Agilent ESA-L Series Spectrum Analyzers Data Sheet Available frequency ranges E4411B E4403B E4408B 9 kHz to 1.5 GHz 9 kHz to 3.0 GHz 9 kHz to 26.5 GHz As the lowest cost ESA option, these basic analyzers are ideal for cost conscious bench-top or manufacturing environments. Customers looking for a more portable solution would benefit from the new Agilent N9340B handheld RF spectrum analyzer.
The ESA-L Series spectrum analyzers are tested to ensure they will meet their warranted performance. Unless otherwise stated, all specifications are valid over 0 to 55 °C. Supplemental characteristics, shown in italics, are intended to provide additional information that is useful in using the instrument.
ESA-L Express Analyzer Option BAS or BTG Receive faster delivery and a favorable price when you order the ESA-L express analyzer Option BAS or BTG. This express analyzer is configured based on the most frequently ordered ESA-L configuration and most popular options. The express analyzer options simplify the ordering process while maintaining the flexibility of the ESA platform. Choose your frequency range: E4411B E4403B E4408B 9 kHz to 1.5 GHz 9 kHz to 3.0 GHz 9 kHz to 26.
Frequency Specifications Frequency range E4411B E4403B E4408B 9 kHz - 1.5 GHz 9 kHz - 3 GHz 9 kHz - 26.5 GHz (75 Ω input Option 1DP) 1 MHz - 1.5 GHz N/A N/A BAS/BTG configuration Custom configuration Frequency range 100 Hz - 3 GHz 2.85 - 6.7 GHz 6.2 - 13.2 GHz 12.8 - 19.2 GHz 18.7 - 26.
Frequency Specifications (continued) Basic analyzer Sweep time and trigger Range 4 ms - 4000 s Span = 0 Hz Span ≥ 100 Hz 4 ms - 4000 s ±1% Accuracy (Span = 0 Hz) Trigger type Free run, single, line, video, offset, delayed, external 1 µs to 400 s Delayed trigger range Sweep (trace) points Range Span = 0 Hz 401 Span ≥ 100 Hz 401 Basic analyzer Resolution bandwidths (1-3-10 sequence) Range (–3 dB) (–6 dB EMI) With 1DRb (–3 dB) (–6 dB EMI) Accuracy 1 to 300 Hz 1 kHz to 3 MHz 5 MHz Selectivity (60 dB
Frequency Specifications (continued) Basic analyzer E4411B E4403B/08B Stability Noise sidebands offset from CW signal with 1 kHz RBW, 30 Hz VBW and sample detector Offset from CW signal ≥ 10 kHz ≥ 20 kHz ≥ 30 kHz ≥ 100 kHz Specification and typical dBc/Hz applies to all frequencies ≤ 6.
Amplitude Specifications E4411B E4403B/08B Amplitude range Measurement range Input attenuator range (5 dB step) Displayed average noise level (DANL) to maximum safe input level 0 - 60 dB 0 - 65 dB Maximum safe input level Input attenuator setting Average continuous power Peak pulse powera DC voltage AC coupled ≥ 15 dB ≥ 5 dB average continuous power; ≥ 30 dB peak pulse power +30 dBm (1 W) +30 dBm (1 W) +50 dBm (100 W) 100 Vdc +75 dBmV (0.
Amplitude Specifications (continued) Basic analyzer E4411B E4403B E4408B Displayed average noise level (dBm) (input terminated, 0 dB attenuation, sample detector) specification Italics indicate typical performance 100 Hz RBW; 1 Hz VBW (Option 1DR); Conditions Frequency 1 - 10 MHz 10 - 500 MHz 500 MHz - 1 GHz 1 - 1.5 GHz 1.5 - 2 GHz 2 - 3 GHz 3 - 6 GHz 6 - 12 GHz 12 - 22 GHz 22 - 26.
Amplitude Specifications (continued) Basic analyzer Display Display range Log scale RBW ≥ 1 kHz RBW ≥ 300 Hz Linear scale Scale units Trace detectors Trace functions Marker readout resolution Log scale 0 to –85 dB 0 to –120 dB (1DR) Linear scale Reference level Range Resolution Log scale Linear scale Accuracyc –10 to > –60 dBm –60 to > –85 dBm –85 to > –90 dBm 0.1, 0.2, 0.
Amplitude Specifications (continued) Basic analyzer Input attenuator switching uncertainty (at 50 MHz) Attenuator setting 0 to 5 dB 10 dB 15 to 60 dB Frequency response (10 dB input attenuation) Absolute a 9 kHz to 3 GHz 3 to 6.7 GHz 6.7 to 13.2 GHz 13.2 to 26.5 GHz ±0.3 dB Reference ±(0.1 dB + 0.01 x attenuator setting) ±0.5 dB ±1.
Amplitude Specifications (continued) Basic analyzer E4411B/03B/08B Spurious responses Third order intermodulation distortion For two –30 dBm signals at input mixera and > 50 kHz separation 100 MHz to 26.5 GHz Second harmonic distortion 2 to 750 MHz –40 dBm tone at input mixer a 10 to 500 MHz –30 dBm tone at input mixer a 500 MHz to 1.5 GHz –30 dBm tone at input mixer a 1.5 to 2.
Tracking Generator Specifications Tracking generator specifications (Options 1DN and 1DQ) Frequency range E4411B Option 1DN, (50 Ω) Option 1DQ, (75 Ω ) RBW range Output power level range E4411B Option 1DN Option 1DQ Output vernier range E4411B Output attenuator range E4411B Output flatness E4411B Option 1DN, (50 W) 9 kHz to 10 MHz 10 MHz to 1.5 GHz Option 1DQ, (75 W) 1 to 10 MHz 10 MHz to 1.
General Specifications Basic analyzer E4411B E4403B E4408B Temperature range 0 to +55 ºC –40 to +75 ºC 10 to +40 ºC Operating Storage Disk drive EMI compatibility Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class A Conducted and radiated interference is in compliance with CISPR Pub. 11/1990 Group 1 Class Ba (Option 060) Audible noise sound pressure at 25 ° C Military specifications Power requirements < 40 dBa pressure and < 4.
General Specifications (continued) Inputs/outputs Front panel Input RF out Probe power External keyboard Headphone Power output AMPT REF out IF INPUT (Option AYZ) LO OUTPUT (Option AYZ) 50 Ω type N (f); 75 Ω BNC (f) (Option 1DP); 50 Ω APC 3.5 (m) (Option BAB) 50 Ω type N (f); 75 Ω BNC (f) (Option 1DQ) +15 Vdc, –12.6 Vdc at 150 mA maximum (characteristic) 6-pin mini-DIN, PC keyboards (for entering screen titles and file names) Front panel knob controls volume 0.
For More Information For the latest information on the Agilent ESA-L Series see our Web page at: www.agilent.com/find/esa Remove all doubt Our repair and calibration services will get your equipment back to you, performing like new, when promised. You will get full value out of your Agilent equipment throughout its lifetime. Your equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts.