User`s guide
Operation
91
3. Ensure the GC and any related support equipment and
software are installed and are displaying a ready status.
i.e. the GC detectors, GC data acquisition systems, etc.
4. If you have any options installed, ensure they are turned
on and allowed to warm-up as required. Refer to the
documentation supplied with each installed accessory.
5. Press the On switch. The power switch is located on the
rear panel of the ATD/TD.
6. The ATD/TD splash screen will be displayed.
7. The log in message will then be displayed. Enter the
password and press OK. (The default password is
“hstd”.)
8. The local ATD/TD control interface will be displayed.
9. Control Software is available that allows you to control
the instrument through your computer. See TD Control
Software on page 199 for more details. If you are using
the TD Control Software, connect the computer to the
instrument as outlined in the control software manual.
10. After power on, the instrument starts an initialization
process lasting approximately 10 seconds and set up the
system with the last method used.
The Touch Screen Display for all
Models of the TurboMatrix Series
You will control the operation of the instrument through the touch
screen display. Lightly touch the screen to enter data and issue
commands, using a stylus or finger.
NOTE: Do not use sharp objects such as pens and pencils to
activate the touch screen as you may damage the display.
Using the three tabs, you can control the operation of the ATD/TD.
The touch screen interface allows you to communicate directly with
the ATD/TD so that you can: