Agilent E5052A Signal Source Analyzer 10 MHz to 7, 26.
Definitions All specifications apply over a 18 °C to 28 °C range (unless otherwise stated) and 30 minutes after the instrument has been turned on. All specified and supplemental values for RF input signals are applicable to sinusoidal-wave carriers unless otherwise noted. Specification (spec.): Warranted performance.
Table 1-2. SSB phase noise sensitivity (standard, < 150 kHz optim., correlation = 1, + 5 dBm input, start frequency = 1 Hz, measurement time = 17.7 sec) Offset from carrier (Hz) Input frequency 10 MHz -114.0 -144.5 -54.0 -94.0 -135.5 -34.0 -77.0 -115.5 Spec. Typ. 7 GHz -74.0 Spec. Typ. 3 GHz 100 Spec. Typ. 1 GHz 10 Spec. Typ. 100 MHz 1 -24.5 -64.5 -106.0 Spec. Typ. -17.1 -57.1 -98.6 1k 10 k 100 k 1M 10 M 40 M -148.5 -156.5 -166.5 -168.5 — — -152.5 -160.5 -170.
Figure1-1. SSB phase noise sensitivity (typical) (standard, < 150 kHz optim., correlation = 1, +5 dBm input, start offset frequency = 1 Hz, measurement time = 17.7 sec) Figure1-3. SSB phase noise sensitivity improved with the crosscorrelation function (typical) (standard, improvement with the correlation, carrier frequency = 1 GHz, < 150 kHz optim., +5 dBm input) Figure1-2. SSB phase noise sensitivity (E5052A-011, typical) (standard, < 150 kHz optim.
Table 1-9. DC current measurement Description Specifications Current range 0 to 80 mA Resolution 10 µA Accuracy ± (0.2% of reading + 160 µA) Table 1-5. Typical measurement time (sec) for phase noise measurement 1 Start frequency (Hz) Stop frequency (Hz) 1 10 100 1k 100 k 10.9 2.7 0.34 0.04 1M 10.7 2.7 0.34 0.04 10 M 11.1 2.8 0.35 0.04 40 M 17.7 4.4 0.56 0.07 Table 1-6.
Transient Measurement Table 1-12. Transient measurement Description Specifications Measurement function Frequency, power, phase Target frequency range 10 MHz to 7 GHz Input power level -20 to +20 dBm Frequency bandwidth Wide band See Table 1-13 Narrow band 3.125 kHz, 25 kHz, 200 kHz, 1.6 MHz, or 25.
Table 1-15. Narrow band transient (frequency bandwidth = 3.125 kHz) Measurement time (sec) 100 m 200 m 500 m 1 2 5 10 Frequency resolution (Hz rms) 0.0095 0.0095 0.0034 0.0012 0.0004 0.0004 0.0004 Time resolution (µsec) 81.92 163.84 409.6 819.2 1638.4 4096 10240 Number of point 1222 1222 1222 1222 1222 1222 978 Table 1-16.
Port Output (DC Power/Control) Table 1-21. DC power voltage output Description Specifications Voltage range 0 to 16 V Resolution 1 mV Setting accuracy ± (0.2% + 2 mV) Maximum output current 80 mA Noise < 10 nV /√Hz at 10 kHz Output resistance < 0.3 ohm (typical) Settling time @error: 0.1% < 20 msec (typical) Table 1-22. DC control voltage output Description Specifications Voltage range -15 to 35 V Resolution 0.1 mV Setting accuracy -15 to 0 V ± ((Setting + 15 V) x 0.
Table 1-26.
Figure 1-5. Dimensions (front view, in millimeters, nominal) Table 1-28. Measurement capabilities Number of Up to 4 measurement windows measurement windows: and 1 user defined window Data markers: 6 independent markers per trace. Reference marker available for delta marker operation.
Table 1-31. Storage Internal hard disk drive: Store and recall instrument states and trace data on 10 GB, minimum, internal hard drive. Trace data can be saved in CSV (comma separated value) format. All files are MS-DOS®-compatible. Instrument states include all control settings and memory trace data. File sharing: Internal hard disk drive (F:) can be accessed from an external Windows® PC through LAN or USB (USBTMC) Disk drive: Instrument states and trace data can be stored on an internal 3.5 inch 1.
System Performance with the E5053A Downconverter The system performance is the combination of the E5052A SSA and E5053A downconverter. All data is typical performance. Table 2-1. System performance characteristics Description Test port Frequency range Input level Carrier search range 1 Phase noise measurement Frequency band Performance characteristics 10 MHz to 3 GHz (E5052A SSA RF IN port) 3 to 26.
System Performance with the E5053A Downconverter Table 2-2. System phase noise sensitivity (typical) Input frequency 3 GHz 10 GHz 18 GHz 26.5 GHz Offset from carrier (Hz) 1 10 -39.0 -78.7 -34.6 -71.8 -31.6 -66.5 -29.0 -63.2 100 -99.5 -91.1 -85.3 -81.8 1k -124.4 -116.1 -110.3 -106.8 10 k -135.3 -124.1 -121.2 -117.8 100 k -137.5 -128.1 -125.3 -121.8 1M -153.5 -147.1 -141.3 -137.8 10 M -164.5 -156.6 -150.7 -147.3 40 M -167.6 -160.0 -154.2 -150.8 Figure 2-2.
E5053A Microwave Downconverter Specification Summary Downconverter Test Port Table 2-3. Input/output Description Test port input Frequency range Maximum Input level Damage level LO output Output frequency LO resolution Output power Specification 3 to 26.5 GHz +10 dBm (3 to 10 GHz band) +5 dBm (9 to 26.
Figure 2-4. Dimensions (front view, in millimeters, nominal Figure 2-5. Dimensions (rear view, in millimeters, nominal) Figure 2-6.
Web Resources www.agilent.com Visit our Signal Source Analyzer Web site for additional product information and literature: www.agilent.com/find/ssa For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is available at: Phase noise measurements: www.agilent.com/find/phasenoise RF and microwave accessories: www.agilent.com/find/accessories Windows‚ and Visual Basic‚ are U.S.