User`s guide
Operation
87
Safety Precautions for the all Models of
the TurboMatrix Series
To protect yourself from harm and to avoid damaging the
instrument, please observe the following notes:
• Before using this instrument, read and observe the safety
information in Chapter 1.
• Do not sample carbon disulfide or other solvents with a self-
ignition point of 100°C or less.
• Always follow the correct safety procedures and the
manufacturer’s recommendations when using any solvent.
Refer to the MSDS sheets for compound-specific
information.
• Do not touch moving parts of the instrument during
operation.
• Do not operate the instrument with any covers or parts
removed.
• A supply of carrier gas must be connected to the ATD/TD
and a column flow established before the GC is powered up.
• Do not start the ATD/TD until you have established a flow
of carrier gas and dry air through the system.
Supplies—The ATD/TD is shipped without sample tubes or caps
.
These parts are necessary to operate the ATD/TD and are available
either individually or in the ATD/TD Starter Kit (M041-3541).
Before beginning operation ensure that you have adequate supplies
of tubes, caps, and other sample handling materials. Ensure that the
materials you are using are compatible both with your sample and
your sampling method.
CAUTION
The Peltier device will always be cooling while the
instrument is switched on.You must connect the dry air purge
before switching on the ATD/TD so that the Peltier cooler
and trap do not become covered with ice from the ambient
air. This will extend the operational life of the Peltier cooling
elements and other components.