specification Typical Nominal power (preset power) –5 dBm Level accuracy6 (stepped sweep mode)1 300 kHz to 1 MHz 1 MHz to 5 MHz 5 MHz to 8.5 GHz 8.5 GHz to 20 GHz ±0.650 dB (at –5 dBm, 50 MHz absolute) +2.0 dB, –6.0 dB ±2.0 dB ±1.0 dB ±2.5 dB Level accuracy (swept sweep mode)2 300 kHz to 1 GHz 1 GHz to 8.5 GHz 8.
Test port output power6 (continued) Table 13. Option 2D5/2K5/4D5/4K5 Description Specification SPD power3, 4 Max leveled 300 kHz to 1 MHz 1 MHz to 10 GHz 10 GHz to 13 GHz 13 GHz to 15 GHz 15 GHz to 18 GHz 18 GHz to 20 GHz 9 dBm 10 dBm 9 dBm 7 dBm 5 dBm 4 dBm Test port output signal purity Table 14. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Typical Harmonics (2nd or 3rd) 9 kHz to 2 GHz 2 GHz to 8.
Test Port Input Test port input levels Table 16. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Typical Maximum test port input level 9 kHz to 8.5 GHz +10 dBm Damage level 9 kHz to 8.5 GHz +26 dBm ±35 VDC Crosstalk1 9 kHz to 300 kHz 300 kHz to 10 MHz 10 MHz to 3 GHz 3 GHz to 6 GHz 6 GHz to 8.
Test port input levels (continued) Table 17. Option 2D5/2K5/4D5/4K5 Description Specification Typical Maximum test port input level 300 kHz to 20 GHz +10 dBm Damage level 300 kHz to 20 GHz +26 dBm or ±35 VDC Crosstalk1 300 kHz to 1 MHz 1 MHz to 5 MHz 5 MHz to 10 MHz 10 MHz to 45 MHz 45 MHz to 4 GHz 4 GHz to 6 GHz 6 GHz to 8.5 GHz 8.
Test port input (compression level) Table 18. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification SPD Compression level Magnitude 9 kHz to 5 GHz 5 GHz to 8.5 GHz (maximum test port input level = +10 dBm) 0.08 dB 0.1 dB Phase 9 kHz to 3 GHz 3 GHz to 5 GHz 5 GHz to 8.5 GHz (maximum test port input level = +10 dBm) 0.3 deg 0.6 deg 1.0 deg Table 19.
Test port input (trace noise) Table 20. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification SPD Trace noise magnitude1 9 kHz to 30 kHz (IFBW = 3 kHz) 0.004 dB rms 0.001 dB rms 30 kHz to 100 kHz (IFBW = 3 kHz) 0.003 dB rms 0.001 dB rms 100 kHz to 10 MHz (IFBW = 3 kHz) 0.003 dB rms 0.0005 dB rms 10 MHz to 4.38 GHz (IFBW = 70 kHz) 0.004 dB rms 0.001 dB rms 4.38 GHz to 5 GHz (IFBW = 70 kHz) 0.006 dB rms 0.
Test port input (trace noise) (continued) Table 21. Option 2D5/2K5/4D5/4K5 Description Specification SPD 300 kHz to 1 MHz (IFBW = 3 kHz) 0.006 dB rms 0.0009 dB rms 1 MHz to 10 MHz (IFBW = 3 kHz) 0.003 dB rms 0.0005 dB rms 10 MHz to 4.38 GHz (IFBW = 70 kHz) 0.004 dB rms 0.0010 dB rms 4.38 GHz to 8.5 GHz (IFBW = 70 kHz) 0.006 dB rms 0.0012 dB rms 8.5 GHz to 13.137 GHz (IFBW = 70 kHz) 0.009 dB rms 0.0024 dB rms 13.137 GHz to 17 GHz (IFBW = 70 kHz) 0.013 dB rms 0.
Test port input (stability)1 Table 22. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Typical Stability magnitude 9 kHz to 3 GHz ±0.005 dB/°C 3 GHz to 6 GHz ±0.01 dB/°C 6 GHz to 8.5 GHz ±0.04 dB/°C Stability phase 9 kHz to 3 GHz ±0.1 deg/°C 3 GHz to 6 GHz ±0.2 deg/°C 6 GHz to 8.5 GHz ±0.8 deg/°C Table 23. Option 2D5/2K5/4D5/4K5 Description Specification Typical Stability magnitude 300 kHz to 3 GHz ±0.005 dB/°C 3 GHz to 6 GHz ±0.
Test port input (dynamic accuracy)1 Table 24. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Accuracy of the test port input power reading is relative to –10 dBm reference input power level. Description Specification Typical Dynamic accuracy magnitude 10 dBm ±0.207 dB –30 dBm ±0.045 dB –100 dBm ±2.00 dB –110 dBm ±3.0 dB Dynamic accuracy phase 10 dBm ±5.03 deg –30 dBm ±0.30 deg –100 dBm ±15.
Test port input (dynamic accuracy)1 (continued) Table 25. Option 2D5/2K5/4D5/4K5 Accuracy of the test port input power reading is relative to –10 dBm reference input power level. Description Specification Typical Dynamic accuracy magnitude 10 dBm ±2.49 dB –30 dBm ±0.046 dB –100 dBm ±2.00 dB –110 dBm ±3.0 dB Dynamic accuracy phase 10 dBm ±20.6 deg –30 dBm ±0.30 deg –100 dBm ±15.0 deg Specification Phase 100 1 10 Accuracy (dB) Accuracy (dB) Magnitude 10 0.1 1 0.01 0.
Test port input (group delay)1 Table 26. Option 230/235/240/245/260/265/280/285/430/435/440/ 445/460/465/480/485 Description Specification Aperture (selectable) (frequency span)/(number of points - 1) Supplemental information Maximum aperture 25% of frequency span Minimum delay Limited to measuring no more than 180° of phase change within the minimum aperture.
Test port input (group delay)1 (continued) Table 27. Option 2D5/2K5/4D5/4K5 Description Specification Aperture (selectable) (frequency span)/(number of points - 1) Supplemental information Maximum aperture 25% of frequency span Minimum delay Limited to measuring no more than 180° of phase change within the minimum aperture. Accuracy See graph below, typical The following graph shows group delay accuracy with 3.5 mm (male) connectors, full 2-port calibration and a 10 Hz IF bandwidth.
General Information Table 28. System bandwidth Description General characteristics IF bandwidth settings Range 10 Hz to 500 kHz Nominal settings are: 10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1 k, 1.5 k, 2 k, 3 k, 4 k, 5 k, 7 k, 10 k, 15 k, 20 k, 30 k, 40 k, 50 k, 70 k, 100 k, 150 k, 200 k, 300 k, 400 k, 500 kHz Table 29.
Table 30. Rear panel information Description Typical General characteristics External trigger input connector Type BNC, female Input level Low threshold voltage: 0.5 V High threshold voltage: 2.
Table 31.
Table 32. Rear panel information (continued) Description Typical General characteristics Line power2 Frequency 47 Hz to 63 Hz Voltage 90 to 132 VAC, or 198 to 264 VAC (automatically switched) VA max 350 VA max.
Table 34. EMC, safety and environment Description General characteristics EMC European Council Directive 2004/108/EC IEC 61326-1:2005 EN 61326-1:2006 CISPR 11:2003+A1:2004 EN 55011:2007 Group 1, Class A IEC 61000-4-2:1995 +A2:2000 EN 61000-4-2:1995 +A2:2001 4 kV CD / 8 kV AD IEC 61000-4-3:2006 EN 61000-4-3:2006 1-3 V/m, 80-1000 MHz / 1.4 GHz - 2.7 GHz, 80% AM IEC 61000-4-4:2004 EN 61000-4-4:2004 1 kV power lines / 0.5 kV signal lines IEC 61000-4-5:2005 EN 61000-4-5:2006 0.
Table 35. Analyzer environmental specifications and dimensions Description General characteristics Operating environment Temperature +5 °C to +40 °C Error-corrected temperature range 23 °C (±5 °C) with < 1 °C deviation from calibration temperature Humidity 20% to 80% at wet bulb temperature < +29 °C (non-condensation) Altitude 0 to 2,000 m (0 to 6561 feet) Vibration 0.
22 19 22 222 426 27 58 20 51 122 37 Figure 3. Dimensions (front view, E5071C with Option 230/235/240/245/260/265/280/285, in millimeters) 22 19 22 222 426 27 58 20 51 41 41 41 Figure 4.
26 19 22 222 426 27 58 20 51 122 Figure 5. Dimensions (front view, E5071C with Option 2D5/2K5, in millimeters) Figure 6.
424 422 128 1 96 21 18 27 25 25 290 25 58 109 148 25 144 146 159 179 190 215 1 25 25 25 165 Figure 7. Dimensions (rear view with Option 1E5, in millimeters) 472 23.4 417 13 18 235 215 222 32 15 Figure 8. Dimensions (side view, E5071C Option 230/235/240/245/260/265/280/285/430/435/440/445/460/465/480/485 , in millimeters) 472 417 23 13 18 215 222 235 32 14 Figure 9.
Measurement Throughput Summary Cycle time for measurement completion1, 2 (ms) Table 36. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: OFF Number of points 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 3.4 4.9 7.1 20 2-port cal 6.3 9.3 14 41 4-port cal 13 19 28 92 Start 1 GHz, stop 1.
Table 37. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 (continued) Number of points 51 201 401 1601 Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 15 18 19 24 2-port cal 28 35 38 48 4-port cal 55 69 74 104 Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 15 18 20 26 2-port cal 28 36 38 52 4-port cal 56 70 75 117 Start 100 kHz, stop 8.
Cycle time for measurement completion1, 2 (ms) Table 38. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: ON Number of Points 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 3.5 5.5 8.0 24 2-port cal 6.4 9.8 15 49 4-port cal 13 20 29 108 Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 3.6 5.4 8.
Table 39. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 (continued) Number of points 51 201 401 1601 Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 15 18 20 27 2-port cal 28 35 38 52 4-port cal 55 69 74 111 Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 15 18 20 27 2-port cal 28 36 38 56 4-port cal 56 70 75 120 Start 100 kHz, stop 8.
Cycle time for measurement completion1, 2 (ms) Table 40. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Sweep mode: Stepped, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: ON Number of Points 51 201 401 1601 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth Uncorrected 4.4 7.7 12 29 2-port cal 8.2 15 22 56 4-port cal 16 29 45 126 Start 1 GHz, stop 1.2 GHz, 100 kHz IF bandwidth Uncorrected 4.8 9.
Table 41. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 (continued) Number of points 51 201 401 1601 Start 100 kHz, stop 8.5 GHz, 500 kHz IF bandwidth Uncorrected 7.5 13 19 48 2-port cal 15 26 37 94 4-port cal 29 50 73 189 Start 100 kHz, stop 8.5 GHz, 100 kHz IF bandwidth Uncorrected 7.9 15 22 60 2-port cal 16 29 43 120 4-port cal 30 56 85 237 Start 100 kHz, stop 8.
Cycle time for measurement completion1, 2 (ms) Table 42.
Table 43. Option 2D5/2K5/4D5/4K5 (continued) Number of points 51 201 401 1601 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth Uncorrected 3.0 3.2 4.9 17 2-port cal 5.6 6.1 9.9 37 4-port cal 11 14 23 84 Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth Uncorrected 3.0 3.4 5.5 18 2-port cal 5.6 6.
Cycle time for measurement completion1, 2 (ms) Table 44.
Table 45. Option 2D5/2K5/4D5/4K5 (continued) Number of points 51 201 401 1601 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth Uncorrected 3.0 3.4 5.9 20 2-port cal 5.6 6.6 12 43 4-port cal 11 14 26 98 Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth Uncorrected 3.0 3.5 5.8 20 2-port cal 5.6 6.
Cycle time for measurement completion1, 2 (ms) Table 46.
Table 47. Option 2D5/2K5/4D5/4K5 (continued) Number of points 51 201 401 1601 Start 11 GHz, stop 12 GHz, 500 kHz IF bandwidth Uncorrected 2.8 6.4 11 31 2-port cal 5.1 13 21 61 4-port cal 9.6 24 41 131 Start 11 GHz, stop 12 GHz, 100 kHz IF bandwidth Uncorrected 3.4 8.0 14 43 2-port cal 5.
Cycle time (ms)1, 2 vs. number of points Table 48. Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Start 1 GHz, stop 1.2 GHz, 500 kHz IF bandwidth, error correction: OFF, display update: OFF, number of traces = 1 Number of points Sweep mode: Fast swept, system error correction: OFF Sweep mode: Fast swept, system error correction: ON Sweep mode: Std stepped, system error correction: ON 3 3.1 3.1 3.4 11 3.3 3.3 3.3 51 3.5 3.5 4.4 101 3.9 4.1 5.6 201 4.9 5.5 7.
Cycle time (ms)1, 2 vs. IF bandwidth Table 50. All options Sweep mode: Swept, analyzer display turned off with: DISP:ENAB OFF, number of traces = 1, system error correction: OFF, NOP = 201 Option 230/235/240/245/260/265/280/ 285/430/435/440/445/460/465/480/485 Option 2D5/2K5/4D5/4K5 Frequency [Hz] IFBW [Hz] Cycle time [ms] Frequency [Hz] IFBW [Hz] Cycle time [ms] 4G 500000 5.0 10 G 500000 2.8 4G 400000 5.0 10 G 400000 2.8 4G 300000 4.9 10 G 300000 2.8 4G 200000 4.
Data transfer time1, 2 (ms) Table 51.
E5092A Configurable multiport test set The section provides test input/output performance without calibration by the E5071C. Table 52. Test set input/output performance Description Specification Frequency range 50 MHz to 20 GHz Damage level Typical 20 dBm, ±35 VDC Table 53.
Table 54. Option E5092A-020 port performance (continued) Description Specification Typical Load match (common port) SPDT switch1 50 MHz to 2 GHz 16 dB 2 GHz to 4 GHz 11 dB 4 GHz to 8 GHz 8 dB 8 GHz to 10 GHz 7 dB 10 GHz to 20 GHz 4 dB SP4T switch2 50 MHz to 1.3 GHz 16 dB 1.3 GHz to 4 GHz 11 dB 4 GHz to 8 GHz 8 dB 8 GHz to 10 GHz 7 dB 10 GHz to 20 GHz 4 dB Insertion loss SPDT switch1 4 dB 50 MHz to 100 MHz 3.5 dB 100 MHz to 2 GHz 4.5 dB 2 GHz to 3 GHz 5 dB 3 GHz to 4 GHz 5.
Table 55. Control line Description Specification Number of groups 4 Group A: 8 bits Group B,C,D: 4 bits Input voltage range1 0 V to +5 V (positive input) –5 V to 0 V (negative input) Maximum current Group A, B: 50 mA in total of each group Group C, D: 500 uA in total of each group Typical Impedance Group A, B: < 10 ohm Group C, D: < 200 ohm Table 56. DC source Description Specification Number of sources 4 Typical 0 V to +5.
Table 59. Front panel information Description General characteristics RF connectors SMA (Female) Test ports 38 ports Control line 15-pin D-sub, female 25-pin D-sub, female Table 60. Rear panel information Description General characteristics USB port Type B-receptacle, provide connection to the E5071C Line power1 Frequency Voltage 47 to 63 Hz 90 to 132 VAC, or 198 to 264 VAC (automatically switched) 300 VA max. VA max Table 61.
40.6 40.6 36.8 17 40.6 27 23.38 132.72 27 27 27 27.8 13.5 27 65.34 27 18.57 23.38 27 69.8 Figure 11. Dimensions (pitch between switches, with Option E5092A-020, in millimeters, nominal) 37 51 83 133 129 Figure 12. Dimensions (rear view, with Option E5092A-020, in millimeters, nominal) 472 417 22 133 146 32 13 18 144 126 23 Figure 13.
SW 1 PORT 1 1A 1B 1C 1D SW 5 5A 5B PORT 2 SW 2 SW 3 6A 6B 6C O M 5C O M 7A 7B SW 7 SW 4 3A 3B 3C 3D 2A 2B 2C 2D SW 6 PORT 3 SW 8 7C O M 8A 8B SW 9 4A 4B 4C 4D 9A 9B 9C O M 8C O M PORT 4 S W 10 10A 10B 10C O M Figure 14.
23 22 21 20 19 18 17 16 Positive input DC source GND Negative input Positive input DC source GND Negative input 12 11 GROUP B 10 9 8 7 6 5 GROUP C 12 11 10 9 GND Negative input Positive input DC source Reserved Reserved Reserved Line 8 Line 7 Line 6 Line 5 Line 4 Line 3 Line 2 Line1 8 7 6 5 4 3 2 1 GROUP C 4 3 Line 1 13 Line 2 Line 3 Line 4 Line 1 GROUP B Line 2 Line 3 Line 4 Line 1 Line 2 Line 3 Line 4 GND 14 15 14 DC source 24 Negative input 13 15 Positive input 25 GND D
Corrected System Performance for 75 Ω Measurements with 11852B 50 to 75 Ω Minimum-Loss Pads (Supplemental Information) Option 230/235/240/245/260/265/280/285/430/435/440/445/ 460/465/480/485 Table 62.
www.agilent.com www.agilent.com/find/ena For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium.