Agilent E5063A ENA Series Network Analyzer 100 kHz to 4.5/ 8.
Deinitions Specification (spec.): Warranted performance. All speciications apply at 23 ºC (± 5 ºC), unless otherwise stated, and 90 minutes after the instrument has been turned on. Speciications include guard bands to account for the expected statistical performance distribution, measurement uncertainties, and changes in performance due to environmental conditions. Typical (typ.): Expected performance of an average unit which does not include guardbands. It is not covered by the product warranty.
Corrected System Performance The specifications in this section apply to measurements made with the Agilent E5063A network analyzer under the following conditions: • No averaging applied to data • Environmental temperature of 23 °C (± 5 °C) with less than 1 °C deviation from the calibration temperature • Response and isolation calibration performed System Dynamic Range Description Specification Typical System dynamic range at test port 1 (IF Bandwidth = 3 kHz) 100 kHz to 300 kHz 63 dB 300 kHz to 8.
Corrected system performance with calibration kit Corrected system performance with type-N device connectors, 85032F calibration kit Network analyzer : E5063A Calibration kit : 85032F (Type-N, 50 Ω) Calibration : Full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from calibration temperature, isolation calibration performed Specification (dB) Description 100 kHz to 10 MHz 10 MHz to 3 GHz 3 to 9 GHz Directivity 49 46 38 Sou
Corrected system performance with type-N device connectors, 85092C electronic calibration (ECal) module Network analyzer : E5063A Calibration kit : 85092C (Type-N, 50 Ω) Electronic calibration (ECal) module Calibration : Full 2-port IF bandwidth = 10 Hz, no averaging applied to data, environmental temperature = 23 °C (± 5 °C) with < 1 °C deviation from calibration temperature, isolation calibration is not performed Specification (dB) Description 300 kHz to 10 MHz 10 MHz to 3 GHz 3 to 9 GHz Directivity
Corrected system performance with 3.5 mm device connector type, 85033E calibration kit Network analyzer : E5063A Calibration kit : 85033E (3.
Corrected system performance with 3.5 mm device connector type, 85093C electronic calibration (ECal) module Network analyzer : E5063A Calibration kit : 85093C (3.
Corrected system performance with 3.5 mm device connector type, 85052D calibration kit Network analyzer : E5063A Calibration kit : 85052D (3.
Corrected system performance with 3.5 mm device connector type, N4691B electronic calibration (ECal) module Network analyzer : E5063A Calibration kit : N4691B (3.
Uncorrected System Performance User correction: OFF System error correction: ON Specification (dB) Description 100 kHz to 300 kHz 300 kHz to 1 MHz 1 to 100 MHz 100 MHz to 3 GHz 3 to 6 GHz 6 to 10 GHz 10 to 13 GHz 13 to 18 GHz Directivity 10 dB 10 dB 25 dB 25 dB 20 dB 15dB 10 dB 10 dB Source match 20 dB 20 dB 25 dB 25 dB 20 dB 15dB 15 dB 15 dB Load match 7 dB (typ.) 11 dB (typ.) 14 dB 11 dB 10 dB 7dB 8 dB (typ.) 6 dB (typ.) Relection tracking ± 3.0 dB ± 3.0 dB ± 1.
Test Port Output (Source) Test port output frequency Description Speciication Frequency range Option 245 Option 285 Option 2H5 100 kHz to 4.5 GHz 100 kHz to 8.5 GHz 100 kHz to 18 GHz Resolution Typical 1 Hz (100 kHz to 6.5 GHz) 2 Hz (6.5 to 13 GHz) 11 Hz (13 to 18 GHz) Source stability CW accuracy ± 7 ppm (5 to 40 ºC) ± 7 ppm Test port output power Description Speciication Nominal power (preset power) –5 dBm Range 100 kHz to 300 kHz 300 kHz to 8.5 GHz 8.
Test Port Input Description Speciication Test port input level Maximum input level Damage level +6 dBm Typical +26 dBm or ± 35 VDC Crosstalk 100 kHz to 300 kHz 300 kHz to 8.5 MHz 8.5 MHz to 4.34 GHz 4.34 to 6 GHz 6 to 13 GHz 13 to 16 GHz 16 to 18 GHz –88 dB –93 dB –115 dB –105 dB –100 dB –90 dB –85 dB Test Port Noise Floor 100 kHz to 8.5 MHz 8.5 to 100 MHz 100 MHz to 4.34 GHz 4.34 to 8.5 GHz 8.
Trace noise Description Speciication Typical Magnitude Transmission: 100 kHz to 300 kHz, 3 kHz IFBW 300 kHz to 8.5 MHz, 3 kHz IFBW 8.5 MHz to 4.34 GHz, 70 kHz IFBW 4.34 to 8.5 GHz, 70 kHz IFBW 8.5 to 13 GHz, 70 kHz IFBW 13 to 16 GHz, 70 kHz IFBW 16 to 18 GHz, 70 kHz IFBW 8 mdB rms 6 mdB rms 5 mdB rms 10 mdB rms 15 mdB rms 25 mdB rms 30 mdB rms 5 mdB rms 3 mdB rms 2 mdB rms 5 mdB rms 8 mdB rms 15 mdB rms 20 mdB rms Relection: 100 kHz to 300 kHz, 3 kHz IFBW 300 kHz to 8.5 MHz, 3 kHz IFBW 8.5 MHz to 4.
Stability 1 Description Speciication Typical Magnitude Transmission: 100 kHz to 300 kHz 300 kHz to 6 GHz 6 to 12 GHz 12 to 18 GHz ± 0.02 dB/°C ± 0.01 dB/°C ± 0.025 dB/°C ± 0.04 dB/°C Relection: 100 kHz to 300 kHz 300 kHz to 6 GHz 6 to 12 GHz 12 to 18 GHz ± 0.02 dB/°C ± 0.02 dB/°C ± 0.035 dB/°C ± 0.05 dB/°C Phase Transmission: 100 kHz to 300 kHz 300 kHz to 6 GHz 6 to 12 GHz 12 to 18 GHz ± 0.4 deg/°C ± 0.2 deg/°C ± 0.5 deg/°C ± 0.
Group delay 1 Description Speciication Aperture (selectable) (frequency span)/(number of points - 1) Typical Maximum aperture 25% of frequency span Minimum delay Limited to measuring no more than 180° of phase change within the minimum aperture. Accuracy See graph below (typical) 1. Group delay is computed by measuring the phase change within a specified step (determined by the frequency span and the number of points per sweep). The following graph shows group delay accuracy with 3.
General Information Description System bandwidth Range General characteristics 10, 15, 20, 30, 40, 50, 70, 100, 150, 200, 300, 400, 500, 700, 1 kHz, 1.5 kHz, 2 kHz, 3 kHz, 4 kHz, 5 kHz, 7 kHz, 10 kHz, 15 kHz, 20 kHz, 30 kHz, 40 kHz, 50 kHz, 70 kHz, 100 kHz, 150 kHz, 200 kHz, 300 kHz Front panel Description Typical General characteristics Test ports Type-N, female, 50 Ω (nominal) Display Type Resolution 10.
Description Typical General characteristics Video output 15-pin mini D-Sub, female; drives VGA compatible monitors GPIB 24-pin D-Sub (Type D-24), female; compatible with IEEE-488 USB host port Universal serial bus jack, type A coniguration, female; provides connection to mouse, keyboard, printer, ECal module, USB coaxial switch, or USB/GPIB interface USB (USBTMC 1) interface port Universal serial bus jack, type B coniguration (4 contacts inline), female; provides connection to an external PC; comp
EMC, safety, environment and compliance Description General characteristics EMC European Council Directive 2004/108/EC IEC 61326-1:2005 EN 61326-1:2006 CISPR 11:2003+A1:2004 EN 55011:2007 Group 1, Class A IEC 61000-4-2:1995 +A2:2000 EN 61000-4-2:1995 +A2:2001 4 kV CD/8 kV AD IEC 61000-4-3:2006 EN 61000-4-3:2006 1-3 V/m, 80-1000 MHz/1.4 GHz - 2.7 GHz, 80% AM IEC 61000-4-4:2004 EN 61000-4-4:2004 1 kV power lines/0.5 kV signal lines IEC 61000-4-5:2005 EN 61000-4-5:2006 0.
Analyzer environmental speciications and dimensions Description General characteristics Operating environment Temperature Error-corrected temperature range Humidity Altitude Vibration +5 °C to +40 °C 23 °C (± 5 °C) with < 1 °C deviation from calibration temperature 20% to 80% at wet bulb temperature < +29 °C (non-condensation) 0 to 2,000 m (0 to 6561 feet) 0.
Dimensions (rear view) Dimensions (side view) (in millimeters) 20
Measurement Throughput Summary Measurement throughput data is typical performance data. Common condition for the measurement throughput data: • Analyzer display turned off with: DISP : ENAB OFF • Number of traces = 1 • firmware version: A.01.0x Cycle time for measurement completion Number of Points 51 Start 1 GHz, stop 1.2 GHz 1-port cal, S11 4 Response cal, S21 4 2-port cal, S21 8 Start 100 kHz, stop 4.5 GHz 1-port cal, S11 8 Response cal, S21 9 2-port cal, S21 16 Start 100 kHz, stop 8.
www.agilent.com myAgilent www.agilent.com/find/myagilent A personalized view into the information most relevant to you. www.lxistandard.org LAN eXtensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. Three-Year Warranty www.agilent.com/find/ThreeYearWarranty Beyond product specification, changing the ownership experience.